Faith FTP9000 Series User Manual

Programmable dc power supply

Advertisement

Quick Links

Programmable DC Power Supply
User's Manual
(FTP9000 Series)
©Copyright Faithtech
Version V1.03
2021-06

Advertisement

Table of Contents
loading

Summary of Contents for Faith FTP9000 Series

  • Page 1 Programmable DC Power Supply User’s Manual (FTP9000 Series) ©Copyright Faithtech Version V1.03 2021-06...
  • Page 3 Please visit our website http://www.faithtech.cn for latest information. Product Warranty Faithtech promises that specifications and features of FTP9000 Series power supplies have completely met all technical indexes claimed in this manual. We strictly check raw materials and manufacturing technology to ensure the stability and credulity of products.
  • Page 4 FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ Safety Abstract Please observe important safety instructions strictly in the process of operating and repairing the power supply. Not observing safety precautions or warnings in this manual will probably weaken protective functions provided by the power supply.
  • Page 5 ____________________________________________________________________________ ---------------------------------------------------------------------------------------------------------------------- The WARNING sign highlights an essential operating or maintenance procedure, practice, condition, statement, etc., which if not strictly observed, could result in injury to, or death of personnel or long term heath hazards. ---------------------------------------------------------------------------------------------------------------------- The CAUTION sign highlights an essential operating or maintenance procedure, practice, condition, statement, etc., which if not strictly observed, could result in damage to, or destruction of equipment.
  • Page 6: Table Of Contents

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ Table Of Contents 1. GENERAL DESCRIPTION............................- 6 - 1.1 Overview..............................- 6 - 1.2 Main features............................- 6 - 1.3 Dimension diagram..........................- 7 - 1.3.1 Model of 5kW, 10kW, 15kW.........................- 7 - 1.3.2 Model of 20kW, 25kW, 30kW........................
  • Page 7 ____________________________________________________________________________ 3.7.4 Output Mode & Status Monitoring....................- 31 - 3.8 Protection..............................- 32 - 3.9 Save And Recall........................... - 33 - 3.10 Power-down Save..........................- 34 - 3.11 Keyboard Lock And Unlock......................- 35 - 3.12 Sequence (SEQ) Function......................- 35 - 3.13 Internal Resistance Simulation (RESI).................
  • Page 8: General Description

    1. General Description 1.1 Overview FTP9000 series high-power programmable DC power supply has the characteristics of high power density, high power factor, high efficiency, and wide range of output. Its wide range voltage / current output and automatic constant power function can greatly increase the application coverage. FTP9000 series DC power supply output power 5kW up to 180kW in a single unit, and expandable up to 1800kW when paralleling 10 identical models.
  • Page 9: Dimension Diagram

    ____________________________________________________________________________ 1.3 Dimension diagram 1.3.1 Model of 5kW, 10kW, 15kW - 7 -...
  • Page 10: Model Of 20Kw, 25Kw, 30Kw

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 1.3.2 Model of 20kW, 25kW, 30kW - 8 -...
  • Page 11: Model Naming

    ____________________________________________________________________________ 1.4 Model Naming 1.5 Specification sheet General specification Voltage rise slew rate Maximum slew rate 6000V/s Voltage fall time No load <2s Full load ≤30ms Transient response Typical 2ms Parallel Parallel up to 10 identical units through master-slave mode for max 1800W output Protection OVP, OCP, OPP, OTP, LVP etc..
  • Page 12 FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ Model FTP9150-80-510 FTP9150-200-210 FTP9150-400-120 FTP9150-500-90 FTP9150-800-75 0~80V 0~200V 0~400V 0~500V 0~800V Voltage 0~510A 0~210A 0~120A 0~90A 0~75A Current 0~15kW Power Model FTP9300-80-1020 FTP9300-200-420 FTP9300-400-240 FTP9300-500-180 FTP9300-800-150 Voltage 0~80V 0~200V 0~400V 0~500V 0~800V...
  • Page 13 ____________________________________________________________________________ Accuracy 0.1%+0.2% F.S. Power measurement① Accuracy 1%F.S. Ripple noise④ 160mV 300mV 550mV 450mV 800mV Ripple Vpp 10kW 320mV 300mV 550mV 450mV 800mV 15kW 320mV 300mV 550mV 450mV 800mV 16mV 40mV 65mV 70mV 200mV Ripple Vrms 10kW 25mV 40mV 65mV 70mV 200mV 15kW...
  • Page 14 FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 0~300A 0~320A 0~240A 0~200A Current 0~100kW 0~120kW Power Voltage programming① Resolution 16bits Accuracy 0.05%+0.1%F.S. Power programming① Resolution 16bits Accuracy 0.1%+0.2% F.S. Power programming① Accuracy 1%F.S. External analog programming① Control voltage 0~10V corresponds to 0~100%F.S.
  • Page 15: Quick Guide

