User Manual
Simpson 4005/ 4006 Series
Analog-Digital Multimeter

Advertisement

Table of Contents
loading

Summary of Contents for Simpson 4005 Series

  • Page 1 ® USA SINCE 1936 RELIABILITY BEYOND MEASURE User Manual Simpson 4005/ 4006 Series Analog-Digital Multimeter...
  • Page 2 AUTO 230.0 50.00 (11) (10) (21) (20) (19) (18) (15) (17) (12) (16) (13) (15) (14) (11)
  • Page 3 (1) Liquid crystal display (2) PMAX / PMIN push button (3) Push button for MIN/MAX functions (4) Pushbutton for manual range selection (5) Multi function push button (6) Function selector switch. (7) Terminal sockets with automatic blocking system. (8) Push button for HOLD function (9) Push button for Back light function (10) Main Display for digits, decimal point and polarity (11) Display for manual range selection, Hold, MIN/MAX, Relative,...
  • Page 4: Table Of Contents

    Contents Page 1. Introduction................4 2. Safety features and safety precautions........4 3. Switching the multimeter "ON"...........5 4. Function and range selection.............6 4.1 Switching the measuring ranges.........6 4.2 Autoranging ................6 5. Liquid crystal display..............7 5.1 Digital display ..............7 5.2 Analog indication ..............8 5.3 Backlit .................8 6.
  • Page 5: Introduction

    1. Introduction Thank you very much for selecting our multimeter. These multimeters are manufactured as per IS 13875 and DIN 43751. 2. Safety features and safety precautions you have chosen a multimeter which provides you a very high degree of safety.
  • Page 6: Switching The Multimeter "On

    Meaning of the symbols on the device Warning of a danger point (Attention, refer to the user manual) Earth (ground) terminal. Double or reinforced insulation Instrument for over voltage CAT II / III / IV category II / III or IV EU conformity mark.
  • Page 7: Function And Range Selection

    Note: Electric discharges and high-frequency influence may cause incorrect information to be displayed and block the measuring process. Reset the meter by switching it OFF and ON again otherwise, check the battery connections. Disconnect the meter from the measuring circuit before you open it, and see section “17.Maintenance”) Automatic Meter-OFF (MoFF) DMM has a default auto meter off function.
  • Page 8: Liquid Crystal Display

    switched OFF when push button AUTO/MAN is pressed (4) for approximately 1s,or when the function selector switch (6) is operated, or when the meter is turned OFF and ON again. Acknowl- AUTO edgement Display Switching sequence at : - 6.6V 8 66V VAC / 660V 1000V...
  • Page 9: Analog Indication

    5.2. Analog indication The analog indication with bar graph is updated 28 times per second. Analog indication is of particular use when observing variations of measured values. The analog bar graph (13) has its own polarity indication in measuring DC quantities, when the positive pole of the measured quantity is applied to the “...
  • Page 10: Peak Measurement

    Apply the measured quantity to the meter and select the measuring range prior to activating the MIN/MAX function. With the function activated, you can select the measuring ranges only manually, if you switch to another range, the stored MIN/MAX values are cleared. After pressing MIN/MAX(3) for the first time, sub-display shows maximum value.
  • Page 11: Current Measurement

    Voltage measurement 50.00 50.00 Black Probe Red Probe +(-)/~ -(+)/~ 10. Current Measurement First disconnect the power supply to the circuit being measured and/ or to the load, and discharge all capacitors within that circuit. With the function selector switch (6), select A for currents >660mA and mA for currents <=660 mA.
  • Page 12: Ac Current Measurement With (Clip-On) Current Transformers

    sections and connection points should be designed to avoid excessive heating. The current measuring ranges up to 660 mA are protected to a short circuit current of 25 A by a fuse 1.6 A/1000V AC/DC in conjunction, with power diodes. The cut-out capacity of the fuse is 10kA at a rated voltage of1000V AC/DC and ohmic load.
  • Page 13 Current measurement ...660 mA ...10 A 9.00 660.0 50.00 50.00 Black Probe Black Probe Red Probe Red Probe -(+)/~ -(+)/~ +(-)/~ +(-)/~ AC current measurement with(clip-on)current transformers 660.0 50.00...
  • Page 14: Resistance Measurement

    11. Resistance measurement Verify that the device under test is electrically dead. External voltages would falsify the measured result! Set the function selector switch (6) to “ ” Connect the device under test as shown. Zero adjustment on the 660 measuring range When measuring small resistance values on the 660 range, you can eliminate the resistance of the leads and contact resistance by REL function.
  • Page 15 As long as the voltage drop does not exceed the maximum display value of 1.999V, you can also test several series-connected elements or reference diodes with small reference voltage. Reverse direction or open circuit: The multimeter indicates overrange “OL”. With the diode function selected, the meter emits a continuous sound signal whenever the reading is less than 30mV.
  • Page 16: Capacitance Measurement (Model 4005, 4006)

