Page 1
This manual is intended as a guide to the use and installation of the product and, therefore, PCME Ltd. shall not be liable for any loss or damage whatsoever arising from the use of any information or details therein, or omission or error in, this manual, or any miss-use of the product.
Page 4
QUALITY ASSURANCE AND SELF TESTS ............55 ....................... 55 NTRODUCTION PCME STACK 990 S ................ 56 ENSOR ESTS Zero check and Span Checks.................. 56 Short Circuit Check ....................56 Contamination Ring Check ..................56 - PCME STACK 990 MANUAL: 4 -...
Page 5
More Sensor Settings .................... 74 Basic System Setup....................75 Other Functions....................76 QA and Calibration....................77 Document Control Information SOP Reference PCME STACK 990 Issue No Issue Date 21/07/09 Software Version 7.00 Filename PCME STACK 990 v3.doc - PCME STACK 990 MANUAL: 5 -...
The procedures given in this manual must only be carried out by suitable trained and qualified personnel. 1.2 Disclaimer PCME Ltd. will not be liable for any injury or damage caused by incorrect installation, set-up, operation or maintenance resulting from a failure to follow the procedures given in this manual. 1.3 Product Safety...
The following abbreviations are used in this manual: Auxiliary Input Module Analogue Output Module Power Supply Unit Relay Output Module 1.7 Associated Documents Control Unit Reference Manual Installation of Network Devices: Reference Manual PCME Installation Notes - PCME STACK 990 MANUAL: 7 -...
Ensure opening/hole in stack wall is at least 45mm diameter Power requirements 24V provided by the control unit RS-485 connection to control unit Local outputs Optional: 4-20mA (250 Ohm, non-isolated) Optional: relay (1A @ 30V) Cable entries 3 x M20 gland/conduit entries - PCME STACK 990 MANUAL: 8 -...
Cable type 4 core screen (see installation section for details) May be required on some applications (consult PCME). Air Purge requirements Requires optional air purge fitting and an external supply of 5-10 litres/min of dry clean, oil free instrument air...
Probe damage may result in inaccurate measurements. Supply Voltage The sensor operates at 24V DC supplied from the Control Unit. 3.2 Data Cabling Specification The cabling provided by PCME has the following specification: • 2 pair (4 core), 0.5mm •...
For large systems it is best to place the control unit in the centre of the network to minimise voltage drop along cables. For large systems you may need to boost power using a Power Supply Repeater – contact PCME for details.
Page 14
4-core Power (24V DC)/ Comms (RS485) 115/230V AC Figure 3-1 Standard System 4-core Power (24V DC)/ Comms (RS485) 115/230V AC 4-core Power (24V DC)/ Comms (RS485) Figure 3-2 PLUS System (“Spur Linked”) - PCME STACK 990 MANUAL: 14 -...
BUS OUT terminals. Make the connections to the BUS OUT terminals as follows from left to right: Blue BUS OUT TERMINAL Comms B Orange *Connect the cable screen to the Comms A Green earth stud Brown (Screen) Do not use* - PCME STACK 990 MANUAL: 16 -...
Page 17
Figure 3-5 PCME Stack 990 Sensor Unit BUS OUT(power and comms) Contamination probe Contamination probe enable/disable BUS IN(power and comms) (shown enabled) 4-20mA and relay outputs (option) Bus termination switch (SW200) Set comms switch (SW101) Main probe enable/disable switch (shown enabled)
The Bus Termination Switch (SW200) identifies the sensor’s position in the network. Set the switch to “U” for all sensors in the chain except the last sensor, which must be set to T. For spur- systems leave all sensors terminated (switch to “T”). - PCME STACK 990 MANUAL: 18 -...
RTU is a faster communications protocol but is not available for all devices; refer to the Control Unit Reference Manual. If RTU mode is required set the DIP switch as follows: Refit the Sensor Cover and secure with the four screws. Do not over tighten the screws. - PCME STACK 990 MANUAL: 19 -...
EARTHING THIS EQUIPMENT MUST BE EARTHED NO User Serviceable parts within. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. - PCME STACK 990 MANUAL: 20 -...
Template Height 110 mm ** the mounting holes are elliptical to allow flexibility of position Data Cabling Specification Data cabling for the MultiController provided by PCME meets the following specification: • 2 pair (4 core), 0.5mm • BS5308 part 2, type 1 •...
Route the data cable for one half of the network through the cable gland nearest to the Data Bus PL1 and make the connections to the Data Bus PL1. - PCME STACK 990 MANUAL: 24 -...
Route a suitable cable through a vacant cable gland and connect to PL3 as shown. The connection should be made between the required input and common ground. - PCME STACK 990 MANUAL: 25 -...
Relay 4: Self-test alarm Relay contacts are shown in the normal operating positions- power is applied and no alarm exists. Route a suitable cable through a vacant cable gland and connect to PL5 as shown. - PCME STACK 990 MANUAL: 26 -...
