WATSON PCR Instruction Manual

WATSON PCR Instruction Manual

4 well portable thermocycler

Advertisement

Quick Links

Advertisement

Table of Contents
loading

Summary of Contents for WATSON PCR

  • Page 2 2 ± 0.5 degrees/sec cooling ramp time. The heated lid prevents water condensation during sample heating step. Power is delivered to WATSON by a USB type C port. Either USB AC wall adapter or USB power bank can power the device. The 10Ah power bank that comes with the device will enable the user to run 4 full PCR assays without charging.
  • Page 3: Safety And Regulatory Compliance

    Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements.  Safety Requirements for Applicable clauses of IEC 61010-2-020: 2003 Electrical Equipment for Measurement, Control, and Laboratory Use, Part 2: Particular Requirements for Laboratory Equipment for the Heating of Materials. P a g e WATSON Manual Ver 1.0...
  • Page 4 Plug the USB power bank or the USB-C AC power adaptor into the USB type C port on WATSON. Both green and the blue lights should turn ON which place the machine into the STANDBY mode for accepting samples. The blue LED is for power and the green LED is for the STANDBY mode.
  • Page 5: Editing Programs

    The red LED will start flashing to indicate that the machine is running. ◦ When WATSON is running DO NOT OPEN the lid as it will be hot and would interfere with the run. It is also a burn hazard.
  • Page 6 • Choose which program to alter by changing the Program Number or by altering the dip switch patterns on WATSON or on the dip switch icon on the main program interface. • Alter temperature and duration of each step by selecting the desired box and changing the numbers.
  • Page 7 Selecting READ (for current settings) will display what program is currently stored for a particular dip switch patten in WATSON. ◦ All programs that are stored in WATSON can be accessed by selecting the program number of a dip switch pattern and pressing the READ button. ◦...
  • Page 8 The top right corner of the interface shows the lid and plate temperature. • The top left corner shows the estimated run time for the PCR program. • The red and blue circles on the graph tracks lid and heating block parameters during the PCR run.
  • Page 9: Led Message

    Dip Switch Settings Figure 4: Dip Switch with all switches down indicating it is in Program 0. The WATSON has the ability to store 16 different PCR programs. The different programs are represented by dip switch positioning. Each unique position of the switches represents a different program that has been uploaded on WATSON.
  • Page 10 Down: 1,2,3 Up: 1,4 Down: 2,3 Up: 2,4 Down: 1,3 Up: 1,2,4 Down: 3 Up: 3,4 Down: 1,2 Up: 1,3,4 Down: 2 Up: 2,3,4 Down: 1 Up: All Down: None 10 | P a g e WATSON Manual Ver 1.0...
  • Page 11: Maintenance

    Maintenance Prior to doing any maintenance, please unplug power supply from WATSON. Cleaning is recommended before every run. Wear gloves and apply 1% Bleach to a microfiber cloth and gently wipe the surface to destroy any DNA left on the machine.
  • Page 12: Product Specification

    Pollution Degree Degree of Protection IP30 Reaction Volumes 10-50 µl Runtime 45-60 min for 3-step s 30-40 cycles PCR Sample Format 4 wells plate; PCR tubes 200 ul capacity Thermocycling System Resistive Heating, Forced-Air Cooling Programmable Plate Temperature range 30-105 °C Programmable Lid Temperature range 30-105 °C...
  • Page 13: Contact Information

    CONSEQUENTIAL DAMAGES OF ANY NATURE WHATSOVER OR EXPENSES ARISING OUT OF OR CONNECTED WITH THE USE OR MISUSE OF THIS PRODUCT. Manufactured by Microbiologique, Inc 8315 Lake City Way NE Seattle, WA 98115 13 | P a g e WATSON Manual Ver 1.0...