ERVICE ANUAL Bair Hugger Series 200 and 500 Warming Units Bair Hugger, Total Temperature Management, 241, and Augustine Medical/light bulb logo/bright ideas...that work, are trademarks of Augustine Medical Inc., registered or pending in the U.S. and other countries.
Page 2
Replacement parts to correct a problem are delivered at no charge. To return a device to Augustine Medical, Inc. for service, first obtain a Return Authorization (RA) number from a technical service representative. Please use this number on all correspondence when returning a device for service.
Hugger blanket before providing skin surface warming therapy. Do not use the 241 fluid warming set with any other forced air warming system other than the 500 Series Bair Hugger system. Fluid temperature outside the indicated range may result. Augustine Medical, Inc.
The blanket’s clear plastic drape may cause visual distortion. Lift the plastic drape to view the patient’s head clearly. Read Instructions See CONTRAINDICATIONS and WARNINGS before administering therapy. Read the Operation Manual, blanket instructions and fluid warming package instructions before use. Augustine Medical, Inc.
(described in this Service Manual), the Over Heat Alarm system, as well as a leakage current test. Proper Use and Maintenance Augustine Medical, Inc. assumes no responsibility for the reliability, performance, or safety of the equipment if: •...
Secure the Phillips-head hose anchor screw. The screw is self-tapping into the hose end. Record the maintenance action taken. Extended length Figure 2. Location of Hose Anchor Screw. Wiring shield Hour meter Hose anchor screw Blower outlet Augustine Medical, Inc.
Use the small screwdriver to remove the fuse carrier. Remove the blown fuse(s) and replace with the new ones. Replace the fuse carrier. Reattach the power cord and the two (2) Phillips head screws that secure the cord retainer. Augustine Medical, Inc.
Page 11
Phillips screws. Unscrew the fuse holder cap, remove the old fuse, and insert the new fuse. Reattach the wiring shield. After replacing the fuses, reconnect the warming unit to the power source and record the maintenance action taken. Augustine Medical, Inc.
1. To replace the fuse, remove the wiring shield by unscrewing the three Phillips screws. 2. Unscrew the fuse holder cap, remove the old fuse, and insert the new fuse. 3. Reattach the wiring shield. Augustine Medical, Inc.
Close the rear access panel and resecure the four (4) Phillips-head screws. Reconnect the warming unit to the power source. Record the maintenance action taken. Figure 6. Location of Air Filter Airflow direction Air filter Augustine Medical, Inc.
Perform all testing and adjustments of warming unit temperature with a calibrated Model 221 Temperature Test Kit or a Model 22110 Temperature Test Unit. Augustine Medical, Inc. assumes no responsibility for the reliability, safety, or performance of the Bair Hugger system if warming unit temperature testing or adjustments are made in any other manner than that described here.
Page 15
• Do not block the rear air vents on the test device. • The temperature setting will not affect the over heat test because the control circuit has been bypassed. Figure 8. Temperature Test Devices Model 22110 Port Port Augustine Medical, Inc.
Page 16
The thermostat interrupts power to the heater and activates the alarm at a preset high temperature of 56° ± 3°C (132.8° ± 5.4° F), or less, at the end of the hose, as determined by the Augustine Medical temperature test device. Model 275 High Temperature Thermostat The thermostat interrupts power to the heater and activates the alarm at a preset high temperature of 53°...
The meter it is connected to must be calibrated. Follow the calibration instructions in the meter service manual or follow standard facility calibration schedules. • When in doubt, contact Augustine Medical Customer Service for assistance in temperature adjustment. Tools/Equipment •...
Page 18
Find the over heat alarm thermostat (see Figure 10). Figure 10. Locating the Thermostat Set screw Over heat alarm thermostat Thermostat connections Set screw Thermostat connections screw Augustine Medical, Inc.
Page 19
P3, if applicable). Put the filter back into place and close the rear access panel. 10. Replace the four (4) Phillips-head screws on the rear access panel. 11. Record the maintenance action taken. Augustine Medical, Inc.
Perform all testing and adjustments of warming unit temperature with a calibrated Model 221 Temperature Test Kit or a Model 22110 Temperature Test Unit. Augustine Medical, Inc. assumes no responsibility for the reliability, safety, or performance of the Bair Hugger system if warming unit temperature testing or adjustments are made in any other manner than that described here.
Page 21
(allow the warming unit to operate at each temperature setting for about three (3) to five (5) minutes before each test). Turn the warming unit OFF. Remove the hose end from the port on the temperature test device. Record the maintenance action taken. Augustine Medical, Inc.
If the HIGH range is properly set, and MED and/or LOW do not meet specs, call Technical Service. • Recalibrate the Temperature Test Kit at least once per year. • When in doubt, contact Augustine Medical Customer Service for assistance in temperature adjustment. Tools/Equipment •...
Page 23
If the HIGH range is properly set, and MED and/or LOW do not meet specs, call Augustine Medical Technical Service. Turn the warming unit power OFF, and remove the hose from the port on the temperature test device.
53° ± 3°C, 127.4° ± 5.4°F, or less, at the end of the hose, as Model 200 determined by an Augustine Medical, Inc. Model 125VAC, 60Hz, 7 or 9.5 Amperes temperature test device.
(replacement only) Standard length with swivel (replacement only) Extended length with swivel (note keyway) Standard length 241 fluid warming hose (replacement for 500 series only) Extended length 241 fluid warming hose (note keyway- standard for 500 series only) Augustine Medical, Inc.
Unscrew the fuse holder cap, remove the old fuse, and insert the new fuse. Replace the fuse holder cap, filter, wing nut and service cover. Figure 2. Accessing the fuses Filter wing nut Service cover Filter Filter seat 10A fuses .062A 10A or 15A fuse fuse Augustine Medical, Inc.