    Continuity Test 00.00 Red Probe Black Probe Black Probe Red Probe DiodeTest Forward direction Reverse direction 0.600 Black Probe Black Probe Red Probe Red Probe 13. Capacitance measurement (Multimeter 4005,4006 ) Verify that the device under test is electrically dead. External voltages would falsify the measured results! Set the function selector switch (6) to “...
  • Page 17: Frequency And Duty Measurement (Model 4005, 4006)

    Notes Connect polarized capacitors with the “ ” pole to the “ ” socket. Resistors and semiconductor junctions in parallel with the capacitor falsify the measured results! Zero adjustment on the 6.600 nF measuring range When measuring small capacitance values on the 6.600 nF range, the internal resistance of the multimeter and the capacitance of the leads can be eliminated by “REL”...
  • Page 18: Temperature Measurement

    Notes The applied frequency must remain constant during the duty cycle measurement. 15.Temperature measurement The Multimeter 4005..4006 allows you to measure temperature with K type thermocouple in the range from 0 C...1300 C, 32°F...2372°F. Set the function selector switch (6) to “ C ” Connect the multimeter probe to the two unblocked terminals and thermocouple output.
  • Page 19: Specifications

    16. Specification Resolution Meas. Measuring 4006 4005 Range Function (TRMS) • • µ 100 V 660.0mV • • 6.600V • • 10mV V(DC) 66.00V • • 660.0V 100mV • • 1000V • • µ 100 V 660.0mV • • 6.600V •...
  • Page 20 Digital display Overload capacity Inherent deviation at Input reference condition Overload Impedance Overload +(...%rdg + ...digits) Values Duration >10 G // <40pF 0.7 + 5 11 M // <40pF 0.4 + 5 10 M // <40pF 0.4 + 5 1000 V 10 M // <40pF 0.4 + 5 10 M // <40pF...
  • Page 21 Resolution Meas. Measuring 4006 4005 Function Range (TRMS) • • 66.00K • • 660.0K • • 6.600M • • 66.00M • • 100m BUZZER 660.0 • • 2.000V DIODE • • 6.600nF • • 10pF 66.00nF • • 100pF 660.0nF •...
  • Page 22 Digital display Overload capacity Inherent deviation at Input reference condition Overload Overload Impedance +(...%rdg + ...digits) Values Duration -1.08V 0.8 + 5 -1.08V 0.8 + 5 -1.08V 1.0 + 5 -1.08V 2.0 + 5 -3.3V 0.8 + 5 3.3V 2.0 + 10 3.0+40 1000 V 2.0+10...
  • Page 23 Influence Quantities and Variations Measured Variation Quantity/ Influence Range of + (...% of rdg. Measuring quantity Influence + ... digits) Range 1 X Intrinsic 0 ºC Temperature error +21 ºC Diode +25 ºC…+40ºC ºC 1.0+3 20 Hz..< 50 Hz 660mV~ 5.0+3 >...
  • Page 24 Influence Measuring Range of Influence Attenuation quantity ranges > 100 dB Noise quantity max. 1000 V Common >100 dB mode interference Noise quantity max. 1000 V ~ voltage 50 Hz, 60 Hz sinusoidal > 60 dB Noise quantity V~ 660mV, 6.6V, >...
  • Page 25 Power supply Battery : 2 AA size batteries alkaline manganese cells as per IEC LR6. Service life : for 4005 : 600 hrs. for VDC, ADC 300 hrs. for VAC, AAC for 4006 : 400 hrs. for VDC, ADC 200 hrs. for VAC, AAC Battery test : Automatic display of “...
  • Page 26: Maintenance

    Reference conditions ± Ambient temperature 23 C/ 73.4 F Relative humidity 45% ... 55 % RH Frequency of measured quantity 45Hz ...65 Hz Waveform of the measured quantity sinusoidal Battery voltage 3V + 0.1V Environmental conditions 0°C...50°C, -13°F...158°F Functional temperature range Storage temperature range -25°C...+70°C (without batteries), -13°F...158°F...
  • Page 27: Fuses

    Special maintenance of the case is not required .Take care that the surface between the connection sockets is clean. For cleaning take a moist cloth. Avoid scrabbing. Simpson Electric Company 520 Simpson Avenue Lac Du Flambeau, WI 54538-0099 (USA) Web: www.simpsonelectric.com...