  • Page 1 NI-9229 Getting Started 2022-07-06...
  • Page 2: Table Of Contents

    Safety Guidelines............3 NI-9229 with Screw Terminal Safety Voltages......3 NI-9229 with BNC Safety Voltages.
  • Page 3: Overview

    Note The guidelines in this document are specific to the NI-9229. The other components in the system might not meet the same safety ratings. Refer to the documentation for each component in the system to determine the safety and EMC ratings for the entire system.
  • Page 4: Ni-9229 With Bnc Safety Voltages

    Explosive atmospheres Channel-to-channel 60 V DC, Measurement Category I Channel-to-earth ground 60 V DC, Measurement Category I NI-9229 with BNC Safety Voltages Connect only voltages that are within the following limits: Isolation Channel-to-channel Continuous 60 V DC, Measurement Category I...
  • Page 5: Safety Guidelines For Hazardous Locations

    Safety Guidelines for Hazardous Locations The NI-9229 is suitable for use in Class I, Division 2, Groups A, B, C, D, T4 hazardous locations; Class I, Zone 2, AEx nA IIC T4 Gc and Ex nA IIC T4 Gc hazardous locations;...
  • Page 6 II 3G and is suitable for use in Zone 2 hazardous locations, in ambient temperatures of -40 °C ≤ Ta ≤ 70 °C. If you are using the NI-9229 in Gas Group IIC hazardous locations, you must use the device in an NI chassis that has been evaluated as Ex nC IIC T4, Ex IIC T4, Ex nA IIC T4, or Ex nL IIC T4 equipment.
  • Page 7: Electromagnetic Compatibility Guidelines

    NI-9229 Getting Started Caution Transient protection shall be provided that is set at a level not exceeding 140% of the peak rated voltage value of 85 V at the supply terminals to the equipment. Caution The system shall only be used in an area of not more than Pollution Degree 2, as defined in IEC/EN 60664-1.
  • Page 8: Special Conditions For Marine Applications

    NI-9229 Getting Started Cable Requirements for EMC Compliance Select and install cables for the NI-9229 with BNC in accordance with the following requirements: Install a clamp-on ferrite bead (NI part number 782801-01) on the BNC cable ■ for each channel that you are connecting to on the NI-9229 with BNC.
  • Page 9: Ni 9229 Pinout

    NI-9229 Getting Started Operating humidity (IEC 60068-2-78) 10% RH to 90% RH, noncondensing Pollution Degree Maximum altitude 2,000 m Indoor use only. Note Refer to the device datasheet on ni.com/manuals for complete specifications. NI 9229 Pinout AI0+ AI0+ AI0– AI0-...
  • Page 10: Floating Connections

    You must use 2-wire ferrules to create a secure connection when connecting ■ more than one wire to a single terminal on the NI-9229 with screw terminal. Wiring for High-Vibration Applications If your application is subject to high vibration, NI recommends that you follow these guidelines to protect connections to the NI-9229 with screw terminal: Use ferrules to terminate wires to the detachable connector.
  • Page 11: Where To Go Next

    NI-9229 Getting Started Where to Go Next CompactRIO NI CompactDAQ NI 9229 Datasheet NI 9229 Datasheet NI-RIO Help NI-DAQmx Help LabVIEW FPGA Help LabVIEW Help RELATED INFORMATION C Series Documentation Services & Resources ni.com/services ni.com/info cseriesdoc Located at ni.com/manuals Installs with the software...