Accuracy - HOPETECH HT3530 User Manual

Insulation resistance meter
Table of Contents

Advertisement

8.2 Accuracy

Test conditions for the following indicators:
Temperature: 20±3℃; Humidity:<80%RH;
Preheating more than 15 minutes
Calibration within 1 year
HT3530/HT3530A:
Insulation resistance: take 11 points typical voltage value,
accuracy 1%(less than 1 G); 3%(1 G -10 G); 5%(10 G -100 G);
10%(100 points -1 T); 20%(more than 1 T).
Range
1
Voltage
1V
0.5K-50K
10V
5K-500K
25V
12.5K-1.2M 125K-12.5M 1.2M-125M 12.5M-1.2G 125M-12.5G 1.2G-125G 12.5G-2500T
50V
25K-2.5M 250K-25M 2.5M-250M 25M-2.5G
75V
37.5K-3.8M 375K-37.5M 3.8M-375M 37.5M-3.8G 375M-37.5G 3.8G-375G 37.5G-7500T
100V
50K-5M
125V
62.5K-6.2M 625K-62.5M 6.2M-625M 62.5M-6.2G 625M-62.5G 6.2G-625G 62.5G-12500T
250V
125K-12.5M 1.2M-125M 12.5M-1.2G 125M-12.5G 1.2G-125G 12.5G-1.2T 125G-25000T
500V
250K-25M 2.5M-250M 25M-2.5G
750V
375K-37.5M 3.8M-375M 37.5M-3.7G 375M-37.5G 3.7G-375G 37.5G-3.8T 375G-75000T
1000V
500K-50M 5M-500M
2
3
5K-500K
50K-5M
50K-5M
500K-50M 5M-500M
500K-50M 5M-500M
50M-5G
4
5
500K-50M
5M-500M 50M-5G
50M-5G
250M-25G 2.5G-250G 25G-5000T
50M-5G
500M-50G 5G-500G 50G-10000T
250M-25G
2.5G-250G 25G-2.5T 250G-50000T
500M-50G
5G-500G
53
6
7
500M-100T
500M-50G 5G-1000T
50G-5T
500G-10000T

Advertisement