Infinity
Instruction Manual
Toll-Free Customer Service
1-877-970-4970
Save this manual for future reference
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Provant Infinity and is the answer not in the manual?

Questions and answers

Summary of Contents for Regenesis Biomedical Provant Infinity

  • Page 1 Infinity Instruction Manual Toll-Free Customer Service 1-877-970-4970 Save this manual for future reference...
  • Page 3: Table Of Contents

    Electromagnetic Compatibility (EMC) User Information System Technical Specifications Copyright 2021 Regenesis Biomedical, Inc. All rights reserved. Regenesis, Regenesis Biomedical, the Energy Starburst logo, and the Provant Infinity trade name and logo are registered trademarks of Regenesis Biomedical, Inc., Scottsdale, AZ.
  • Page 4: Indications For Use

    Indications for Use The Provant Infinity is indicated for the adjunctive use (as part of another therapy) in the palliative treatment of postoperative pain and edema of soft tissue. Contraindications (Reasons not to use this device) • If you have a pacemaker or defibrillator.
  • Page 5: Advisories

    • DO NOT USE this device in oxygen-rich environments. • DO NOT SUBMERGE this device in water or other liquids. • DO NOT dispose of the Provant Infinity device. Please contact Regenesis Biomedical at 1-877-490-4970 to return the device. • This nonthermal shortwave therapy (SWT) device generates PEMF energy to produce clinical effects.
  • Page 6: Product Description

    4) Applicator Cable 5) A/C Power Supply 6) A/C Power Cable 7) POWER Button 8) Treatment Button 9) LCD Screen 10) Indicator LED 11) Carrying case (not pictured) If any items are missing or appear damaged, please contact Regenesis Biomedical at 1-877-970-4970...
  • Page 7: General Information

    There is no need to remove these items for treatment. • DO NOT stand on the treatment applicator pad. • Provant Infinity is an electrical device; access to a three-prong electric outlet is required for use. Use the A/C power supply provided with this device.
  • Page 8: Instructions For Use

    1c. Check all device components for damage and if any items are missing or appear damaged, contact Regenesis Biomedical at 1-877-970-4970. 1d. Choose a location for your treatment and place the Base Unit of the device in a convenient, stable location (e.g., side table, floor, etc.).
  • Page 9: Starting A Treatment Session

    2b. POSITIONING THE TREATMENT APPLICATOR PAD • Find a comfortable position sitting or lying down. o If you are sitting or lying on a metallic surface, make sure comfortable padding (e.g., a mattress or cushion) is between the Treatment Applicator Pad and the metal, so as not to interfere with the delivery of the therapeutic PEMF energy.
  • Page 10: Pausing Or Interrupting A Treatment Session

    4. PAUSING OR INTERRUPTING A TREATMENT SESSION 4a. Pause treatment: Press “Pause” button on the Control Panel. • The indicator light will flash orange. • A 20-minute Pause Remaining Timer will display on the top center of the LCD screen, indicating how much time is left for the treatment to be paused.
  • Page 11: Cleaning

    • DO NOT SUBMERGE this device in water or other liquids. • Before cleaning, unplug the device from the electrical outlet. • Provant Infinity can be cleaned using a damp cloth moistened with 70% rubbing alcohol. Do not use cleaning products containing bleach.
  • Page 12: Troubleshooting

    • Do not use an alternate A/C Power Supply. If A/C Power Supply is required, replace with Regenesis A/C Power Supply P/N 315-1005-01. • If the problem persists, call Regenesis Biomedical at 1-877-970-4970. SERVICE REQUIRED MESSAGE: • Please call Regenesis Biomedical at 1-877-970-4970.
  • Page 13: Servicing

    • Products not covered under the Limited Warranty may also require repairs. Repairs outside of warranty will be charged for the retail cost of parts, plus labor and shipping charges. • Regenesis Biomedical recommends routine maintenance every two years. For any maintenance related information, call Regenesis Biomedical at 1-877-970-4970.
  • Page 14: Summary Of Clinical Performance

    Reports received at a rate of 0.1% or less generally related to preexisting pain, the underlying condition being treated, or to the surgical procedures that preceded the use of Provant. Regenesis Biomedical does not consider these latter reports as being associated with use of Provant.
  • Page 15: Electromagnetic Compatibility (Emc) User Information

    • Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Provant Infinity, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • Page 19: System Technical Specifications

    System Technical Specifications...
  • Page 20 250-0014-00 E Protected under US patents, 6,334,069; 6,353,763; 6,967,281; 6,974,961; 7,024,239; 7,572,985 and 7,579,555. Additional patents are pending. Provant, Regenesis, Regenesis Biomedical, SAFE-T-ENERGETICS, the Provant yellow, and the Energy Starburst Logo are registered trademarks of Regenesis Biomedical, Inc., Scottsdale, AZ.