Medi CAD babyDOPPY User Manual

Ultrasonic fetal heartbeat detector

Advertisement

Quick Links

babyDOPPY
Ultrasonic fetal heartbeat detector
USER MANUAL
1011
v.: 1.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the babyDOPPY and is the answer not in the manual?

Questions and answers

Summary of Contents for Medi CAD babyDOPPY

  • Page 1 Ultrasonic fetal heartbeat detector USER MANUAL 1011 v.: 1.
  • Page 2 Ultrasonic fetal heartbeat detector Medi-CAD Kft. 3529 Miskolc, Lázár Vilmos u. Tel: +36 (46) 369-377, 431-395 Fax: +36 (46) 560-423 http://www.medicad.hu/ kiadás: 1...
  • Page 3: Table Of Contents

    Contents Contents 1.Fields of application...........2 2.Main parts of the device........3 3.Measuring fetal heart rate........4 4.Checking and charging the battery.....5 6.Cleaning and disinfection........7 7.Environmental and other conditions of the use, storage and transportation of the device..8 8.Technical specifications........10 9.Contents of the kit...........12 10.
  • Page 4: Fields Of Application

    1. Fields of application babyDOPPY is a portable, battery-powered, ultrasonic diagnostic equipment. Using its 2 MHz sensor, human fetal heart activity can be detected from the second trimester of pregnancy. From then, the fetal heart rate can be monitored to the end of pregnancy.
  • Page 5: Main Parts Of The Device

    2. Main parts of the device  Sensor  Push button (On/Off switch)  wheel (VOLUME) Volume control  (PHONE) Headphone socket  Sensor/ (PROBE) battery charger connector  LCD display  Charger adapter       ...
  • Page 6: Measuring Fetal Heart Rate

    3. Measuring fetal heart rate   Connect the sensor to the PROBE socket. Switch on the device by briefly pressing the push button on the handle of the sensor. After the unit has been turned on, three number 8’s appears on the LCD display of the device, followed by three “-“...
  • Page 7: Checking And Charging The Battery

    DO NOT use salt-containing ECG conductive gels, as they may harm the sensor! 4. Checking and charging the battery babyDOPPY is operated by a 9V battery, independently of the electric network. The fully charged battery lasts for a total of up to 4 hours of operation in case of normal signals and use at an average volume level.
  • Page 8 most probable that one battery-charge will operate the device for several workdays even if the device is in constant use. The beginning of low-battery status is indicated by the brief flash of “Lob” on the LCD display when the device is turned on or being operated during an examination.
  • Page 9: Cleaning And Disinfection

    6. Cleaning and disinfection Use a soft, damp cloth to clean the main unit and the sensors. Be sure to prevent water or other liquids to seep in through the gaps or crevices (e. g. speaker grid, socket, switch) of the main unit and the sensor. Never submerge the sensor or the main unit in any liquid (e.
  • Page 10: Environmental And Other Conditions Of The Use, Storage And Transportation Of The Device

    7. Environmental and other conditions of the use, storage and transportation of the device Temperature data: +5 °C - +40 °C Operating temperature: -40 °C - +65 °C Transportation temperature: -20 °C - +55 °C Storage temperature: Humidity data: Operating humidity: 30% - 95 % Transportation and storage humidity: max.
  • Page 11 At the end of its life-cycle, the device must be disposed of at the manufacturer or at collection sites specialized for the recycling and disposal of electronic equipment.
  • Page 12: Technical Specifications

    8. Technical specifications Sensor: Frequency: Type Nominal Max. ultrasound output power: [MHz] diameter [mm] cm² 2±10% DO-02 Method of heart rate (FHR) calculation: Calculated based on the average of 8 heartbeats : 60/min. - 210/min. Range Classification: Device operated by internal power source BF type patient-side part Power supply: 9V rechargeable battery (PP3)
  • Page 13 Operating period: approx. 4 hours (total, with a fully charged battery) Dimensions: main unit: 117 x 78 x 25 mm sensor: Ø16/Ø 20 x 105 mm Weight: main unit: approx. 150 g sensors: approx. 120 – 140 g/each charger: approx. 70 g The manufacturer reserves the right to change the specifications!
  • Page 14: Contents Of The Kit

    9. Contents of the kit babyDOPY unit 1 pc 1 pc Sensor Conductive gel 1 tube Charger adapter 1 pc Case 1 pc User’s manual 1 pc...
  • Page 15: Repair And Maintenance

    10. Repair and maintenance The repair and maintenance of the device may only be performed by Medi-CAD Kft. or its authorized technicians. Address: Medi-CAD Kft. 3529 Miskolc, Lázár Vilmos u. 4. Tel.: +36 (46) 369-377, 431-395 Email: medicad@medicad.hu...
  • Page 16: Declaration Of Conformity

    Hungary Tel: +36 46 369 377 Fax: +36 46 560 423 Product: Ultrasonic fetal heartbeat detector Type: DOPPY product family: babyDOPPY Classification: Class II a. Notified body: GYEMSZI Equipment Evaluation and Hospital Technology Directorate 1125 Budapest Diós árok u. 3.
  • Page 17: Quality Certificate / Warranty

    Quality Certificate / Warranty Name of Product: babyDOPPY Serial Nos.: • Main unit: • DO-02: Based on quality inspection performed on the finished product, the device meets the performance requirements specified in the User’s Manual. Miskolc, 201 _________________________ This warranty is valid for the above product for 24 months from purchase.