1. Liner sizing tool and blade sizing tool 1) Remove the existed acetabular liner. EZX is an a�achment of Surgical Handpiece that delivers the rota�onal force of Surgical 2) Measure the inner diameter of the cup by using the Liner sizing tool included in the tray.
Page 3
Keep clean and be careful about moisture and dust. 2. Maintenance method 1. When assembling the liner to the EZX sha� head, push the liner onto the distal end on the Appropriate regular maintenance on the product and its components is required to sha�...
Page 4
Manual cleaning method Steriliza�on method 1) Rinse soiled device under running cold tap water for at least (2) minutes. Remove gross Steam steriliza�on is a safe and effec�ve method, and there are no restric�ons in sterilizing soil using a sponge, so� lint-free cloth or so�-bristled brush. For cannula�ons of the the product.
Page 5
6. Do not use the blade to cut cups, pins, or screws. Complica�ons related to EZX are very rare, but when they occur, they can be serious. 7. Do not subject the device to strong impact such as bending, breaking, or twis�ng as it Most complica�ons are related to surgery techniques and possible complica�ons would...
Page 6
c. Regular inspec�on must be conducted as prescribed to preserve the product in its op�mal status. d. Do not arbitrarily adjust or alter the product. e. Please return devices and a�achments no longer in use to your local representa�ve of IMEDICOM for disposal.