Calibration Of Reservoir Gauges - PRESTON PRESSURE PS-500 User And Maintenance Manual

Pitot-static/air data tester
Table of Contents

Advertisement

10.
After the calibration point has been entered, the display will briefly read
"Calibration Saved" and then exit the calibration mode.
11. When a satisfactory calibration is obtained at this test point, go through each test
point in the following chart and enter the corrected value at each point. (The correction
chart may be photocopied, cut out, and fixed to the test box for handy reference.)
12. Make a written note of the calibration correction value. In the very rare case that a
calibration factor is lost from the memory, it can be re-programmed without a complete
calibration procedure. The correction value can also serve as a record to see if the sensors
are drifting with age.
Maximum allowable error at any airspeed is 2 Knots. (Generally, a maximum error of 1
Knot should be expected.) The airspeed has only one calibration point.
PS-500 S/N
FOR TEST
DISPLAY
FOR TEST
AIRSPEED
AIRSPEED
AIRSPEED
20 Kts
275 Kts
40 Kts
300 Kts
60 Kts
325 Kts
80 Kts
350 Kts
100 Kts
375 Kts
120 Kts
400 Kts
140 Kts
425 Kts
160 Kts
450 Kts
180 Kts
500 Kts
200 Kts
550 Kts
225 Kts
600 Kts
250 Kts
650 Kts
Calibrated by: ___________ Date: _________
Next Calibration Due Date: _______________
Calibration Notes:
 Do not forget to wait the minimum of 10 minutes before calibrating.
 The calibration procedure can be entered only when the mbar mode is selected.
 Do not forget to check bead colors in the air pressure reservoir/dryer. If they are
pink or clear, it is time to change the unit to avoid a build-up of moisture in the
pressure system. Replacement air dryers can be purchased at many hardware
stores, or on the internet (do a search for "mini desiccant dryer").
The reservoir gauges are for reference only, and do not need calibration. The operator
merely references these gauges as an indication of how much reserve pressure or vacuum
there is in the reservoirs, and as a guide to know when to turn on the pump(s).
DISPLAY
AIRSPEED

Calibration of Reservoir Gauges

18