Q-Sense E4 Flow System Setup, with IPC-N Pump
Part 1 – Ismatec IPC-N Pump
Before switching the pump on,
check the voltage setting and fuse
on the backside:
220-240V: use 250mA fuses
110-120V: use 500mA fuses
Keep the "SETTINGS" button
pressed while switching the pump
on. This makes the pump start in
settings mode (the blue words
above the buttons are active).
Cycle through the settings, using
the up/down arrows, until "TUBE" is
shown, select by pressing "OK".
Select inner diameter 0.64 mm,
(double-check with numbers given
on the tubing itself!). Press "OK" to
confirm.
Press "ESCAPE" to save settings.
Attach the adaptors on each
cassette, and then the tubing.
Ready to go!
Typical flow rates: for filling the
system 100-400 μl/min; for
experiment 25-100 μl/min.
For details and calibration function, see supplier's manual.
1
Voltage settings
window
2
3
4
Q-SENSE E4 LIQUID HANDLING SET UP
50 μl/min
100 μl/min
Dec 2007, Rev B