Tonghui Electronics TH6400 Series Operation Manual

Programmable dc power supply

Advertisement

Quick Links

Operation Manual
TH6400 Series
Programmable DC Power Supply
Changzhou Tonghui Electronic Co., Ltd.
www.tonghui.com.cn

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TH6400 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Tonghui Electronics TH6400 Series

  • Page 1 Operation Manual TH6400 Series Programmable DC Power Supply Changzhou Tonghui Electronic Co., Ltd. www.tonghui.com.cn...
  • Page 2: Table Of Contents

    TH64 Series Operation Manual Contents Chapter 1 Introduction........................1 Introduction to Instrument .................. 1 Working Condition ..................... 1 Chapter 2 Precautions for Usage ....................2 2.1 Dimension and Weight ....................2 2.2 Safety Requirements ......................2 2.3 Electromagnetic Compatibility ..................2 Chapter 3 Panel ..........................4 3.1 Front panel ........................
  • Page 3 TH64 Series Operation Manual 4.4.4 R-Wait ......................... 11 4.4.5 Bus Mode ........................11 4.4.6 Baud Rate ........................11 4.4.7 GPIB-Addr ........................11 4.4.8 Combine ........................11 4.4.9 Date ..........................12 Chapter 5 Operation ........................ 13 Chapter 6 Interface and communication ................14 6.1 Remote Control ......................
  • Page 4 TH64 Series Operation Manual Declaration The descriptions contained in this manual may not cover all information about this instrument. Introductions to the improvements of the instrument in performance, function, internal structure, outer appearance, accessories, packing material, etc. are subject to change without notice. If you find any inconformity of this manual with our instruments, please contact us for further consultation by the address listed on the cover.
  • Page 5: Chapter 1 Introduction

    1.1 Introduction to Instrument TH6400 series is a programmable linear DC power supply with double range. The series of instruments have powerful functions and superior performance. The use of LCD screen display makes the display clear and menu operation easy, which adapt to the needs of rapid operation on the production site and high precision and stability of high-precision laboratory.
  • Page 6: Chapter 2 Precautions For Usage

    TH64 Series Operation Manual Chapter 2 Precautions for Usage ⚫ Please do not use the tester in dusty, vibrative, direct sunlight and corrosive gases and other adverse environments. ⚫ When the instrument is not used for a long time, please put it in the original box or similar box and stored in a ventilated room with temperature of 5℃~40℃...
  • Page 7 TH64 Series Operation Manual Radiated interference, refer to requirements of GB6833.10.
  • Page 8: Chapter 3 Panel

    TH64 Series Operation Manual Chapter 3 Panel In this chapter, there is only rough description, please refer to more information in chapter 4. 3.1 Front panel Name Function ① USB interface Import or export the data via the USB interface ②Trademark and model Mark the manufacturer and model of the instrument ③LCD display...
  • Page 9: Rear Panel

    TH64 Series Operation Manual 3.2 Rear panel GPIB Communicated with PC to build GPIB test system Controlled by PC via USB DEVICE RS232 Function is same as No.2 Power socket Connecting 220V/50HZ AC power supply Remote test Same function as output in front panel, 4- terminal terminal sampling function is added Nameplate...
  • Page 10 TH64 Series Operation Manual (2.) Status and error information zone: Display the error information or current working status. (3.) Fast setting and display zone: In this zone, you can set the voltage, current and real time sampling numbers & timing can be displayed . (4.) Softkey menu zone: The content is flexibly, display the function matched with current softkey...
  • Page 11: Chapter 4 Operation

    TH64 Series Operation Manual Chapter 4 Operation 4.1 Output display Press OUPUT DISPLAY to enter <OUTPUT> page <OUTPUT>is displayed as below You can set the voltage and current in 3 channels, also the measured voltage, current, power, output time and output status are displayed. 4.1.1 Value setting The setting range of voltage and current in each channel is from 0 to Max.value (The Max.
  • Page 12: Set

