Do you have a question about the OtoPet-USA Earigator and is the answer not in the manual?
Questions and answers
Summary of Contents for DermaZoo OtoPet-USA Earigator
Page 1
OtoPet-USA Earigator Setup and Use Instructions VETERINARY USE ONLY OtoPet-USA is a division of DermaZoo P P e e t t 8909 Iverleigh Court, Potomac, MD 20854 - - USA Office: 301-365-0192 • Fax: 301-983-2369 Email: info@otopet-usa.com www.OtoPet-USA.com...
Page 2
OtoPet-USA Earigator Manual Introduction Lavage and Suction Your OtoPet-USA EARIGATOR is a versatile Irrigation, Using the EARIGATOR for lavage and suction: Suction, and Insufflation system for veterinary clinical • Be sure the unit is powered OFF use. Use it in conjunction with your OtoPet-USA •...
Page 3
OtoPet-USA Earigator Manual Insufflation • Using the EARIGATOR for insufflation: Insufflation is now applied by SLOWLY AND • CAREFULLY opening the insufflation valve to Turn the insufflation control fully clockwise to the achieve the desired inflation. OFF position. • •...
Page 4
OtoPet-USA Earigator Manual Power Cord and Switch Power Switch Power Cord Earigator Front Panel Controls Figure 2 Irrigation and Insufflation Suction Vacuum Gauge Pressure Gauge 0-15 PSI 0-30” Hg Irrigation Regulator Control Suction Regulator Control - Turn CW to increase...
Page 6
OtoPet-USA Earigator Manual Earigator Trouble Shooting LOW IRRIGATION PRESSURE: LOW SUCTION: Before any checks, ensure that the middle insuf- With the machine on, and regulator fully counter flation knob is turned fully clockwise. With the ma- clockwise (for maximum suction), and the suction...