CAUTION: Federal (US) law restricts this device to sale by or on the order of a licensed professional.
INDICATION FOR USE
The MediLectric System is intended to convert pneumatic output from a dental treatment
center into electrical energy in order to drive an electric micromotor that operates electrical-
ly-driven dental handpieces. This system is designed for use by a trained professional in the field
of general dentistry.
ELECTRICAL SAFETY INFORMATION
• Do not handle the power cord with wet hands to prevent an electric shock.
• Do not expose the device to water to prevent an electric shock.
• Do not use the device in a place where flammable materials are found due to risk of explosion.
• No part of the control unit shall be repaired by the user. Do not disassemble the device.
• Place the device on a stable location. Do not drop the device or exert an external force that
may damage the device.
• Do not use a mobile phone or wireless RF communication device close by as such adjacent
use may adversely influence this electrical device.
• This product shall be used only under the rated power specified. Power outside the rate
indicated may damage unit
• Do not use device with a handpiece that does not meet the standard requirements.
• If motor begins to run rough, discontinue use and send in for request to the supplier.
• Check if the speed is set within the acceptable speed range prior to using the speed
control button on the control panel.
• Use only the adapter that is provided by the manufacturer for this device.
• Do not disassemble or tamper with micromotor or control unit as it may damage the device.
• Micromotor generated greater power and torque than traditional air turbines of air motors.
Therefore, handpiece that equipped with micromotor with poorly serviced, damaged or
used improperly can overheat which can cause serious burn injuries to the patients.
• The following guide lines should be followed to ensure the safe operation of electric
handpieces:
• Follow maintenance instruction of handpiece manufacturer.
• Examine the handpiece for damage before each use
• Never apply pressure on the chuck button while handpiece is in operation or use handpiece
as a cheek or tongue re-tractor.
INSTRUCTIONS FOR USE
Read before use. Keep in a safe place for future reference.
Item Code: EJ0008-MEDI

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MediLECTRIC EJ0008-MEDI and is the answer not in the manual?

Questions and answers

Summary of Contents for Medidenta MediLECTRIC EJ0008-MEDI

  • Page 1 Item Code: EJ0008-MEDI INSTRUCTIONS FOR USE Read before use. Keep in a safe place for future reference. CAUTION: Federal (US) law restricts this device to sale by or on the order of a licensed professional. INDICATION FOR USE The MediLectric System is intended to convert pneumatic output from a dental treatment center into electrical energy in order to drive an electric micromotor that operates electrical- ly-driven dental handpieces.
  • Page 2 4. Scrap and recycle the device and its parts in accordance with the applicable local laws and regulations. 5. For any issues relating to the use of the product including technical inquiries, please contract Medidenta. SPECIFICATIONS Input 100-240V~, 1.4-0.7A, 47-63Hz Power Consumption 105 VA(Max.)
  • Page 3: Display And Controls

    DISPLAY AND CONTROLS NO. DESCRIPTION FUNCTION Rotational direction Used to select a rotational direction (Fwd/Reverse) of the micromotor. Micromotor operation direction. Displays the direction of the micromotor. Micromotor LED ON/OFF button Used to turn on and off LED. Micromotor LED indicator Indicates whether micromotor LED is running (LED On signals LED is turned on).
  • Page 4 DESCRIPTION FUNCTION Tubing connector-standard 4-hole Connect to handpiece hose from dental unit to handpiece hose connector supply air & water, tubing connector per ISO 9168 type 3 Motor tubing / cable Motor tubing/cable, supplies power to the motor, distributes air and water (it cannot be separated from the control unit.) Power switch Main power switch for unit.
  • Page 5: Installation And Setup

    INSTALLATION & SET UP • Check for damage to the cable and the connection pins prior to connecting with the micromotor. • Check system connection diagram and cable / motor connection illustration below prior to installation. System connection diagram Medilectric Control box 2.
  • Page 6 Handpiece Tubing Connector 1. Confirm that the delivery system tubing matches with the back terminal of the control unit for connection (Fig. 6) (coupling connection per ISO 9168 type 3 standard 4-hole handpiece hose) 2. Connect the tubing to the control unit and secure the connection to prevent the air or water leakage (Fig.7) Fig.
  • Page 7 display. Selection of the program mode 2. Press and hold the button for selecting micromotor’s operational direction and the LED on/off button simultaneously for 3 seconds or longer.. Memory program 1. How to save memory program : Customize settings of rotational direction, optic and so on; Press and hold one of the three program buttons M1, M2, or M3 for 3 seconds or longer.
  • Page 8 Manual external cleaning for micromotor and motor tubing: Recommendation for external cleaning accessories: • Clean cotton cloth • Soft brush-Nylon Bristles • Alcohol 70% Isopropyl • Protective glove uses during cleaning procedure. 1. For control box, micro motor cable / tubing (non removable) can be disinfected by wiping off using soft cotton cloth with isopropyl alcohol.
  • Page 9: Troubleshooting

    • There is no specific replacement period. Changing the O-ring To replace the motor O-rings: (Fig.13) Fig. 13 1. Squeeze the O-ring together between your thumb and forefinger to form a loop. 2. Pull the O-ring off the motor. 3. Repeat steps 1 and 2 for the other three O-rings. 4.
  • Page 10: Emission Test

    Poor connection with a power Check the power connection. cable Poor connection with the Check the connection condi- When the micromotor is not connection cable tion with the cable. running Control panel malfunction Contact the manufacturer. Controller malfunction Contact the manufacturer. Micromotor malfunction Contact the manufacturer.
  • Page 11: Immunity Test

    described below. IMMUNITY TEST EC 60601-LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT GUIDANCE Electrostatic ± 6kV contact ± 6kV contact Floor should be discharge(ESD) ± 8kV air ± 8kV air wood, concrete or IEC61000-4-2 ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %...
  • Page 12 IMMUNITY TEST EC 60601-LEVEL COMPLI- ELECTROMAGNETIC ENVIRONMENT ANCE LEVEL GUIDANCE Portable and mobile RF communi- cations equipment should be used no closer to any part of the product, including cables, than the recom- mended separation distance calcu- lated from the equation applicable to the frequency of the transmitter.
  • Page 13: Warranty

    This product is warranted against manufacturing defects in materials and workmanship for years from date of purchase. Medidenta will repair or replace product at its sole discretion. This warranty shall be limited to replacement or repair of the product or its parts and shall not extend to any other claims, including but not limited to loss of profit, cost of removal or replacement, incidental or conse- quential damages or other similar claims arising from use of this product.