5. Measurement
5.4.3.
Setup of table conditions
Click
icon to display the setting screen of table condition and select the type of sample
and the unit. The data set using this function can be changed after the measurement.
Peak
Select HEIGHT (peak height) or AREA (peak area). Usually AREA is used. Should use
HIGHT if the peak is small and large variation is resulted when AREA is selected.
Calibration Curve
Select from 3 kinds calibration curves. Normal y=ax+b is selected. It is recommended to
select y=ax+Blank for the measurement of a sample in the vicinity of a blank.
Drift correction
Usually set to ON. If abnormal peak form is observed, the detector and heating condition
of a heater can be verified by setting the drift correction to OFF.
Set the other items such as Unit, according to the target to be measured.