Instructions for Use
Zest CHAIRSIDE® Denture Prep and Polish Kit
IMPORTANT: This document contains the most current Instructions for Use. Please read and retain.
PRODUCT DESCRIPTION
The Zest CHAIRSIDE Denture Prep and Polish Kit contains the necessary instruments to prepare and
finish an overdenture for pickup of a LOCATOR® or SATURNO® attachment. The system includes a recess
bur, undercut bur, vent hole bur, trim bur, grind bur, and polisher.
INDICATIONS FOR USE
The Zest CHAIRSIDE Denture Prep and Polish Kit is intended to prepare and polish full and partial acrylic
dentures as part of a chairside or laboratory procedure.
CAUTION
Federal (USA) law restricts this device to sale by or on the order of a licensed dentist.
MULTI-USE DEVICES
CHAIRSIDE Denture Prep and Polish Kit are multi-use devices and are provided non-sterile. Reusable
tools and instruments must be cleaned and sterilized prior to use or reuse on patients. Reuse of a multi-
use devices without cleaning and sterilization may cause harm to the patient in the transfer of blood,
tissue, or saliva that may contain infectious disease.
PREPARATION
Impurities need to be removed from the products directly after use (within two hours
maximum). To do so, use running water or an FDA approved and/or CE Marked disinfectant
solution. The disinfectant's effectiveness should be established, suitable, and compatible with
the instruments.
CLEANING INSTRUCTIONS
Place the instruments in the cleaning bath for the prescribed contact time and concentration,
take care that the instruments do not contact each other.
In case of severe contamination of the instrument, it is recommended to use an ultrasonic bath
for cleaning.
The Bur Block must be cleaned and disinfected without instruments loaded.
The Bur Block is not suitable for cleaning using a thermo disinfector (disinfector RDG).
Remove instruments from the cleaning solution and rinse for at least 3 minutes with tap water.
Remove excess moisture from the components with a clean, absorbent, and non-shedding wipe.
RECOMMENDED STERILIZATION INSTRUCTIONS
Page 1 of 4

Advertisement

loading

Summary of Contents for Zest Dental CHAIRSIDE

  • Page 1 INDICATIONS FOR USE The Zest CHAIRSIDE Denture Prep and Polish Kit is intended to prepare and polish full and partial acrylic dentures as part of a chairside or laboratory procedure.
  • Page 2 ENGLISH • All components and instruments are supplied NON-STERILE. Check instruments for damaged surfaces, chipped off edges, deformations and eliminate • damaged instruments. Products that are still contaminated need to be cleaned and disinfected until all visible contaminants have been removed. •...
  • Page 3 12,000 No coolant is needed during use. STORAGE AND HANDLING The Zest CHAIRSIDE Prep and Polish Kit in its undamaged, original packaging is not subject to any special considerations for storage or handling (during transport and storage). DISPOSAL Dispose of used devices which pose a risk of infection according to facility clinical waste procedures and applicable local and state regulations.
  • Page 4 9464 Rüthi, Switzerland www.zestdent.com Made in Switzerland +41 (0) 71 767 79 99 CHAIRSIDE and ZEST are registered trademarks and Zest Dental Solutions is a trademark of ZEST IP Holdings, LLC U.S. Patents: Refer to www.zestdent.com/patents L8114-ZD Rev D 2021-09...