Plymouth Nitronox Plus Maintenance

Advertisement

Nitronox Plus
®
Maintenance/Cleaning Methods
Maintenance
Maintenance
Field
adjustments and replacement parts obtained through a Porter authorized distributor and
as outlined in this user manual. To assure a long useful product life, perform the
periodic field performance tests and adjustments, including the Field Leak Test, Alarm
Function Test, Failsafe Test and Emergency Air Intake Valve Test (as outlined in the
Porter user manual pp. 15-17).
Porter advises (not required) an inspection by an authorized Porter representative every 2
years. Check local and state guidelines (if any) for certification regulations and requirements.
It is advised that the Nitronox Plus
out and calibration after 10 years of product operation in the field. The Nitronox Plus
device can be serviced and repaired for 20 years. Contact your Porter distributor for
pricing information and more details.

Cleaning Methods

We recommend the use of an approved disinfectant for the healthcare patient environment
for cleaning the outside of the Nitronox Plus
disinfectant directly onto housing. Spray disinfectant into disposable towel, or use
disinfectant
surface
thoroughly removing excess disinfectant to eliminate buildup. Cleaning of the device
has been validated with Super Sani-Cloth Germicidal wipes. Follow the disinfectant
manufacturer's directions for use (p. 18).
Introduction of moisture or other contaminants into this device may result in
defective operation. Never oil or grease any part of this system (minimize fire or explosion
potential).
WARNING: The breathing circuit, mask, and mouthpiece accessories are single use only. Do not
attempt to clean, sterilize, or reuse these items. Always use a new breathing circuit and mask/
mouthpiece for each procedure.
WARNING: Do not wipe the inside of the inlet fittings, gas supply hoses or connection port as
this may deposit cleaning agents into the breathing pathway of the device.
WARNING: Do not immerse the device or any of the device's components or accessories in
water, cleaning solution, or any other liquid. Always ensure the device and the device's
components and accessories are completely dry before use.
WEB – plymouthmedical.com
COPYRIGHT © 2020 PLYMOUTH MEDICAL LLC. ALL RIGHTS RESERVED.
repair
of
the
Nitronox Plus
device have a complete factory elastomer change
®
wipe
(avoid
excess
TEL – 888.392.5076
PLYMOUTH MEDICAL
FAX – 888.392.1455
135 PLYMOUTH ST., SUITE 410
BROOKLYN, NY, 11201
is
limited
®
device and accessories. Do not spray
®
disinfectant
liquid),
to
maintenance
®
and
wipe
unit
V2-12-21

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Nitronox Plus and is the answer not in the manual?

Questions and answers

Summary of Contents for Plymouth Nitronox Plus

  • Page 1 It is advised that the Nitronox Plus device have a complete factory elastomer change ® out and calibration after 10 years of product operation in the field. The Nitronox Plus ® device can be serviced and repaired for 20 years. Contact your Porter distributor for pricing information and more details.
  • Page 2 5. Perform the setup and pre-use checks as specified in Section 3.1 (p. 12) Servicing the Demand Valve The demand valve utilized in the Nitronox Plus® requires periodic service (every 5 years). For information on servicing the demand valve, contact your authorized distributor.

Table of Contents