Intended use: The Econocare Nebuliser is designed to deliver the prescribed medicated solution to assist patients with the treatment of respiratory disorders such as asthma, allergies and bronchitis. Nebulising converts the medicated solution into an aerosolized mist which is inhaled by the patient through the mouthpiece or mask.
Description: 1. Air outlet connector 4. Power cord 7. Air inlet connector 2. Filter cap 5. T-mouthpiece 8. Tubing 3. Power switch 6. Nebuliser cup 9. Foam filters Power switch Air Outlet Connector Air Openings Nebuliser Cup Holder Filter & Cover Accessories: Nebuliser machine Air tube...
Operating Instructions: 1. Place the Econocare Nebuliser on a stable, sturdy and flat surface so that the unit can be easily reached when you are seated. 2. Pull the power cord out from the base of the unit, and make sure that the unit is in the “off”...
Do NOT operate without a filter. Warranty: We warrant the Econocare nebuliser to be free from defective workmanship and materials for a period of (1) year from the date of purchase. Check with your retailer for more details.
Page 7
Do not untwist the jet nebuliser during operation. Follow your medical practitioner’s instructions on type and dosage of medication required. Do not pour more than 6ml of solution into the jet nebuliser. DANGER: TO REDUCE THE RISK OF ELECTROCUTION. Always unplug the unit immediately after using. Do not use while bathing.
Aerosol Output: **0.33 ml (2ml,1%NaF) Aerosol Output Rate: **0.06ml/min (2ml,1%NaF) Econocare nebuliser the particle size distribution curve as well: Result of cascade impactor **measurements for particle size with Econocare nebuliser and nebuliser kits MMAD = Mass Median Aerodynamic Diameter oklet_REV_042020.indd 10...
Important information regarding Electro Magnetic Compatibility (EMC) The Econocare Nebuliser needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided. This unit can be affected by portable and mobile RF communications equipment.
Page 10
Guidance and manufacture’s declaration – electromagnetic immunity This is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Electrostatic dis- ±6 kV contact ±6 kV contact...
Page 11
Guidance and manufacture’s declaration – electromagnetic immunity This is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should...
Page 12
Trou Chec Recommended separation distances between portable and mobile RF communications equipment and the nebuliser. This is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the nebuliser as recommended below, according to the maximum output power of the communications equipment.
Note: If the suggested remedy does not solve the problem, do not try to repair the device. No parts of the unit are user serviceable. Return the unit to an authorized Econocare retail outlet or distributor. oklet_REV_042020.indd 15...
Disposal: Correct Disposal of this product (Waste Electrical & Electronic Equipment) This marking shown on the product or its literature, indicates that it should not be disposed of with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.