READ THESE INSTRUCTIONS BEFORE USING THE EQUIPMENT
111-0063-00 (Doc 556M)
INSTRUCTIONS FOR USE
OXYGEN DEMAND VALVE
Page 1 of 8
18/03/2021
PAGE
2
2
2
3
4
5
5
6
6
6
6
6
8
ISSUE H

Advertisement

Table of Contents
loading

Summary of Contents for Meditech OXYGEN DEMAND VALVE

  • Page 1: Table Of Contents

    INSTRUCTIONS FOR USE OXYGEN DEMAND VALVE PAGE 1. SYMBOLS 2. INTENDED USE OF DEVICE 3. TECHNICAL DESCRIPTION 4. WARNINGS 5. INSTRUCTIONS FOR USE 6. CLEANING AND DISINFECTION 7. MAINTENANCE 8. SERIAL NUMBER 9. PRODUCT LIFE SPAN 10. SPECIFICATION 11. SPARE PARTS 12.
  • Page 2: Symbols

    Next Maintenance or Service Due date 2.INTENDED USE OF DEVICE The B.N.O.S. Meditech Oxygen Demand Valve is intended for the self-administration of Medical Oxygen. Oxygen should only be used for short-term/temporary therapy by self-administration of the Demand Valve user and under supervision of suitably qualified medical personnel. Please refer to any rules or restrictions that may cover the use of Oxygen in the specific country of use of the device.
  • Page 3 The demand valve has been specifically designed to used with low cost single use generic mouth piece and filter systems that are widely available using standard connections that conform to BS EN ISO 5356-1. 111-0063-00 (Doc 556M) Page 3 of 8 ISSUE H 18/03/2021...
  • Page 4: Warnings

    WARNINGS! 4.1 Please read these instructions before use. Do not use the demand valve if you do not understand the instructions detailed in this user guide. Do not use any form of grease or oil (hydrocarbon based substances) with this device. Do not use this device near any source of ignition e.g.
  • Page 5 5.1 Before connecting to a pressurised gas source, please ensure the unit is clean and in good condition. If you have any doubts about the condition, please do not connect or use the device. Please check that the inhalation diaphragm is fitted to the device and seated correctly (laying flat against the body of the device).
  • Page 6: Cleaning And Disinfection

    Demand Valve should be thoroughly dried before storage. 7. MAINTENANCE 7.1 Maintenance must be carried out on the unit on a five yearly basis by B.N.O.S. Meditech or engineers certified by B.N.O.S. Meditech. This activity involves dismantling the unit and replacing all internal seals, diaphragm, medical gas supply hose and any components which show significant wear and tear.
  • Page 7: 12. Relevant Applicable Standards

    The responsibility and cost of returning and collecting the unit from the manufacturer or their authorised representative is the owners. B.N.O.S Meditech Ltd reserves the right to change design without prior notice. Any disassembly of the device beyond that detailed in this manual will invalidate the warranty and the manufacturers disclaim any liability for products that have undergone unauthorised repair.
  • Page 8 2797 COMPANY CONTACT DETAILS This device is designed and manufactured by: B.N.O.S. Meditech Ltd. 9 Fifth Avenue, Bluebridge Ind. Est., Halstead, Essex CO9 2SZ, England Tel: +44 (0)1787 479475, Fax: +44 (0)1787 477747 E-mail: sales@meditech.uk.com www.meditech.uk.com EC Representative Medical Device Management...