Section 1 Installation and Start-up Installation Environment ➢ Environment temperature: 5 ℃ ~ 28 ℃ , do not exceed 32 ℃ , the best temperature is 18℃~25℃, please use air conditioning system if necessary. ➢ Environment humidity: <80% RH. It shall be lower than 57% RH if the highest work temperature is 32℃.
➢ If the socket is not installed with grounding wire, please install grounding wire by qualified technician. ➢ Please install the independent overload and leakage protector at the power supply terminal of the storage cabinet. Choose Location When the freezer is in position, set the front casters’ brakes. Note: The freezer must not be moved with the product load inside.
Page 4
➢ Please use power socket with grounding wire. If the socket is not installed with grounding wire, please install grounding wire by qualified technician. ➢ Do not connect this equipment with the ground through gas pipe, water pipe, telephone wire or lightning rod because above mentioned grounding will cause electric shock.
adjust settings. Wrong operation may cause abnormal working of the device, such as bad temperature uniformity and abnormal alarm. Parameter Settings When power on the medical refrigerator, the display screen displays internal temperature value. To change setting temperature, for example, the original set temperature is 2℃...
Notes: When normal display long press key 9 5s,to delete a record high and low temperature data, check high and low temperature data at this time, showing “00” ,time display code +00,such as “HD00”.The current data logging time interval is 10min, the default is 17:35 minutes to clear records of high and low temperature data.
Error Display ➢ When faults of the refrigerator temperature display sensor including short circuit, open circuit, malfunction and incorrect connection appear, “AD1S” is displayed. When faults of the refrigerator temperature control sensor including short circuit, open circuit, malfunction and incorrect connection appear, “AS1S”...
Page 10
should confirm in advance that whether inside temperature of the medical refrigerator is in accordance with the set temperature for storing to avoid losses caused by goods damage due to the improper temperature. ➢ Due to the refrigeration inertia, the medical refrigerator cannot be kept at constant temperature.
Section 3 Daily Maintenance Clean the Storage Cabinet ➢ Please clean the storage cabinet once a month. It can keep the new appearance of the storage cabinet after periodical cleaning. ➢ Wipe off dust from the enclosure and internal chamber and all accessories of the storage cabinet.
Section 4 Trouble Shooting Please check the followings before notifying the repair technician if there’s fault: The Equipment Can’t Start ➢ Check if power is correctly connected, if the main power switch is turned ➢ Check if the power voltage is too low? ➢...