Troubleshooting; Cleaning And Sterilization; Storage, Maintenance And Transportation; Storage - WALDENT ENDO PRO SMART-TOUCH User Manual

Table of Contents

Advertisement

the automatic calibration, which is used when changing the contra-angle;
9) Click the "
" button in the lower right corner of the screen, and then click "
factory settings.
10) Click the "
" button in the lower right corner of the screen, then click "
enter the user guide, and press the "Quit" button to exit.

5. Troubleshooting

1. The incorrect battery
No display on the screen after
1
installation.
turning it on
2. Low power.
1.The contra-angle handpiece
2
File is not working
is not connected correctly.
2.Defective contra-angle handpiece
1. Resistance is too large
Endo file stops
3
2. Root canal not in ideal condition
1. Root canal in a bad Situation
Root canal measurement values
2. Electromagnetic interference
4
are inaccurate
1. The mode with automatic
positive / reverse rotation
function is not selected.
Endo file does not reverse rotate
5
2. A too large torque value was
set
3. Endo file does not reach the
root tip region.
1. Root canal is too dry.
2. A poor connection of file clip or
Endo file doesn't work in
measuring wire.
6
3. Oral mucosa is too dry.
combination Enlargement with
4. The doubIe-head measuring
apex locator mode
wire is connected to main unit.
5.Electromagnetic interference.
6.Poor conductivity of endo file
15
" to restore
" to
1.Install the battery correctly.
2. Recharge the battery.
1.Reconnect the contra-angle
2.Replace the contra-angle
1. Set a higher torque
2. Change to non-root-canal measuring mode.
1. Removal of liquid and residual pulp
2. Please check the equipment used
nearby, and be away from interference
1. Select automatic positive/reverse
rotation model
2. Set a lower torque value
3. Endo file will automatically rotate
when reaching root tip.
1. Lubricate the canal.
2. Secure the connection first.
3. Wet oral mucosa.
4. Replace with a single-head measuring
wire.
5 Please check the equipment used
nearby, away from interference sources.
6.Use endo file with good
conductivity.
1. Resistance is too large
2. Root canal not in ideal condition
3. Root canal is narrow
File reversed too frequently
7
4. Endo file is oversized
5. Electromagnetic interference
Notice: lf the problem can't be solved , please contact the local distributors or manufacturer.

6. Cleaning and Sterilization

1) After use, all parts that have contacted with the patients should be wiped by sterilized towel.
2) Cleaning with chemical reagents may cause damage to the instrument.
3) Contra-angle, lip hook, clip, handpiece base,rubber case and probe must be autoclaved
before use.
4) Main unit, handpiece and measuring wire can't be autoclaved
pressure.

7. Storage, maintenance and transportation

7.1 Storage

1) Handle with care, far away from vibration source. Install or store in a cool, dry, well-ventilated area.
2) Do not store with toxic, corrosive, flammable, explosive items.
3) Store in environment of relative humidity no more than 85%, atmospheric pressure 70kPa ~ 106kPa,
temperature -10℃~ +50℃.

7.2 Maintenance

1) The product does not contain any user-serviceable accessories. Instrument maintenance
should only be carried out by the professional trained maintenance personal.
2) If the battery is fully charged but does not allow the instrument to function for at least one day,
the battery must be replaced as soon as possible. Please use the original accessories, so as
not to cause damage to the instrument.

7.3 Transportation

Avoid excessive shock during transport, handle with care, keep away from dangerous goods
and avoid sun and rain.
1. Set a higher torque
2. Removal of liquid and residual pulp
or change to non-root-canal measuring
model.
3. Root-canal enlarge
4. Replace with a smaller size file
5. Please check the equipment used
nearby, away from interference sources.
under
high temperature and
16

Advertisement

Table of Contents
loading

Table of Contents