Operations and Safety Manual Before using Skiin Underwear, please read this manual.
Page 2
Operation and Safety Manual for Skiin Underwear. Version information System: Skiin Underwear System Software: Skiin Connected Life App Version 1.0.0...
1. Introduction Contraindications This Operation and Safety Manual for the Skiin The Skiin Underwear is not for use by: Underwear System provides instructions for the • Persons under 18 years of age safe and efficient use of the Skiin Underwear •...
3. Contents The standard configuration for the Skiin Connected (B) USB Charging Kit (A) Skiin Pod Clothing System includes the following components: (A) 1 Skiin Pod (B) 1 USB Charging Kit: (C) Skiin Garments - 1 USB Charging Dock - 1 Micro-USB Cable - 1 Wall Plug (C) Skiin Garments: Each kit has one "Try Me On"...
It is recommended to keep your smart device close to you whenever possible. When you put on the Skiin Underwear, it can start sensing your heart rate and physical activity, which will be captured and stored as data by Skiin Pod and transmitted to Sensor Zones the Skiin Connected Life App.
STEP 3: Download the Skiin Connected Life App SIZE RANGE (INCHES) Women’s On your smartphone, download the Skiin Connected Life App from the Apple App Store or the Google Play Store. 29.5-32 Open the App and follow the onboarding steps to setup 32-34.5 and pair the Skiin Pod, start logging your health data and 34.5-37...
6. Skiin Underwear System Components Serial Number and markings Only the components listed in this section WARNING may be connected to the Skiin Underwear EN-W4 Pod Connected State System. If components other than those specified in this section are connected to...
Data Storage and Syncing 6.2 Skiin Charging Kit Your Skiin Pod will automatically sync data to Charging The Pod your smart device when it is within range and When the Pod is inserted into the Charging Dock and the app is opened or in the background. connected to a power source, the Pod’s LED will Appear YELLOW to indicate that the Pod is currently charging.
Rapid Blink Red Low battery level & the Pod is NOT connected to the App • Before wearing for the first time, please wash 6.3 Skiin Underwear • Wash after wearing the garment for 24-48 hours Materials • Remove the Pod and put all garments in the washbag, use cold water when washing Skiin Garments can be washed like regular clothing.
Connections Machine Wash on Cold The Connections page helps you to connect, share and communicate with your friends, family and health care Tumble Dry on Medium professionals. You can connect with your connections individually and also as a part of a group, sharing all Do not Wring the Garment your captured biometric and contextual data.
6.4.2 Getting Help Battery Level: Garment Signal Quality: Navigate to Menu (User Settings)> Help to Battery Empty access Contact Information for Technical Good Signal Support and Frequently Asked Questions Noisy Signal Navigate to Menu (User Settings)> About for information about Software, Terms and Battery Full Conditions and the Privacy Policy No Signal...
Myant will only replace the defective part Operational Time, 1 year warranty of the product after it has been returned and found to be defective due to normal use. Myant will not replace any products that have been tampered with or Physical Characteristics found to have been broken due to user negligence.
Operational Humidity ..15% to 90%, non-condensing, the band of the Skiin Underwear. These electrodes water vapour partial pressure <50hPa Operational make contact with the user’s skin. The Skiin Underwear Altitude............700hPa to 1060hPa Electrodes may be cold washed and tumble dried low Storage Humidity ..........15% to 90%,...
Page 15
Indicates authorization under the FCC FCC title 47, Part Symbols on the Skiin Underwear System Device Label (Skiin Pod) 18.209 b Declaration of Conformity procedure. (symbol shown in Symbol Meaning Reference 47 CFR 15.19 b) MMWWWWW Serial Number Temperature range limitation...
Page 16
Symbols on the Wall Plug Full name of Standard Indicates authorization FCC title 47, Part IEC 60601-1:2012-Ed.3.1 Medical Electrical Equipment Part 1: under the FCC Declaration of 18.209 b General requirement for basic safety and essential performance Conformity procedure (symbol shown in 47 CFR 15.19 b) IEC 60601-1-2:2014-Ed.4.0 Medical Electrical Equipment Part 1: General requirement for basic safety and essential performance,...
EN-C2 latter as to how the device should be used. anesthetics, drugs or pressurized oxygen. WARNING Contraindications The Skiin Underwear is not for use by: EN-W8 The Skiin Pod contains magnets, use with • Persons under 18 years of age.
Do not use the Skiin Pod, or Skiin Charging WARNING a particular installation. If this equipment causes harmful Kit or Skiin Underwear if they are damaged EN-W14 interference to radio or television reception, which can in any way.
Page 19
Table 1 – Emission Class and Group Compliance The Skiin Underwear is intended for use in the electromagnetic environment specified below. The customer or the user of the Skiin Underwear should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment –...
Page 20
Table 2 – Immunity Test Level Compliance The ME EQUIPMENT is intended for use in the electromagnetic environment specified below. The customer or the user of the ME EQUIPMENT should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment –...
Page 21
0% U Mains power quality should be that of a typical Voltage Dips, short 0% U (100% dip in U commercial or hospital environment. If the user of the interruptions and (100% dip in U ME EQUIPMENT requires continued operation during for 0,5 cycle voltage variations on for 0,5 cycle...
Page 22
Table 3 – Immunity Test Level Compliance The ME EQUIPMENT is intended for use in the electromagnetic environment specified below. The customer or the user of the ME EQUIPMENT should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment –...
Page 23
Table 4 – Recommended Separation Distances Recommended separation distances between portable and mobile RF communications equipment and the ME EQUIPMENT The ME EQUIPMENT is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ME EQUIPMENT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ME EQUIPMENT as recommended below, according to the maximum output power of the communications equipment.