Page 30
The unit will not run unless the interlock switch is properly seated. Record the maintenance action taken. Connect the warming unit to the power source and turn it on to verify that the filter interlock is properly seated. Augustine Medical, Inc.
Remove the fuse holder. Place the new fuse into the fuse holder. For knob-type fuse holders: Push the fuse holder in with your fingers and turn 90° counterclockwise. Remove the fuse holder. Place the new fuse into the fuse holder. Augustine Medical, Inc.
Page 32
To access the control board, remove the lower filter seat. If the filter seat is one piece, remove the screws holding it in place. Figure 5. Accessing the control board fuse Filter Service lid Power head Lower filter seat Control board Augustine Medical, Inc.
Page 33
Replace the fuse with the same type. Reattach the lower filter seat. Reattach the power head. Reattach the filter (make sure the interlock switch is engaged) and reattach the service lid. Figure 6. Watlow A007 control board Fuse Augustine Medical, Inc.
Connect the warming unit to the power source and turn it ON to verify that the filter interlock is properly seated. Figure 7. Replacing the Air Filter Service lid Air filter Power head Power head plug connectors Filter seat Power head plug connectors Augustine Medical, Inc.
Perform all testing and adjustments of warming unit temperature with a calibrated Model 221 Temperature Test Kit or a Model 22110 Temperature Test Unit. Augustine Medical, Inc. assumes no responsibility for the reliability, safety, or performance of the Bair Hugger System if warming unit temperature testing or adjustments are made in any other manner than that described here.
Page 36
To access the control board, remove the lower filter seat. If the filter seat is one piece, remove the screws holding it in place Figure 8. Accessing the Control Board Service lid Power head Lower filter seat Control board Augustine Medical, Inc.
Page 37
50°C to 56°C (122°F to 132°F) at the end of a hose, as determined by the Augustine Medical temperature test device. Figure 9. Locating TP3 and TP4 on the Watlow A007 control board TP3 and TP4 Augustine Medical, Inc.
Page 38
P-3 on the control board. 10. Replace the filter seat and reassemble the power head. 11. Replace the filter. Make sure the interlock switch and the service lid are engaged. Figure 10. Locating P-3 on the Series 80 control board Augustine Medical, Inc.
K style thermocouple (e.g., a Fluke Model 52 K/J Thermometer). • If the connector does not fit your meter, you can remove the provided connector and attach a connector that fits your meter. Augustine Medical, Inc.
Page 40
Adjusting the Over Heat Alarm Thermostat Locate the over heat adjustment thermostat (see Figure 11). Figure 11. Location of Thermostat in Model 500 Units Bulb type thermostat Set screw Bimetallic thermostat Set screw Interlock switch Screw Bulb type thermostat Control board Augustine Medical, Inc.
Page 41
(or reattach the P3 lead, if applicable). Replace the filter seat and reassemble the power head. 10. Replace the filter (make sure the interlock switch is engaged) and the service lid. 11. Record the maintenance action taken. Augustine Medical, Inc.
Perform all testing and adjustments of warming unit temperature with a calibrated Model 221 Temperature Test Kit or a Model 22110 Temperature Test Unit. Augustine Medical, Inc. assumes no responsibility for the reliability, safety, or performance of the Bair Hugger system if warming unit temperature testing or adjustments are made in any other manner than that described here.
Page 43
Figure 13. Room Temperature v.s. High Temperature Output Warming unit temperature setting should be on HIGH when conducting this test 15 16 17 18 19 20 21 22 23 24 25 26 27 Ambient Room Temperature (°C) Ambient Room Temperature (°F) Augustine Medical, Inc.
Units with the Series 80 control board Watlow A007 control board Temperature Adjustment Potentiometer (Labeled “HI”) Temperature Adjustment Potentiometer WARNING: Use the “HI” temperature adjustment potentiometer only. If temperature is still out of range, please call Technical Service. Augustine Medical, Inc.
Page 45
• If the tested operating temperature was below specified limits, turn the screwdriver CLOCKWISE to increase the operating temperature. • If the tested operating temperature was above specified limits, turn the screwdriver COUNTERCLOCKWISE to lower the operating temperature. Augustine Medical, Inc.
Page 46
If the HIGH range is properly set, and MED and/or LOW do not meet specs, call Augustine Medical Customer Service. 14. Turn the warming unit power OFF, and remove the hose from the port on the temperature test device.
100VAC, 60Hz, 9.5 Amperes the alarm at a preset high temperature of 53°C ±3° (122°F to 132°F), at the end of the hose, as determined by an Augustine Medical, Inc. temperature test device. Temperature Characteristics Air temperatures reaching the patient are approximately 2°C lower than the listed temperatures.
Returning Units for Service Call Technical Service (see inside front cover for phone number) to get a Return Authorization (RA) number and a service carton. Using a marking pen, write your Return Authorization number on the outside of the carton. Augustine Medical, Inc.
Ensure that P-3 is connected or that Warming unit heats to alarm TP3 and TP4 are not shorted condition on all settings Thermostat is out of calibration Recalibrate over temperature set point Runaway heat failure Call Technical Service Augustine Medical, Inc.
Maintenance Log Date Maintenance Action Performed Hour Meter Serial No. Augustine Medical, Inc.
Page 54
TEL 952-947-1200 • 800-733-7775 • FAX 952-947-1400 • 800-775-0002 www.augustinemedical.com Augustine Medical’s Authorized Representative in the European Community (as defined in Article 14 of the Medical Device Directive: 93/42/EEC): Actamed Limited, Calder Island Way, Wakefield, WF2 7AW, United Kingdom. TEL (44) 1924 200550 • FAX (44) 1924 200518 Watlow is a registered trademark of Watlow Electric Manufacturing Company, St.