VDW nuve DirectFREEZE DF 290 Instruction Manual

Ultra low temperature deep freezer

Advertisement

Quick Links

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş.
DF 290 / DF 490 / DF 590
DirectFREEZE™
ULTRA LOW TEMPERATURE
DEEP FREEZER
INSTRUCTION MANUAL
1984
Z14.K 25 276
Rev.No: 08
Rev. Date: 01/2018

Advertisement

Table of Contents
loading

Summary of Contents for VDW nuve DirectFREEZE DF 290

  • Page 1 NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. DF 290 / DF 490 / DF 590 DirectFREEZE™ ULTRA LOW TEMPERATURE DEEP FREEZER INSTRUCTION MANUAL 1984 Z14.K 25 276 Rev.No: 08 Rev. Date: 01/2018...
  • Page 2 Dear Nüve User, We would like to take this opportunity to thank you for preferring this Nüve product. Please read the operating instructions carefully and keep them handy for future reference. Please detain the packing material until you see that the unit is in good condition and it is operating properly.
  • Page 3 WARRANTY CERTIFICATE 1. Nüve warrants that the equipment delivered is free from defects in material and workmanship. This warranty is given for a period of two years. The warranty period begins from the delivery date. 2. Warranty does not apply to parts normally consumed during operation or general maintenance or any adjustments described in the operating instructions provided with the instrument.
  • Page 4: Table Of Contents

    CONTENT Page INTRODUCTION ..........................5 1.1. USE AND FUNCTION ......................... 5 TECHNICAL SPECIFICATIONS ......................6 2.1. TECHNICAL SPECIFICATIONS TABLE ................... 6 2.2. ACCESSORIES ..........................7 2.2.1. FACTORY FITTED ACCESSORIES ..................7 2.2.2. OPTIONAL ACCESSORIES ....................7 PRECAUTIONS AND LIMITATIONS ON USE ..................7 SYMBOLS AND LABELS ........................
  • Page 5: Introduction

    INTRODUCTION 1.1. USE AND FUNCTION DF series Ultra Low Temperature Freezer is designed for the storage of the samples needed to store at ultra-low temperature and it is used in the hospitals, blood banks, epidemic diseases research centers, research laboratories in the universities, genetic engineering and bio- pharmaceutical industry.
  • Page 6: Technical Specifications

    DF Series Freezers are designed and manufactured in accordance with national and international directives, EN 50419 and EN 61000-6-3 standards under the supervision of total quality management systems ISO 9001 and ISO 13485. This device is in compliance with WEEE Regulation. If the warnings mentioned in this manual are not considered, NUVE will not be responsible from their results.
  • Page 7: Accessories

    2.2. ACCESSORIES 2.2.1. FACTORY FITTED ACCESSORIES DF XXX Y Weekly temperature chart recorder 0°C / -100°C 2.2.2. OPTIONAL ACCESSORIES A 08 191 AlerText™ GSM alarm module K 13 009 NuveWarn™ Remote alarm system with 10 m cable A 08 195 NuveCloser™...
  • Page 8: Symbols And Labels

    SYMBOLS AND LABELS Symbol in the operating instructions: Attention, general hazard area. This symbol refers to safety relevant warnings and indicates possibly dangerous situations. The non-adherence to these warnings can lead to material damage and injury to personal.  Symbol in the operating instructions: This symbol refers to important circumstances.
  • Page 9: Installation

    INSTALLATION ENVIRONMENTAL CONDITIONS The instrument is designed to operate safely under the following conditions:  Indoor use only  Ambient temperature: 5°C to 30°C  Maximum relative humidity for temperature up to 31°C: 80 %  Maximum altitude: 2000 m ...
  • Page 10: Handling And Transportation

    HANDLING AND TRANSPORTATION All handling and transportation must be carried out by using proper equipment and experienced staff. The instrument must be supported underneath, moved in vertical position and never be turned upside down. UNPACKING Remove the cardboard box packing and ensure that no damage has occurred during transportation.
  • Page 11: General Presentation

    GENERAL PRESENTATION DOOR HANDLE CONTROL PANEL CHART RECORDER CONDENSER (OPTIONAL) FILTER Figure 1 CO2 BACK-UP USB PORT-1 OUTPUT POWER LED USB PORT-2 FUSES ETHERNET POWER CORD RS 232 POWER KEY NC-CONTACT OUTPUT REMOTE ALARM OUTPUT Figure 2...
  • Page 12: Chamber Dimensions

