Static Feeder Calibration – OP12
Description
The technical instruction describes the step by step process to conduct the
static balance of the scale of the OP12.
Status:
01.06.2022
Static Feeder Calibration – OP12

Advertisement

loading

Summary of Contents for Brabender OP12

  • Page 1 ECHNICAL NSTRUCTION Static Feeder Calibration – OP12 Description The technical instruction describes the step by step process to conduct the static balance of the scale of the OP12. Status: 01.06.2022 Static Feeder Calibration – OP12...
  • Page 2 Step 1 Press the button on the bottom right Step 2 Press on the Key button Step 3 Type in the code 08434 and press the Enter button Technical Instruction - Static Feeder Calibration – OP12...
  • Page 3 Step 4 Choose the scale whose leverage factor needs to be changed Step 5 Press the second button from the bottom right Step 6 Press the second button from the top right Technical Instruction - Static Feeder Calibration – OP12...
  • Page 4 Step 7 Press the button to enter the weight of the test weight, (Example: 10kg) Step 8 Put the test weigth on top of the feeder Remove the test weight Step 9 Technical Instruction - Static Feeder Calibration – OP12...
  • Page 5 Step 10 Press button to send the new lever arm into the system. The lever arm is now set! Control again with the test weight. Technical Instruction - Static Feeder Calibration – OP12...