Mustang
Modular Sample Control Panel
®
Installation, Operation & Maintenance
MMSCP
®

Advertisement

loading
Need help?

Need help?

Do you have a question about the MMSCP and is the answer not in the manual?

Questions and answers

Summary of Contents for Mustang Sampling MMSCP

  • Page 1 Mustang Modular Sample Control Panel ® Installation, Operation & Maintenance MMSCP ®...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Safety Warnings Product Description Technical Specifications Product Dimensions & Parts Flow Diagram Installation Instructions Operation Instructions Maintenance Instructions...
  • Page 3: Safety Warnings

    SAFETY WARNINGS Failure to abide by any of the safety warnings could result in serious injury or death. • Electrical power must be “OFF” before and during installation and maintenance or personal injury may result. • Do not exceed any equipment pressure ratings. •...
  • Page 4: Product Description

    PRODUCT DESCRIPTION The Mustang Modular Sample Control Panel (MMSCP ) uses Analytically Accurate ® ® technology to deliver a representative sample to an analyzer. The sample is controlled with a regulator and membrane filter to reduce pressure and remove entrained liquid and particulates.
  • Page 5: Technical Specifications

    TECHNICAL SPECIFICATIONS Maximum allowable working pressure 200 psig (14 bar) Wetted materials Machined parts: 316 stainless steel/NACE compliant All other metal parts: stainless steel/NACE compliant Seal Viton ® other materials available upon request...
  • Page 6: Product Dimensions & Parts

    PRODUCT DIMENSIONS & PARTS Item Description Number Membrane Separator Optional Single or Multi-Stage Regulator Particulate Filter Item Description Number From Cell Bypass To Cell Drain Sample In...
  • Page 7 Front View Side View...
  • Page 8: Flow Diagram

    FLOW DIAGRAMS...
  • Page 9: Installation Instructions

    INSTALLATION INSTRUCTIONS NOMENCLATURE • MAOP—Maximum Allowable Operating Pressure • BTU—British Thermal Unit • TCO—Thermal Cut-Off • Regulator adjustment screw - pressure set point adjuster TOOLS REQUIRED • Flat or Phillips Head Screwdriver. MOUNTING • Only use the two holes indicated below for mounting to a surface. DO NOT drill additional holes into the unit. TUBE FITTINGS •...
  • Page 10: Operation Instructions

    OPERATION INSTRUCTIONS START-UP 1. Open V1. 2. Open FIC-01 (Cell) and adjust the flow 1/2 to 3/4 of the meter unless it is dead-headed to the analyzer. In this case, turn fully counter clockwise to open the flow meter. 3. Open FIC-02 (By-Pass) and adjust the flow 1/2 to 3/4 of the meter. NOTE: •...
  • Page 11: Maintenance Instructions

    MAINTENANCE INSTRUCTIONS CHANGING THE PARTICULATE FILTER First, depressurize the system. With a wrench, Next, turn the Remove and replace the filter. Reassemble. Do loosen and turn metal disk counter- not overtighten the plastic container. the clear plastic clockwise and container counter- remove it and the clockwise and filter.