Ventilation holes on the product housing must be kept clear to avoid malfunctions due to high temperatures, abnormal operation, shortened life and fire. Please strictly follow the instructions in this manual, otherwise the product's protective device may be damaged. Page 4 www.ceyeko.com...
Page 5
When not using this product, be sure to turn off the power switch. If you find smoke from the product, smell odor, abnormal noise, etc., please turn off the power switch immediately and contact the company in time. Page 5 www.ceyeko.com...
Controller Manual Certificate After opening the box, please confirm the package contents before starting the operation.If you find that the model and quantity are incorrect or there is physical damage in appearan- ce, please contact us. Page 6 www.ceyeko.com...
Can be pH / ORP measurement, temperature measurement, upper and lower limit control, transmission output, RS485 communication. Configurable manual and auto temperature offset function. Configurable upper/lower limit warning and delay. Configurable hummer and LCD backlight switch. Page 8 www.ceyeko.com...
With ambient temperature over 60 degrees in operation. With humidity over 85% in operation. Nearby electromagnetic source. In strong mechanical vibration. With varying temperature and dew condensation. With oil smoke, steam, humidity, dust and corrosive gases. Page 10 www.ceyeko.com...
2 . 1 . 3 I n s t a l l a t i o n M e t h o d s Install a 92.5*92.5 mm mounting hole on the instrument cabinet or mounting panel. The thickness of the mounting panel is 1.5mm~13mm. Page 11 www.ceyeko.com...
❷ ❺ Pipe installation Flow-through installation ❸ ❻ The interface must be in 15°oblique angle, or it will affect the normal test and use of the electrode. We won’t be responsible for any results due to this. Page 12 www.ceyeko.com...
Under “Monitoring Page” - Enter the MENU MENU Under “Menu Page” - Exit the MENU Under “Menu Page” - Select the related menü DOWN Modify the values in the configuration state Under “Menu Page” - Enter the sub-menu or confirm modifi- ENTER cation Page 14 www.ceyeko.com...
The instrument is equipped with monochrome lattice LCD, 128*64 resolution. Push [MENU] to enter password verification page; input password to enter the home page. Push [EXIT] to enter alarm inquiry page, to inquire the current warning configuration information. pH Monitoring Page ORP Monitoring Page Page 15 www.ceyeko.com...
: Calibrate or correct pH and ORP signal. Remote Setting : Settings of RS485 parameters and current transmission output. Alarm Setting : Settings of parameters of high and low warning. Information Inquiry : Current version number. Page 16 www.ceyeko.com...
ORP electrode two types. Temperature Compensation : Set the automatic warming or manual temperature compensation, the temperature range of -10-130°C. Temp Switch: auto temperature offset NTC10K or PT1000 or manual temperature compensation, temperature range: -10°C—130.0°C. Page 17 www.ceyeko.com...
4 . 3 . 5 R e m o t e S e t t i n g RS485 Setting : Settings of 485 communication address and baud rate. Current Transmission : Settings of 4mA corresponding value and 20mA corresponding value of 4-20mA output. Page 18 www.ceyeko.com...
4 . 3 . 7 I n f o r m a t i o n I n q u i r y Information Inquiry : Inquire the current hard software version, high traceability. Page 19 www.ceyeko.com...
Unsigned pH Value (default two 0x0000 4 Bytes 0x03 Read Only Long decimal places) Temperature 0x0002 Short 2 Bytes 0x03 Read Only (default 1 decimal) Unsigned ORP Value 0x0003 4 Bytes 0x03 Read Only Long (signed integer) Page 20 www.ceyeko.com...
Page 21
00 FA is the returned temperature value, 25.0°C, unit: °C. The value obtained shall be divided by 10 to get the current temperature value. The range: -10.0-130.0°C. 05 C7 is the CRC16 check code, which is varying depending on different data. Page 21 www.ceyeko.com...
When the above methods are invalid, sandpaper grinding can be used to remove the mechanical method of grinding. This method may cause the sand under the grinding into the liquid interface. Causing permanent clogging. Page 22 www.ceyeko.com...
Answer: Check if the standard solution is polluted. Replace the solution and calibrate again. The response of number is slow? Answer: If the electrode is covered by dirt, the response would be slow. Clean the pollutant in a corresponding method. A slow response is normal in winter. Page 23 www.ceyeko.com...