Page 2
Mouth Rejuvenation Faster: Teeth whitening performed in 15 to 30 minutes, compared to one hour on average with current treatments. Safer: Eliminate the potential risks from abuse or poor application during patient home follow-up treatments, and prevent hydrogen peroxide associated side effects. Brighter: Effectively whiten even the very dark or light teeth shades, known to be difficult to treat with present methods.
Page 3
Technology The Y10 technology provides the treatment results at the level of professional treatment. The treatment can be repeated as needed. The use of application of RF electric field having certain time and frequency pattern (function of time and/or frequency) to unique Y10 toothpaste...
Page 4
Important Guidelines for the Treatment / 1 • Please make sure to slightly tilt the patient’s head to the back as described in the following sketch: • Please make sure the patient doesn’t bite too hard on the mouthpiece or chews it.
Page 5
Important Guidelines for the Treatment / 2 • Please make sure the toothpaste was refrigerated for at least an hour prior to the treatment. The toothpaste should fill out the entire electrode cavity of the mouthpiece from both sides, to reach the following •...
Page 6
Important Guidelines for the Treatment / 3 Do not use the Y10 device in case of the following: • The patient has an implanted pacemaker, an internal defibrillator or any other active electrical implants anywhere in the body. • The patient has one of the following: a bleeding disorder, leukemia, cirrhosis, vascular disease, sarcoidosis, active hepatitis or HIV.
Page 7
Preparation Step Please make sure to carefully read the operator manual prior to the first use. The following items are required for Y10 operation:...
Page 8
Step 1 - Connecting the Y10 System Connect the applicator cable to the appropriate connection port. Power Cable Then proceed to connect the system’s power cable to the clinic’s power supply. Activation Key Finally, connect the activation key to the...
Page 9
Step 2 - Initializing the Y10 System To initialize the devise, please press the button at the bottom of the screen. After about 20 seconds, the main menu will appear. On / Stand by Button...
Page 10
Step 3 System Licensing Approach your BrightTonix product NOTE: The BrightTonix system is configured distributor to receive the license key. to operate with a To update the system’s license key: current license for a • Press the on the screen, the pre-determined time.
Page 11
Step 4 Setting the System Up To customize the system operations, press the button on the screen. The system Configuration screen will appear. Then, press the General button- the Secondary Setup screen will appear. NOTE: To view the on screen version of the operator manual, press the ‘Operator’s Manual’ button.
Page 12
Step 5 Screen Navigation Press the button at any Press the ‘Return’ button at time to navigate to the any time to navigate to the main menu screen. previous screen.
Page 13
Step 6 Mouthpiece preparation Connect a new, disposable mouthpiece to the applicator by inserting the tab on the mouthpiece into the slot on the top of the applicator. Make sure to connect a mouthpiece to the distal end of the umbilical. NOTE: Prior to connecting the mouthpiece, please make sure there are no electrodes sticking out of it.
Page 14
Step 7 Treatment Customization Press the ‘Treatment’ button in the screen to customize the system operations. The Main treatment screen will appear.
Page 15
Step 8 Shade Selection Select the client’s current teeth shade and tap the corresponding shade number to highlight it. VITA Bleachedguide 3D-MASTER NOTE: Hold the shade guide approximately 10-50cm from the open mouth of the client to determine the shade.
Step 9 Preset Selection Select a treatment Preset: Low energy treatment, 15-minute duration (automatically selected as default). Medium energy treatment, 20-minute duration. High energy treatment, 30-minute duration.
Page 17
Step 10 Toothpaste Application A-Apply the Y10 toothpaste to the mouthpiece generously on the bottom side of the mouthpiece. B-While placing the mouthpiece in the patient’s mouth, apply the toothpaste to the top side of the mouthpiece. A-Bottom side B-Top side...
Page 18
Step 11 Preparation Press the ‘Next’ button, the Treatment screen will appear.
Page 19
Step 12 Performing the Treatment Tap the applicator NOTE: If the customer reports discomfort, button to commence at any time you may treatment, a press the ‘Pause’ percentage progress button to interrupt the treatment. To continue, dial will count from press ‘Resume’.
Page 20
Step 13 Treatment Termination Once the percentage progress dial reaches 100%, the treatment light will extinguish and the screen will be displayed. Press the button at the top of the mouthpiece to safely remove the applicator from the mouthpiece. Remove the mouthpiece from the patient's mouth and allow him to wash his mouth with water.
Page 21
Step 14 Final Shade Selection Select the final shade achieved at the end of the treatment and press the button to reset the system for the following customer. VITA Bleachedguide 3D-MASTER NOTE: Leaving the system without any action for NOTE: Hold the shade guide approximately more than five (5) minutes 10-50cm from the open mouth of the will switch the system to...