Refit the unit cover and secure with the four screws. Do not over tighten the screws. Carry out the initial set-up procedures given in the Initial Setup section. - PCME STACK 990 MANUAL: 27 -...
EARTHING THIS EQUIPMENT MUST BE EARTHED NO User Serviceable parts within. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. - PCME STACK 990 MANUAL: 28 -...
If the unit is located in an area which is accessible to unauthorised personnel it is recommended that access to the unit be restricted by some form of lockable cabinet. - PCME STACK 990 MANUAL: 29 -...
Route a suitable cable through a vacant cable gland and connect as shown to the HOST/PC PORT. The maximum cable length for RS232 is 25 metres. The maximum cable length for RS485 is 1000 metres. - PCME STACK 990 MANUAL: 31 -...
The Interface Module is fitted with two voltage-free SPCO with a 3A current rating. The relays can be used to switch mains voltages. The maximum current through the alarm contacts must not exceed 3 Amps. - PCME STACK 990 MANUAL: 32 -...
Refit the unit cover and secure with the four screws. Do not over tighten the screws. Carry out the initial set-up procedures given in the Initial Setup section. - PCME STACK 990 MANUAL: 33 -...
When using the multicontroller for the first time, communication must be established between the multicontroller and all the sensors in the network. If the multicontroller is being re- configured, all existing settings can be cleared by performing a ‘Master Reset’. - PCME STACK 990 MANUAL: 34 -...
Select ‘Setup’ display Menu route: Other Functions ↵ Reset Functions ↵ Master Reset The Communications Mode is set by PCME prior to installation and should have been checked during the system installation. To check the Communications Mode Select ‘Setup’ display Menu route: ↵...
When the calibration factor is adjusted, the new calibration factor is only applied to NEW logged data. Existing logged data remains calibrated to the old calibration factor. The control unit maintains a history of calibration factor changes for use by the PCME Dust Reporter PC software - PCME STACK 990 MANUAL: 37 -...
Alarm Delay is set to 5 secs. Increase this as required to the duration of the dust spikes you are seeing (look at the pulse log to access this). Notes: Note 1: The alarm delay is applied to both instant and average alarms. - PCME STACK 990 MANUAL: 38 -...
(i.e. if the time is put back 1 hour from 10.00 to 09.00 the Controller will not log data until the time reaches 10.00. If the time is put back by more than 24 hours the Controller will clear all logs and start logging again immediately. - PCME STACK 990 MANUAL: 39 -...
The following screen is displayed: 4-20mA Calibration Select Channel Start 4mA Test Digital Value 0665 Increase Current Decrease Current Start 20mA Test Digital Value 3373 Increase Current Decrease Current Back Figure 4-6 4-20mA Calibration - PCME STACK 990 MANUAL: 41 -...
4-20mA Filter Times: sets the filter time (walking window filter) applied to the instantaneous output. The default setting is to read between 0 and 100 units with a filter time of 10 secs. - PCME STACK 990 MANUAL: 44 -...
Comms Error All devices Limit Alarm All devices Warning Alarm All devices Self Test Fail All devices Back Relays can be assigned to individual channels/sensors and the alarm types can be user specified. - PCME STACK 990 MANUAL: 45 -...
Page 46
In the event of a power failure of a power loss to the Control Unit the relays will de-energise and alarms generated. Select ‘Setup’ display Menu route: Other Functions ↵ FailSafe Relays? - PCME STACK 990 MANUAL: 46 -...
All devices Limit Alarm All devices Warning Alarm All devices Self Test Fail All devices ROM 1 ROM 2 Back Select the ROM to be configured and the following screen will be displayed: - PCME STACK 990 MANUAL: 47 -...
Page 48
The FailSafe function only detects power failure to a ROM. It does not detect a communication failure. To monitor for communication errors set up a fail-safe relay to monitor the ROM for Comms Error. - PCME STACK 990 MANUAL: 48 -...
The 990 sensor relay output has the following specification • volt-free DPCO relay • current rating 1Amp per contact • option to wire normally open or normally closed • non fail-safe operation only - PCME STACK 990 MANUAL: 49 -...
Page 50
The default settings for the local alarm relays are to be triggered by Both Limit alarms. It is not possible to latch the relays. It is not possible to run the relays in fail-safe mode - PCME STACK 990 MANUAL: 50 -...
Select ‘Setup’ display Menu route: More Settings The following screen is displayed: Sensor Settings Add a new device Edit an existing device Delete an existing device Edit All Devices Back - PCME STACK 990 MANUAL: 51 -...
• only specific sensors are to be added. Select ‘Setup’ display Menu route: More Settings↵ Add a New Device Choose the correct channel type from the list and set the following parameters - PCME STACK 990 MANUAL: 52 -...
Delete an existing device Select the device to be deleted. This will permanently remove all settings and logged data associated with the device. Entries in the Event Log for the device will be marked as deleted. - PCME STACK 990 MANUAL: 53 -...