    ____________________________________________________________________________ 2. Quick Guide 2.1 Checking Goods Upon reception of the power supply, please examine the equipment according to procedures stipulated as below. (1) Check if there is damage during transportation. Contact authorized dealer or Faithtech after-sale service immediately if the packing or protective cushion is severely damaged.
  • Page 16: Front Panel

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 2.2 Front Panel ① Power switch ② TFT color screen ③ Keyboard ④ Rotary knob 2.3 Keyboard Figure 2-2 FTP9000 keyboard Table 2-2 Key Function Descriptions Button Key Name Description Voltage...
  • Page 17 ____________________________________________________________________________ Power Switch to output setting mode from other functions, and Power is Setting Key now settable; Advanced For performing Sequence Test, Voltage RAMP, Test Key RESI internal resistance simulation, etc. SAVE Key Save the settings of the static output to internal memory. Recall Key Recall the saved settings from internal memory.
  • Page 18: Screen Display

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 2.4 Screen Display Figure 2-3 MAIN PAGE Display ① Specification info of the power supply; ② Readback voltage, current, etc. parameters display area; ③ Status prompt area; ④ Parameters setting area;...
  • Page 19: Rear Panel

    ____________________________________________________________________________ 2.5 Rear Panel Figure 2-4 FTP9000 Rear Panel Item Name Description Share Master-Slave parallel ports This connector compensates for any voltage drop generated by the cable resistance. Be sure to connect the V SENSE remote sense connector “+” to the positive output terminal and the “–”...
  • Page 20: Installing

    (3) For indoor use only, ambient temperature 0°C to 40°C. 2.6.2 Requirements of Input Power Please pay attention to the following items for FTP9000 series power supply input connection: ① Three-phase input: 340VAC~460VAC, 47~63Hz; ② Ensure reliable grounding;...
  • Page 21: Connection

    ____________________________________________________________________________ 2.7 Connection 2.7.1 Input Connection The user connects the three-phase wire and the ground wire to the power supply input terminal as shown in Figure 2-6. For the size of the wire diameter, please refer to "Table 6-1 Recommended Wire Diameter for Cables". Figure 2-6 Power supply Input Connection ---------------------------------------------------------------------------------------------------------------------------- 1.
  • Page 22: Output Connection

    ---------------------------------------------------------------------------------------------------------------------- 2.7.4 Remote Sensing Connection FTP9000 series power supplies are able to compensate for the voltage drop caused by output connecting cables. Use two wires to connect the positive/negative connectors of the load to the corresponding positive/negative remote sensing connector on the rear panel, symbol "Sense" will be prompted in the status prompt area of the screen if the power supply output is turned on.
  • Page 23: Pro-Int Analog Interface

    ____________________________________________________________________________ 2.7.5 PRO-INT Analog Interface FTP9000 series power supply provides optional PRO-INT analog Interface at its rear panel. This interface has the functions of analog programming input, voltage/current monitoring output, external control, power supply status and work mode indication. Figure 2-11 PRO-INT Analog Interface...
  • Page 24: Function Introduction

    3. Function Introduction This chapter describes the main functions and characteristics of the power supply. Reading this chapter, you will have a deeper understanding of FTP9000 series power supplies. 3.1 Local & remote operation mode The power supply provides two operation modes: local operation and remote operation. In the local operation mode, the user mainly operates through the keyboard and knob on the front panel, and checks the power status through the LCD display;...
  • Page 25 ____________________________________________________________________________ Data transmission speed, choices: 1000Hz/500Hz/200Hz/100Hz/10Hz, Samp Rate default value 10Hz Prio SLT Choice for CV or CC priority. Default value: Volt (CV Priority) Voltage rise slew rate, range: 0.0006V/s~ slew-UP 6000V/s, default value: 50% of max value App Set Voltage fall slew rate, range: 0.0006V/s~ slew-DN 6000V/s, default value: 50% of max value...
  • Page 26: System Menu