    TH64 Series Operation Manual 4.2 CH SET CH 和 CH to switch Press CH SET to enter CH1 SET, and set another channels by CH2 SET is displayed as below: On this page, you can set the parameter of one channel, the following parameter and function can be set: ⚫...
  • Page 13: Time Set

    TH64 Series Operation Manual above setting. OVP function is used to set the protective point. When the voltage is over the point then the power supply cuts the output, meanwhile OVP is displayed. 4.2.4 Time Set The range of time set is from 0 to 99999.9s, same as above setting. When the Time is on, the timer starts working in the process of outputting also the count down value is displayed.
  • Page 14: Recall And Delete

    TH64 Series Operation Manual 4.3.1 Recall and delete There are 4 pages in CH FILE, and 10 groups of data on each page, use the direction key or PAGE UP and PAGE DOWN to check the data. When you want to recall some data, just use knob to select the item then click RECALL, then the data is this group will cover the current setting.
  • Page 15: Key Voice

    TH64 Series Operation Manual 4.4.2 Key voice ⚫ ⚫ 4.4.3 P-Mem The function is used to select the user data or default ⚫ DEFAULT: Factory setting ⚫ User Data: When there are some settings in the tester, and the settings need to be loaded when turning on the tester next time, then you should select it.
  • Page 16: Date

    TH64 Series Operation Manual ⚫ Track:CH1+CH2 in synchronous mode, CH2+CH3 in synchronous mode and 3 channels in synchronous mode. In this mode, the set voltage and current are changed based on some ratio. Before selecting this mode, the voltage and current must be set in advance. E.g.
  • Page 17: Chapter 5 Operation

    TH64 Series Operation Manual Chapter 5 Operation (1) Connect to power and press power switch (2) Warm up 20 min (3) According to the actual requirement, select the suitable connection cable, connect the load to the [+,-] terminal in the tester. In Combine mode, the connection cable in the output terminal must be connected in serial or parallel.
  • Page 18: Chapter 6 Interface And Communication

    TH64 Series Operation Manual Chapter 6 Interface and communication RS232C, GPIB and USB are available in this tester, which can be used for data communication and remote control but not at the same time. The command is same for all ports, but the hardware configuration and communication protocol is different.
  • Page 19: Communicate With Pc

    TH64 Series Operation Manual 6.1.1.2 Communicate with PC The connection of tester to PC is displayed as below: Based on the table above, the pin definition of this tester is same as the 9-core connector of IMB AT. User can make 3 wore connection cable (length is less than 1.5m) by using double core shielding wire or buy from us.
  • Page 20 TH64 Series Operation Manual the number of devices connected on the bus (the GPIB controller counts as one device) and the total length of cable must not exceed 20 meters. 2. A maximum of 15 devices can be connected on one bus system. 3....
  • Page 21: Gpib Bus

    TH64 Series Operation Manual Way 2 6.1.2.2 GPIB bus Code Function Support all data source contact function Support all receptor contact function Basic Talker;Cancel in MLA;Serial call is not supported Basic Listener;Cancel in MTA;no Listen only function Remote/Local function Device clear function Device trigger function No controller function Open Electrode Drive...
  • Page 22 TH64 Series Operation Manual New Hardware will show on the right bottom of the computer desktop, as is shown below Click Next, the table below is displayed and select Install from a list or specific location Then click Next, select the routine of the driver provided by us then Next When the driver is installed, you can see USB Test and Measurement Device in the device manager.
  • Page 23: Usb Cdc Virtual Serial Port

    TH64 Series Operation Manual 6.1.4 USB CDC virtual serial port Select the bus mode to “USBCDC”, it can make USB interface configure as Vcom 6.1.4.1 System configuration Connect the USB interface on the rear panel of device to the USB interface in PC via USB cable 6.1.4.2 Install the driver Same as USB TMC Now USB V Com port can be considered as a serial port, when there is no serial port in PC, the...
  • Page 24: Command For Reference