Select the parameter to be edited and set the value. Select Save to apply the new settings to all the sensors/devices. Advanced Settings For information on more advanced sensor/device settings refer to the Control Unit Reference Manual. - PCME STACK 990 MANUAL: 54 -...
7.2 PCME STACK 990 Sensor Self Tests The QA screen allows the status of the hardware self-checks to be reviewed for each sensor. These may be set to run automatically and can also be manually forced (i.e. during installation or maintenance).
Yes / No Select Save to save the settings and return to the main menu. Adjusting the Contamination Ring Check Alarm Level The influence of the contamination ring signal may vary for different installations. - PCME STACK 990 MANUAL: 57 -...
Tests the communication between the controller and the selected sensor. If this test fails check the wiring between the controller and the sensor. Check the address and comms settings on the sensor and controller. - PCME STACK 990 MANUAL: 58 -...
To put all sensors into Maintenance Mode select Start Maintenance (All Devices). To return all sensors into Enabled Mode select Stop Maintenance (All Devices). To place individual sensors in Maintenance Mode: select Individual Device Maintenance: The following screen will be displayed: - PCME STACK 990 MANUAL: 59 -...
Page 60
Active Set to 0 Outputs Active Set to 0 Bar Graph Displays Active (labelled MAINT) Active Logging Active Inactive (labelled MAINT)* Maintenance Alarm Log Alarm *Data may be excluded from Dust Reporter reports. - PCME STACK 990 MANUAL: 60 -...
Action Required installation Inspection due 6 months Remove sensor from stack for inspection and cleaning Arrange service visit by PCME Service Engineer or a local Service due 12 months PMCE representative Recommended return to PCME for testing and refurbishment Major Service due...
Check the time and date are correct. The control unit does not automatically adjust 1 hour for Daylight Saving – you may wish to leave it unchanged. • Check that the connections to the control unit are correct and secure. - PCME STACK 990 MANUAL: 62 -...
Page 63
Watch the instrument performance with the process running and ensure that the process is still being followed correctly. This is best done by downloading data using PCME Dust Reporter software. Compare the data before and after cleaning: if cleaning has significantly altered the reading then more regular maintenance may be required.
10 CALIBRATION 10.1 Introduction The PCME Stack 990 sensors attached to the control unit can be calibrated in mg/m³ in certain applications. Calibration of each sensor is necessary on at least an annual basis and should be carried out either by PCME or an authorised PCME representative.
If required the Units can be set to mg/m Select ‘Setup’ display Menu route: ↵ More Settings ↵ Edit an existing device Select device to edit Units Name Select Save to save the settings and return to the main menu. - PCME STACK 990 MANUAL: 65 -...
The calibration tool can be used either to: • carrying out a simple calibration with a single sample. • obtaining sensor averages for sample runs to send to PCME Select ‘Quality Assurance / ’Self Tests’ display Menu route: Device Select Device...
Factor in the sensor configuration and return to the QA main menu. If the new calibration factor is unacceptable select Back to return the calculation screen and repeat the calculation procedure or select Cancel to return to the QA main menu. - PCME STACK 990 MANUAL: 67 -...
Page 68
: 45 Duration Max Reading 7.3200 ShortTerm Sensor Average 5.6315 Test Result 17.94 Calc Exit Sensor Average 5.6315 Test Result 17.94 Current Cal 1.0000 New Cal 3.1856 Save New Calibration? Save Back Cancel - PCME STACK 990 MANUAL: 68 -...
Each 990 Sensor has a Red LED located internally that flashes at approximately 1Hz (with equal spacing: 0.5 secs on, 0.5 secs off) when power is correctly applied to the sensor (before the - PCME STACK 990 MANUAL: 69 -...
• the controller software is too busy reading other sensors. To solve this power the control unit off/on then the control unit will restart sharing its time between all sensors on the network. - PCME STACK 990 MANUAL: 70 -...
Remove rust from the thread and apply copper slip or grease If the stack if not metallic or you are concerned the stack may not provide a suitably earthed shield for the sensor then contact PCME. - PCME STACK 990 MANUAL: 71 -...
AIM Digital 3 Multi Cal facotrs AIM Digital 4 Diagnostics? Software Version 7.01 Multi Calibration Factors Back Cal Factor 0 1.0000 (fixed) Cal Factor 1 2.0000 Cal Factor 2 3.5000 Cal Factor 3 4.7000 Back - PCME STACK 990 MANUAL: 76 -...
: 45 Duration Max Reading 7.3200 ShortTerm Sensor Average 5.6315 Test Result 17.94 Calc Exit Sensor Average 5.6315 Test Result 17.94 Current Cal 1.0000 New Cal 3.1856 Save New Calibration? Save Back Cancel - PCME STACK 990 MANUAL: 77 -...
Page 78
New sections: Grounding of sensors Network layout 6 Initial Set-up Make joint for Multicontroller and Interface Module versions Add blue borders to screen drawings Tidy up display of menu routes 9 Troubleshooting Expanded - PCME STACK 990 MANUAL: 78 -...