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 3.2.2 System Menu Press to enter main menu, then select “System” and press to pop up sub menu as below: Table 3-2 System Menu Layout Default 192.168.1.123 S-Mask Default 255.255.255.0 Baud Serial port baud rate, 4800/9600/19200/38400/115200, default: 9600.
  • Page 27: Static Output Function

    ____________________________________________________________________________ 3.3 Static Output Function FTP9000 series power supply provides auto ranging wide range output within its rated power. The output range is 0 ~ 102% rated output of voltage, current and power. Press , the power supply enters static output function, the screen displays...
  • Page 28: Power Supply Output Modes

    3.3.2 Limit Voltage/Current/Power Set FTP9000 series power supply allows for setting upper and lower limits on voltage, current and power. When the upper and lower limit is set to non-zero, the setting range of static output voltage, current, and power parameters is limited between the upper and lower limit.
  • Page 29: Voltage Slew Rate And Current Slew Rate

    3.5 CV/CC Priority Start function FTP9000 series power supply provides CC priority and CV priority start function, which forces the power supply to operate in CC or CV mode at the moment the output is turned on, effectively avoids the current or voltage overshoot resulted from capacitive or inductive load.
  • Page 30: Output Control

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ Operation: 1. Press 2. Select “Set” >> “App Set”, press “Enter” to enter the application setting page shown as below: System Edit About Samp Rate 10Hz Prio SLT Volt 5000.0 V/s slew-UP 5000.0 V/s...
  • Page 31 ____________________________________________________________________________ Analog programming control speed is 1000 points per second, external 0 ~ 5V DC voltage corresponds to 0 ~ 100%F.S. output voltage, current, power. Figure 3-7 Input Programming Signal And Output Relation Operation: 1. Press to enter the main menu. 2.
  • Page 32: External Control

    ____________________________________________________________________________ 3.7.2 External Control FTP9000 series power supply analog interface provides an input “INHIBIT” for trigger signal, as shown below: Figure 3-9 Trigger Signal Connection This “INHIBIT” port receives 5V TTL level signal, low level is effective. You can use a physical switch such as "Figure 3-9 Trigger Signal Connection"...
  • Page 33: Voltage, Current Monitoring Output

    Figure 3-11 Output Voltage, Current And Monitoring Output 3.7.4 Output Mode & Status Monitoring The FTP9000 series power supply has an output mode signal port and a fault indication signal port. The port outputs a 5V TTL level signal, and its functions are as follows: Output mode signal port, low level indicates that the power supply is working in CV mode, and high level indicates that the power supply is working in CC or CP mode;...
  • Page 34: Protection

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 3.8 Protection FTP9000 series power supply provides protections against overvoltage, overcurrent, overpower, low voltage, output reverse voltage, communication timed out, overtemperature. When a protection occurs, the power supply will turn off output, sounds an alarm, and prompt corresponding protection information on display.
  • Page 35: Save And Recall

    ____________________________________________________________________________  Hardware overpower protection (OP): It is triggered once the output power exceeds 110% rated output power, and “OP” is prompt on display.  Software overpower protection (OPP): It is programmable as shown in “Figure 3-12 Protection Set”. Users are able to set the OPP threshold, set to 0 means to turn off OPP.
  • Page 36: Power-Down Save

    10. 3.10 Power-down Save FTP9000 series provides power-down save function. The power supply will resume last time working parameters at its power on moment when power-down save function is activated, otherwise the power supply will initialize using default parameters when it is turned on.
  • Page 37: Keyboard Lock And Unlock

    ____________________________________________________________________________ 3.11 Keyboard Lock And Unlock Press to lock the keyboard. When keyboard locked, only ON/OFF button and “Shift+9” operations are valid, the other key operations are invalid. At the same time, the parameter setting area in MAIN PAGE will shown in gray color to differentiate with normal unlocked state.
  • Page 38 FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ System Edit About SEQ Step File Length 100 V_Set 0.000 V Cycle I_Set 0.000 A Link Step-Time 86400.0 s Figure 3-13 Sequence File Edit 4. Set “Length”, press to validate it, the editing cursor will move to “Cycle” automatically;...
  • Page 39: Internal Resistance Simulation (Resi)

