Gastops ChipCHECK CC120 Maintenance And Service Manual

Chips analyzer
Hide thumbs Also See for ChipCHECK CC120:

Advertisement

Quick Links

CC120 Maintenance and Service
CHIPS ANALYZER (B166500)
CAUTION – USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER
THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE
THE INFORMATION, TECHNICAL DATA AND DESIGNS DISCLOSED HEREIN ARE THE EXCLUSIVE PROPERTY OF GASTOPS LTD. AND ARE NOT TO BE
USED OR DISCLOSED TO ANY OTHERS WITHOUT THE WRITTEN CONSENT OF Gastops LTD. ANY RECIPIENT, BY ITS RETENTION AND USE, AGREES
TO HOLD IN CONFIDENCE ANY INFORMATION, TECHNICAL DATA AND DESIGNS DISCLOSED HEREIN. THE FOREGOING SHALL NOT APPLY TO
PERSONS HAVING PROPRIETARY RIGHTS TO SUCH INFORMATION, TECHNICAL DATA OR DESIGNS TO THE EXTENT THAT SUCH RIGHTS EXIST.

Advertisement

loading
Need help?

Need help?

Do you have a question about the ChipCHECK CC120 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Gastops ChipCHECK CC120

  • Page 1 THE INFORMATION, TECHNICAL DATA AND DESIGNS DISCLOSED HEREIN ARE THE EXCLUSIVE PROPERTY OF GASTOPS LTD. AND ARE NOT TO BE USED OR DISCLOSED TO ANY OTHERS WITHOUT THE WRITTEN CONSENT OF Gastops LTD. ANY RECIPIENT, BY ITS RETENTION AND USE, AGREES TO HOLD IN CONFIDENCE ANY INFORMATION, TECHNICAL DATA AND DESIGNS DISCLOSED HEREIN.
  • Page 2: Maintenance And Service

    Finally, if the fuse blows, then the User is permitted to replace the fuse with the proper part. To ensure the integrity of the unit, it is recommended that this fuse be sourced directly from Gastops. Note that, to ensure optimum performance and longevity of the equipment the Gastops in-service support package must be maintained throughout the life of the equipment.
  • Page 3 Gastops Ltd. Figure 6-1: Containment Window Location 6.1.1 Replacement of the Containment Window If the Containment Window is to be completely replaced, then the following procedure must be followed. Note that gloves are required to complete this procedure and the User must refer to Figures 6-1, 6-2 and 6-3 to supplement the procedure below.
  • Page 4 Gastops Ltd. Carefully place the metallic cover back over the seated Containment Window – ensure that • the cover is fully seated (note that the Containment Window metallic cover can only be assembled in one orientation). Lastly, return to the software interface and press the ‘Maintenance Complete’ button.
  • Page 5: Replacing The Fuse

    In the event that a fuse requires replacement, then the following procedure must be followed. WARNING: Disconnect supply before servicing (note that service/servicing can only be completed by Gastops authorized technicians). Disconnect supply before replacing fuse. C009423, Rev. 4 page 45...
  • Page 6 There is absolutely NO SERVICE ALLOWED by the User. Service activities are restricted to only Gastops authorized technicians who have the appropriate laser safety training. There are no User serviceable parts. This unit is to be serviced or repaired only by Gastops authorized technicians.
  • Page 7: Basic Troubleshooting

    Class 4 laser radiation. Such panels include the unit’s side panels, back panel, and main aluminum wrap. Tamper proof fasteners have been used for your protection. Only Gastops authorized technicians may remove these fasteners. Users are strictly forbidden from removing any fastener external to the unit.
  • Page 8 Then power up and restart the unit as Software interface does not respond per Section 3.3. to user input If this does not work, then contact Gastops • Customer Support (See Section 8.0). Conduct a complete unit shutdown and power •...
  • Page 9 Then power up and restart the unit as Software interface does not respond per Section 3.3. to user input If this does not work, then contact Gastops • Customer Support (See Section 8.0). Conduct a complete unit shutdown and power •...
  • Page 10 Gastops Ltd. 7.0 Transport/Shipping and Storage Transport/Shipping The ChipCHECK unit comes packaged in MIL-STD-810 compliant case for field deployability (see Figure 7-1). Within the transport case, a removable foam insert protects the front face of the ChipCHECK unit. When unboxing the unit, this insert must be removed before the ChipCHECK can be removed from the transport case.
  • Page 11 Gastops Ltd. The unit itself weighs approximately 40 lbs (18 kg). The Transport Case that was provided with the instrument must be used any time the instrument is moved positions or locations. Where possible, two people shall transport the unit and the instrument shall be kept in its vertical orientation at all times.
  • Page 12 Gastops Ltd. Storage slots beneath foam insert for Sample Cases Space for Operator’s Manual located beneath ChipCHECK unit Figure 7-3: Transport Case Storage Locations (Foam insert removed) External Keyboard (optional) External USB Wash Bottle (optional) Sample Preparation Tools Hub (optional) (optional) Calibration &...
  • Page 13 Gastops Ltd. Included in the Transport Case: Quantity Description ChipCHECK Unit Unit Power Cable Sample Tray (Chip) 200x Sample Patches Standardization Standard Calibration Standard Sample Patch Cases USB Ethernet Adapter Additional optional accessories may include the following: External keyboard •...
  • Page 14: Customer Support

    Gastops Ltd. 8.0 Customer Support Gastops technical support representatives for ChipCHECK can be reached at: Phone: 1-613-744-3530 Fax: 1-613-744-8846 Toll Free: 1-800-363-8658 Email: support@gastops.com Gastops Limited 1011 Polytek Street Ottawa, ON K1J 9J3 Canada C009423, Rev. 4 page 54 All use of this publication and ChipCHECK by user shall be subject to the restrictions and conditions as set out in this publication.