Environmental Warnings
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
iCOquit® monitor. Otherwise, degradation of the performance of this equipment could
result.
The iCOquit® monitor should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the iCOquit® monitor should be observed to verify
normal operation. If operation is not normal, the iCOquit® or the other equipment should be
moved.
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Avoid exposure to known sources of EMI (electromagnetic interference) such as Magnetic
Resonance Imaging (MRI) systems, diathermy, lithotripsy, electrocautery, RFID (Radio
Frequency Identification), and electromagnetic security systems such as metal detectors.
Keep the iCOquit® outside the MRI scanner room.
Note that the presence of RFID devices may not be obvious. If such interference is suspected
reorient equipment if possible, to maximize distances.
30