Download Print this page

Advertisement

Quick Links

07
Quick Guide Manual
Blood Glucose
Monitoring System

Advertisement

loading

Summary of Contents for OKRA 07

  • Page 1 Quick Guide Manual Blood Glucose Monitoring System...
  • Page 3 Quick Guide Manual Blood Glucose Monitoring System...
  • Page 4 OKRA...
  • Page 5 Using your lancing device Twist the cap to the displayed arrow direction. Gently pull the cap to remove. Insert the lancet into the lancet holder, push it all the way in. OKRA...
  • Page 6 Twist and remove the protective cap from the lancet to expose the needle. Place the cap back on your lancing device. Twist the cap to the direction as shown in arrow. OKRA...
  • Page 7 Select the depth of the lancet by dialing the lancing device. Setting of the lancing depth begin on one, for the shallowest setting. Increase the dial if more depth is needed. Pull the handle out until it clicks. OKRA...
  • Page 8 Discard 6 months after opening. Do not use after expiration date. Do not use after expiration date. Level 2. Level 2. Each OKRA test strip contains Each OKRA test strip contains Glucose Dehydrogenase (Aspergillus sp.): 7.2% Glucose Dehydrogenase (Aspergillus sp.): 7.2% Level 3.
  • Page 9 Level 1. Do not Refrigerate. Do not Freeze. Discard 6 months after opening. Do not use after expiration date. Level 2. Each OKRA test strip contains Glucose Dehydrogenase (Aspergillus sp.): 7.2% Level 3. Mediator (Hexaammineruthenium Chloride): 42.8% Binder: 2.2% Stabilizer: 47.8% Made in Korea Rev.2020-10-01...
  • Page 10 *Do not apply blood on top of the test strip *Caution: Inserting blood at the end of the test strip Do not put blood on TOP of the test strip OKRA...
  • Page 11 The test result appears on the display after 5 seconds Dispose the used strip by pushing the ejector OKRA...
  • Page 12 Model : IGM-1003 Rating : 3V ,4mA Made in Korea IMC01932 INFM001LF Rev.2020-10-01 Tel : (833-977-1339) www.okra.care the door and pull the door up and open Release the old battery and replace with a new Blood Glucose Monitoring System Model : IGM-1003 Rating : 3V...
  • Page 13: Troubleshooting

    Trouble-Shooting Blood Glucose Monitoring System OKRA...
  • Page 14 To solve this issue, please take out a new test strip from the container and insert into the meter. When blood insert icon appears, please apply blood sample at the end of the test strip OKRA...
  • Page 15 When you see Er3 on the meter display, please check whether enough blood has been inserted into the test strip. To solve this issue, make sure to insert the blood at the end of the test strip, not on top of the test strip OKRA...
  • Page 16 To solve this issue, please take out a new test strip from the container and insert into the meter. Make sure the blood insert icon has appeared before applying the blood sample. OKRA...
  • Page 17 Memo OKRA...
  • Page 18 Memo OKRA...
  • Page 20 DISTRIBUTED BY CGETC Inc. CITY OF INDUSTRY, CALIFORNIA 91748 Quality Guarantee www.okra.care Customer service is available Monday thru Friday, 9 to 5pm Eastern Time. TEL: (833-977-1339) E-MAIL: support@okra.care INFM001LF IMC01933 Rev. 2021-07-01...