Download Print this page
Cell Dynamics W8 First Installation Manual

Cell Dynamics W8 First Installation Manual

Physical cytometer
Hide thumbs Also See for W8:

Advertisement

Quick Links

Work instruction:
W8 PHYSICAL CYTOMETER UNBOXING
AND FIRST INSTALLATION

Advertisement

loading

Summary of Contents for Cell Dynamics W8

  • Page 1 Work instruction: W8 PHYSICAL CYTOMETER UNBOXING AND FIRST INSTALLATION...
  • Page 2: Purpose Of The Document

    • Power supplier • One USB 3.1 SuperSpeed cable • W8 Caps (two standard for Vessel A and Vessel B, one for the Sample Holder tube and one for the 50 mL W8 Analysis Solution) • 250 mL Waste bottle (including lid and connection tube) •...
  • Page 3 • Anti-virus software. This system is intended for use as a stand-alone station, not as part of a network. Please follow the instructions described below for a proper unboxing of CellDynamics W8 package. The tutorial video for the W8 system unboxing is available at the W8 Academy page on CellDynamics website.
  • Page 4 • One Corning® 50 mL centrifuge tubes to be used for the W8 Analysis Solution • the W8 Chip box. NOTE: This box is empty: the W8 Chip is already inserted in the W8 Physical Cytometer to prevent any damage throughout the shipment.
  • Page 5 • Maintain at least 15 cm of distance between the W8 Physical Cytometer and any wall, object or PC. This distance is also needed for proper ventilation.Moreover, keep a free area in front of the instrument to allow the opening of the safety door for proper use (Fig.
  • Page 6 Do not use the system without the air filter. NOTE: For proper daily use of the W8 Physical Cytometer replace the Filter once a month. To insert a new Filter into the appropriate slot, simply screw the Luer-Lock connector (Fig. 4).
  • Page 7 Step 6 W8 Caps positioning, W8 Chip positioning and centrifuge tubes insertion Please refer to the W8 Physical Cytometer User Manual 6.4, 6.5 and 6.6 sections for the complete procedures. Alternatively, please consult the W8 Physical Cytometer Quick Start Guide STARTING UP THE...