a. To specify a Segment Type, click the Segment Type box in the
b. To specify a Beginning Stimulus or an Ending Stimulus, click the
c. To specify a Beginning Response or an Ending Response, click the
d. Use the tab key to advance from left to right across a row. Use
Figure 5-10
Empty Limit Table
Table 5-4
Limit Table Entries for Insertion Loss
Segment
Number
Figure 5-11
Completed Limit Table for Insertion Loss
Agilent Technologies Z5623A Option H48
desired row, then use the drop-down menu.
respective box in the desired row, then enter the desired value
followed by "k" if the value is in kilohertz, or by "g" if the value is in
gigahertz. For example, to specify a Beginning Stimulus of 300 kHz,
type, "300k" (without the quotation marks). To specify an Ending
Stimulus of 1.3 GHz, type, "1.3g" (without the quotation marks).
respective box in the desired row, then enter the desired value in dB,
then hit the space bar. For example, to specify a Beginning Response
of −1.8 dB, type, "−1.8<space>".
<shift><tab> to advance from right to left across a row. (When
moving from right to left, the cursor will not move until you release
the shift key.)
Limit
Type
1
Min
2
Min
3
Min
4
Min
Beginning
Ending
Stimulus
Stimulus
300 KHz
1.3 GHz
1.3 GHz
3 GHz
3 GHz
6 GHz
6 GHz
9 GHz
Performance Verification
Setting Up Limit Testing
Beginning
Ending
Response
Response
−1.8 dB
−1.8 dB
−2.3 dB
−2.3 dB
−2.8 dB
−2.8 dB
−3.8 dB
−3.8 dB