ACCURACY: The degree of absolute conformity to a TEMPERATURE standard. NIST: National Institute of Standards and Technology Founded in 1968, Alpha Technics Precision Temperature has become a recognized leader PRECISION: The accuracy with which a number in the precision hand-held thermometer can be represented;...
- T A B L E O F C O N T E N T S - (continued) INTRODUCTION ........1 PROBES: OPERATING INSTRUCTIONS ....... 3 ALPHA TECHNICS 1400 SERIES PROBES for Model 4400-2 & 4600-2 Thermometers RS-232 INTERFACE ........11 EUT1401 0.030” Stainless Steel Needle Probe THERMAL VIEW SOFTWARE ...... 12 EUT1402 7.5”...
Space for the 4600, several probes, power NIST Traceable supply, Thermal View software, Cal plugs, etc. The Model 4600 Series are multi-purpose, economical, ultra-high accuracy high- EUT4653 Calibration Plug Kit resolution, hand-held digital thermometers Allows the user to field check the calibration which accept standard YSI Series 400 of the 4600 at several temperatures.
Page 5
The model 4600 accepts standard YSI Series 400 thermistor probes with phone jack termination. Alpha Technics high-accuracy probes are also available. C o m p u t e r I n t e r f a c e An RS-232 computer interface allows the user to download information from the 4600 to their own PC to create reports and graphs.
Battery needs symbol replacement** *Some versions of 4600 Series thermometers will have different temperature ranges. See the special specifications in the back of this manual or packed separately with the unit. **To ensure that erroneous results are not displayed, the numerical segments are disabled (all turned on or all off) one hour, of running time, after the battery symbol starts flashing.
Page 7
Model 4600 Manual-3.5x8.5-rev.indd, Spread 7 of 16 - Pages (26, 7) 1/11/2007 2:03 PM 4 6 0 0 S E R I E S T H E R M O M E T E R 4 6 0 0 S E R I E S T H E R M O M E T E R - N O T E S - - O P E R A T I N G I N S T R U C T I O N S -...
Page 8
Model 4600 Manual-3.5x8.5-rev.indd, Spread 8 of 16 - Pages (8, 25) 1/11/2007 2:03 PM 4 6 0 0 S E R I E S T H E R M O M E T E R 4 6 0 0 S E R I E S T H E R M O M E T E R - O P E R A T I N G I N S T R U C T I O N S - - S P E C I F I C A T I O N S -...
Page 9
MODEL 4600-2 SPECIFICATIONS: THERMISTOR: 10k Ohms @25°C used in When the MODE key is pressed, the active Alpha Technics’ Series 2400 probe housings. function icon flashes and the other function icons turn on solid. Pressing MODE again PROBE CONNECTION: 1/4” phone jack, scrolls the display to the next function.
Page 10
Model 4600 Manual-3.5x8.5-rev.indd, Spread 10 of 16 - Pages (10, 23) 1/11/2007 2:03 PM 4 6 0 0 S E R I E S T H E R M O M E T E R 4 6 0 0 S E R I E S T H E R M O M E T E R - O P E R A T I N G I N S T R U C T I O N S - - S P E C I F I C A T I O N S -...
Page 11
Model 4600 Manual-3.5x8.5-rev.indd, Spread 11 of 16 - Pages (22, 11) 1/11/2007 2:03 PM 4 6 0 0 S E R I E S T H E R M O M E T E R 4 6 0 0 S E R I E S T H E R M O M E T E R - S P E C I F I C A T I O N S - - O P E R A T I N G I N S T R U C T I O N S -...
Page 12
Model 4600 Manual-3.5x8.5-rev.indd, Spread 12 of 16 - Pages (12, 21) 1/11/2007 2:03 PM 4 6 0 0 S E R I E S T H E R M O M E T E R 4 6 0 0 S E R I E S T H E R M O M E T E R - O P E R A T I N G I N S T R U C T I O N S - - S P E C I F I C A T I O N S -...
Page 13
Alpha Technics. Alpha Technics assumes no risk allows the user to vary the auto shutoff time for damage in transit. Alpha Technics will, at its period. The thermometer can be configured option, repair or relace the defective instrument to shutoff from 3 to 998 minutes, or not auto free of charge.
Instruments covered by the limited warranty will be promptly repaired or replaced, at Alpha The 4600 Series thermometers enter the HOST mode when the RTS Technic’s option, and returned at no charge. line becomes active true. When the RTS line becomes FALSE, the 4600 returns to the normal mode of operation.
Recalibration should be performed at Alpha Technics. When the Savable “Protocol” windows allow the user to Thermometer requires recalibration, forward select start/stop times and temperatures, set...
- B A T T E R Y R E P L A C E M E N T - - C A R E & C L E A N I N G - The Model 4600 Series uses a standard 9 Volt Extensive testing has proven this case to be alkaline battery.