Summary of Contents for Onpharma Onset Mixing Pen

  • Page 1 S y s t e m Reference Guide...
  • Page 2 (sold separately by Onpharma). For more information, please refer • After giving the injection buffered to the Instructions for Use provided with the Onset Mixing Pen. with Onset, stay with your patient instead of leaving the operatory.
  • Page 3 Step-By-Step Guide Step 1. Prime the Onset Mixing Pen First, Reset Piston Rod Before loading the Mixing Pen, unscrew the Dosing Mechanism from the Cartridge Chamber and check to make sure the Piston Rod is fully retracted. If the Piston Rod is not fully retracted, as in...
  • Page 4 Be aware that when priming by pushing the Dispensing Button, it is important to depress the button straight along the axis of the pen to decrease the possibility of jamming the pen. If the Volume Dial reaches and you have not yet felt the piston rod make contact with the cartridge, repeat this third step.
  • Page 5 Step 3. Load a 1.7mL Cartridge Load the 1.7mL Cartridge First, insert a 1.7mL cartridge (for example, a 1.7mL cartridge of lidocaine with epinephrine) into the Cartridge Connector by pushing the septum end of the 1.7mL cartridge straight into the open end of the Cartridge Connector, pushing the 1.7mL cartridge straight in and firmly into place.
  • Page 6 Step 5. Mix Solution Push Dispensing Button Dispense Solution to Mix. Depress the center of the dispensing button all the way until it stops. Do not force the button, but rather press gently down along the pen axis until the volume window reads "0". It is important to push the dispensing button straight in, without pressing against the rotating dial or pressing at an angle.
  • Page 7 Once the solutions are mixed, start your injection within one minute for best results. Reloading the Onset Mixing Pen Note: You will need to remove and discard the Cartridge Connector each time you replace the 3mL cartridge in the Pen.
  • Page 8 Customer Care 877.336.6738 customercare@onpharma.com Hours Monday - Thursday | 7AM - 5PM PST Friday | 7AM - 2PM PST Please see accompanying package insert and usage instructions. These materials are copyrighted by Onpharma Co., 2021. MK047, Rev. E SKU 503...