    Since a battery has internal resistance, its output voltage will decrease along with the increasing output current. FTP9000 series power supply provides internal resistance simulation feature to imitate the output characteristics of batteries. With this function, the power supply output voltage satisfy below formula: : Actual output voltage;...
  • Page 40: Voltage Ramp-Up, Ramp-Down (Ramp)

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 80V/510A/15000W RESI 0.0000 V_MAX 0.000 V 0.0000 I_MAX 0.000 A 0.000 P_MAX Ω 0.000 W 80V/510A/15000W RESI 0.0000 Resistance 0.000 Ω 0.0000 0.000 Ω Figure 3-16 Output Test For Battery Simulating 3.
  • Page 41 ____________________________________________________________________________ FTP9000 series power supply allows for output voltage ramp-up and ramp-down according to the set slew rate. Operation: 1. Press to enter test modes selection interface; 2. Use “Rotary” knob to select “RAMP”, press , the screen displays as below:...
  • Page 42: Master-Slave Parallel

    3.15 Master-slave Parallel 3.15.1 Master-slave parallel connection FTP9000 series power supplies provide a master-slave parallel mode, so that users can achieve greater output power by paralleling same models. The user uses proper wires to connect the "Share" interface in the rear panel, the "RS485"...
  • Page 43: Master-Slave Parallel Setting

    ____________________________________________________________________________ Figure 3-19 Master-Slave Parallel Connection Master-slave parallel cable requirements It is OK to use cross-sectional area 0.5 mm² ~ 1 mm² double strand wire for RS485 and “Share” ports. ---------------------------------------------------------------------------------------- 1. Master-slave parallel support identical power supplies only; 2. Max 10 units; 3.
  • Page 44: Master-Slave Parallel Function Operations

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ System Edit About M_or_S Master S.NUM MS_CTRL Figure 3-20 Master-Slave Parallel Set 1. Press to enter main menu interface; 2. Select “Set” >> “Parallel”, then press to display master-salve parallel set page;...
  • Page 45 ____________________________________________________________________________ 80V/510A/15000W V.I.P 0.0000 0.0000 Slave1 Figure 3-22 Slave display When a protection occurred in the slave unit, please press to clear fault in corresponding slave unit and in master unit. When protection occurred in master unit only, just press on master unit to clear fault.
  • Page 46: Communication Introduction

    ____________________________________________________________________________ 4. Communication Introduction FTP9000 series power supply provides users with 5 types of communication interfaces: standard USB (serial), LAN (Ethernet), optional GPIB or CAN or RS485 interfaces. FTP9000 supports standard Modbus, SCPI protocol, default protocol is ModBus. Users are able to configure communication ports and protocols base on requirements.
  • Page 47: Voltage, Current Acquisition Speed

    ____________________________________________________________________________ 4.2 Voltage, current acquisition speed After the above communication configuration is completed, if the user needs to increase the voltage, current acquisition speed, the configuration can be as follows: 1. Press to enter main menu page; 2. Select “Set” >> “App Set”, press to show as below: System Edit...
  • Page 48: Lan

    FTP9000 Series Programmable DC Power Supply Operation Manual ____________________________________________________________________________ 4.3.2 LAN The FTP9000 series power supply has an Ethernet communication interface and adopts UDP communication mode. The default IP address is 192.168.1.123, the default subnet mask is 255.255.255.0, and the default port number is 7000.
  • Page 49: Communication Protocol

    ____________________________________________________________________________ 4.4 Communication Protocol FTP9000 series power supplies support SCPI, Modbus and CANopen communication protocols. For LAN, USB serial port, RS485 interface, users can choose SCPI or Modbus; for GPIB interface, only SCPI protocol is supported, and CAN interface only supports CANopen protocol.
  • Page 50: Troubleshooting

    Follow the actions described in this chapter to inspect the instrument and troubleshoot the problem if the FTP9000 series DC Power Supply does not operate normally. Contact Faithtech authorized agent or Faithtech after-sales if the information provided in this manual does not resolve the problem.
  • Page 51: Appendix

    ____________________________________________________________________________ 6. Appendix Table 6-1 Recommended Wire Diameters for Cables Cross-sectional Type Different temperature conditions area 60 ℃ 75 ℃ 85 ℃ 90 ℃ Wire type: Wire type: Wire type: Wire type: RUW, T, UF RHW, RH V, MI TA, TBS, SA, AV Rated Current (A) 2.08 3.31...