    TH64 Series Operation Manual (5) Command can be abbreviation or can be full spelling E.g. FUNCTION:IMPEDANCE CPD = FUNC:IMP CPD (6) Question mark(?) is behind the command, execute an inquire to the command E.g. FUNC:IMP? 6.2.2 Command for reference 6.2.2.1 The following GPIB command is supported *RST used for reset the device *IDN inquire the model information 6.2.2.2 SCPI command...
  • Page 25 TH64 Series Operation Manual E.g. disp:page opd switch the page to <OUTPUT DISPLAY>. (2) DISPlay? Inquire the current page, and return the parameter, like “opt” ⚫ INSTrument channel selection command (1) INSTrument:NSELect <1|2|3> Select the channel to operate, which is the base of other commands (2) INSTrument:NSELect ? Inquire the current channel ⚫...
  • Page 26 TH64 Series Operation Manual The command is used to set over voltage protective value, and Min. value is 0, and the Max. value is determined by the model. ◼ VOLTage:PROTection ? Inquire the over voltage protective value of the selected channel. ◼...
  • Page 27 TH64 Series Operation Manual (3) TIMer: DATA<0~99999.9> The command is used to set the time of selected channel, Min. value is 0 and Max. value is 99999.9s (4) TIMer: DATA? Inquire the time value of selected channel ⚫ APPly Complex control command (1) APPLy: VOLTage <voltage value>,<...
  • Page 28 TH64 Series Operation Manual (4) MEASure: POWer? The command is used to inquire the measured power value (5) MEASure:VOLTage:ALL ? The command is used to inquire the voltage value of all channels (6) MEASure:CURRent:ALL ? The command is used to inquire the current value of all channels (7) MEASure:TIMer:ALL ? The command is used to inquire the timer value of all channels (8) MEASure: POWer:ALL ?
  • Page 29 TH64 Series Operation Manual English (2) MENu:LANG? The command is used to inquire the current language,, return cn or en. (3) MENu:VOICe <0|1|ON|OFF> The command is used to turn on or off the beeper (4) MENu:VOICe? The command is used to inquire the status of beeper (5) MENu:PMEM <DEFault|USEr>...
  • Page 30 TH64 Series Operation Manual hour, minute, second, and in second “< >”is the detailed value (2) DATe? The command is used to inquire the current date...
  • Page 31: Chapter 7 Specification

    TH64 Series Operation Manual Chapter 7 Specification TH6402 Parameter Voltage 0~30V 0~30V 0~6V Voltage limit 0~36V 0~36V 0~11V Rated output (0℃~40℃) Current 0~3A 0~3A 0~5A Power 210W ≤0.01%+3mV Voltage Load regulation ≤0.1%+3mA Current ≤0.01%+3mV Power supply Voltage ≤0.1%+3mA regulation Current Resolution of set Voltage value...
  • Page 32 TH64 Series Operation Manual TH6402A Parameter Voltage 0~30V 0~30V 0~5V Voltage limit 0~36V 0~36V 0~11V Rated output (0℃~40℃) Current 0~3A 0~3A 0~3A Power 195W ≤0.01%+3mV Voltage Load regulation ≤0.1%+3mA Current ≤0.01%+3mV Voltage Power supply ≤0.1%+3mA regulation Current Voltage 10mV Resolution of set value Current Voltage...
  • Page 33 TH64 Series Operation Manual TH6412 Parameter Voltage 0~30V 0~30V 0~6V Voltage limit 0~36V 0~36V 0~11V Rated output (0℃~40℃) Current 0~6A 0~6A 0~5A Power 390W ≤0.01%+3mV Voltage Load regulation ≤0.01%+3mA Current ≤0.01%+3mV Voltage Power supply ≤0.01%+3mA regulation Current Voltage Resolution of set value Current 0.1mA...
  • Page 34 TH64 Series Operation Manual Current 0~3A 0~3A 0~5A Power 390W ≤0.01%+3mV Voltage Load regulation ≤0.01%+3mA Current ≤0.01%+3mV Voltage Power supply ≤0.01%+3mA regulation Current Resolution of set Voltage value Current 0.1mA Voltage Resolution of reread value Current 0.1mA ≤0.03%+10mV Resolution of set Voltage ≤0.1%+5mA value...