    5.6.1. CHAMBER DIMENSIONS DF 290 * DF 490 DF 590 575 mm 610 mm 715 mm 310 mm 310 mm 310 mm 310 mm 310 mm 310 mm 310 mm 310 mm 310 mm 310 mm 310 mm 1070 mm 965 mm 1120 mm 420 mm...
  • Page 13 This symbol denotes the graph screen. The last 24 hours of operation of the device can be monitored on the graph of the temperature vs. time by pushing the button corresponding to this symbol. It is recorded every 10 minutes. This symbol denotes the alarm range settings.
  • Page 14: Prior To Operation

    PRIOR TO OPERATION  Plug-in the instrument to correctly grounded sockets.  Insert the key to the power switch lock at the back.  Power the freezer on by rotating the key clockwise.  Observe that the microprocessor control system is activated. ...
  • Page 15: Settings

    Password screen will appear when the MENU button (F1) is pushed. You can enter the password by using the value increase (F1) and decrease keys (F2). The password is user’s password and can be changed on settings page (See section 6.1.2.). The password is 0000 for the first usage.
  • Page 16 “Select Language” submenu is selected on the settings page by using the value increase (F1) and decrease (F2) keys. The page on the left appears when enter key (F4) is pushed. The language of the control panel can be set as Turkish, English, French, Russian or Spanish.
  • Page 17: Memory

    6.1.3. MEMORY “Memory” submenu is selected on the menu page by the value increase (F1) and decrease (F2) keys. The page on the left appears when enter key (F4) is pushed. INTERNAL MEMORY “Failure History” submenu is selected on the memory page by the value increase (F1) and decrease (F2) keys.
  • Page 18: External Memory (Usb Stick)

     When the temperature sensor is broken, “–“appears in the temperature column since temperature cannot be read by the sensor. “History (GRAPH)” submenu is selected on the memory page by the value increase (F1) and decrease (F2) keys. “Select Record Date” page appears when enter key (F4) is pushed. Enter a date to analyze corresponding operation and then push the enter key (F4).
  • Page 19: Sms

    Copy Failure History: As USB stick is connected to the device, all error history in the internal memory is transferred to USB stick when “Copy Failure History” is selected. When loading is finished completely, push the enter key (F4) to exit from this page. 6.1.4.
  • Page 20: E-Mail

     Tab key (F3) is used for changeover from one digit of phone number to another.  After all phone numbers from “Phone 1” to “Service 2” is entered as mentioned above, push the enter key (F4) and return to main page. 6.1.5.
  • Page 21: Set Temperature

    “Login” is selected on the e-mail page by the value increase (F1) and decrease (F2) keys. The page on the left appears when enter key (F4) is pushed. Each character of e-mail addresses is written one by one. The first character is entered by pushing value increase (F1) and decrease keys (F2) and tab key (F3) is pushed to enter the next character.
  • Page 22: Periodic Maintenance And Cleaning

    When the temperature alarm setting key (F4) is pushed in the operation mode, the “Alarm Range” page shown on the left appears. The desired alarm value can be adjusted by pushing value increase (F1) and decrease (F2) keys. Enter key (F4) is pushed to save the value and returned to the working screen.
  • Page 23: Disposal Management Concept

    DISPOSAL MANAGEMENT CONCEPT The currently valid local regulations governing disposal must be observed. It is in the responsibility of the user to arrange proper disposal of the individual components. All parts which may comprise potentially infectious materials have to be disinfected by suitable validated procedures (autoclaving, chemical treatment) prior to disposal.
  • Page 24: Fuse Replacement

    A list consisting of the codes of errors can be accessed from the memory page (See  section 6.1.3.1). ‘EoFF’, ‘Err3’, ‘Err4’, ‘Err5’ and ‘Err6’ do not appear on working screen and they are only shown in the list of the error history. If an error occurs, please contact to an authorized NÜVE agent to seek technical help.
  • Page 25: Nüvewarn ™ Remote Alarm System

    SIM card which will be used for GSM module should not have PIN code.  Plug-in the GSM module to correctly grounded sockets.  Connect the end of RS 232 cable of GSM module (See Figure 6) to the RS 232 port on the ultra-low freezer (See Figure 2).
  • Page 26: Chart Recorder

    CHART RECORDER The below mentioned are provided with the chart recorder, please check them;  100 ea. circular diagram paper for chart recorder  1 ea. pen for chart recorder  1 ea. battery  2 ea. circular recorder keys. Chart recorder measures the temperature in the chamber and records this value by using a pen.
  • Page 27 Figure 11...
  • Page 28: Electrical Circuit Diagram

    ELECTRICAL CIRCUIT DIAGRAM DF 290, DF 490 AND DF 590 ELECTRICAL CIRCUIT DIAGRAM...