Opti-Med JM-LED02 User Manual

Led shadowless lamp

Advertisement

Quick Links

Opti-Med
®
LED Shadowless Lamp
JM-LED02
USER MANUAL
Hangzhou Jupiter Medical Equipment Co., Ltd
1116, Gudun Road, West Lake District, Hangzhou, China
Tel: +86-15021634329
Web: www.opti-med.net
Email: info@opti-med.net

Advertisement

Table of Contents
loading

Summary of Contents for Opti-Med JM-LED02

  • Page 1 Opti-Med ® LED Shadowless Lamp JM-LED02 USER MANUAL Hangzhou Jupiter Medical Equipment Co., Ltd 1116, Gudun Road, West Lake District, Hangzhou, China Tel: +86-15021634329 Web: www.opti-med.net Email: info@opti-med.net...
  • Page 2: Expected Use

    Opti-Med ® 1. Expected Use This product is used in various surgical lighting in hospital operating room. It can also be use in medical lighting situations in emergency room. 2. Conditions of Use 2.1 Normal working conditions Temperature: 10℃~30℃; Relative humidity: ≤70%;...
  • Page 3: Simplified Diagram

    Opti-Med ® 4. Simplified Diagram rotati ng arm decorati ve l ayer bal ance arm 2000 ±45 ° bal ance arm 3000 360 ° ±45 ° 360 ° 360 ° assi st ant 360 ° l am p 300 °...
  • Page 4: Installation

    Opti-Med ® Installation The equipment is fixedly mounted on the roof. The floor must bear the weight of the device,but also bear the force moment because of partial load and the operation. In order to ensure the safety, the floor should have enough bearing capacity and strength.
  • Page 5 Opti-Med ® thi ckness: 8m m thi ckness: 12m m sup erfi ci al treatm ent:anti rust l acq uer sup erfi ci al treatm ent:anti rust l acq uer p i cture 4: p ressure p l ate p i cture 5: m o unti ng d i sk...
  • Page 6 Opti-Med ® Mounted diagram www.opti-med.net...
  • Page 7 Opti-Med ® Mounted steps 1. according to picture 1 and picture 2 in the operation instruction, drill four Ф16mm holes on the floor plate with impact drill in determined place. pressure plate and mounting disk must be tightened with screw M12 in the appropriate length, according to picture 3.
  • Page 8 Opti-Med ® Position adjustment Pull the handhold on the lamp in order to the position and the point which meet operators’ requirement. Illumination adjustment Press the power switch, the lamp lights. each time you press the button “ increasing brightness”, the illumination will rise a gear. When up to the maximum 10 gear, the maximum illumination symbols will lights;...
  • Page 9 Opti-Med ® Damping adjustable rotating arm The center seat of 2 rotating arms has 2 damping adjustment screws. If it drifts after using for some time, you can tighten the damp with wrench according to picture 8. picture 9 Lamp head lamp adjustment Lamp head and two brackets should have appropriate damping, to avoid transposition.
  • Page 10: Warranty

    Opti-Med ® Precautions when maintaining (A) the product is not anti-inlet device. It can not be cleaned with water when cleaning. You can wipe it with wet cloth containing neutral detergent solution, then wipe it with wet cloth containing clear water for several times. It also can be wiped with wet cloth containing disinfectant.