YIMI Life YM201 User Manual

YIMI Life YM201 User Manual

Finger pulse oximeter

Advertisement

Pulse Oximeter User Manual
Product Model:YM201/YM301
Version: 2.0
Date: 2020-07-08

Advertisement

Table of Contents
loading

Summary of Contents for YIMI Life YM201

  • Page 1 Pulse Oximeter User Manual Product Model:YM201/YM301 Version: 2.0 Date: 2020-07-08...
  • Page 2: Front View

    1. Product Introduction and Operation Guide 1.1 Front View Figure 1 Front View of YM201/YM301 1.2 Operation Method A. Open the battery cover, and put the two AAA batteries into the battery compartment in correct polarities, then replace the cover;...
  • Page 3: Lanyard Installation

    ----------------------------------------------------------------------------------------------------- WARNINGS:  Battery polarities should be correctly installed, otherwise, damage may be caused to the equipment.  Please remove the batteries if the equipment will not use for a long time. ----------------------------------------------------------------------------------------------------- Figure 2 Battery Installation 1.4 Lanyard installation A.
  • Page 4: Attention For Operation

    1.5 Attention for Operation A. Before use check and confirm that the people or finger size were applicable; B. Before use check and confirm that the environment should be non- combustible material, as well as to avoid high or low temperature and humidity, but also need to pay attention to the following: a) To avoid glare and direct sunlight exposure;...
  • Page 5 Figure4 Finger Placement Diagram 1.6 Function and menu operation After turning on the oximeter, press and hold the power button for about 2 seconds. The oximeter will call up the parameter setting interface and set it by pressing the button. Defined here, long- press indicates that the button hold time reaches 1-2s, shor t-press indicates that the button hold time is less than 0.5s.
  • Page 6 during the measurement, and when Beep is set to off, no sound is output.  While the "*" symbol stays on the Restore option, hold the button to restore factory settings.  Press the button to select a Brightness level ranging from 1 to 5. The greater the value, the greater the brightness of the screen.
  • Page 7: Specification

    2. Specification 2.1 Classification Type of protection against electric shock..Ⅱ (Internally powered equipment) Degree of protection against electric shock......Type BF-Applied part Operating mode .................Spot checking Degree of protection against hazards of explosion........P22 2.2 Power Requirements Specification of alkaline batteries...........Two AAA (LR03) Operating current..................25-50mA 2.3 Physical Specifications Width*Height*Depth..............57×30×31 mm...
  • Page 8 Atmosphere Pressure Operating..................70~106kpa Storage/Transportation...............50~107.4kpa 2.6 Display Display Color ................YM201: 0.96",Yellow&Blue YM301: 1.3",Blue Display content: SpO2%, Pulse Rate, PI%, Bar Graph Battery Indicator,Pulse Wave Notes: 1) The claim for oxygen saturation accuracy should be supported by clinical studies covering the entire claimed range,The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steady-state saturation periods over the specified SpO2 accuracy range (e.g.
  • Page 9: Maintenance, Cleaning, Disinfection

    2) The clinical trial included 11 subjects, including 6 males and 5 females, with an age range of 18 to 46 years, the subjects skin color included dark black, medium black, light color and white. 3. Maintenance, Cleaning, Disinfection 3.1 Maintenance The equipment's design life expectancy is about 2 years, keep your equipment and accessories free of dust and dirt, and follow these rules: A.
  • Page 10 3.2 Cleaning Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital's regulations for cleaning the equipment. Recommended cleaning agents are: a) Mild soap (diluted).
  • Page 11 3.3 Disinfection Clean the pulse oximeter before disinfecting it. The recommend disinfectant is ethanol 70%. Disinfection step are the same as cleaning. ----------------------------------------------------------------------------------------------------- CAUTION  Never use ETO or formaldehyde for disinfection. ----------------------------------------------------------------------------------------------------- 3.4 Disposal Dispose of the pulse oximeter in accordance with local environment and waste disposal laws and regulations.
  • Page 12: Troubleshooting

    5. Troubleshooting Trouble Possible Reason Solution The battery is drained away or almost Please replace batteries. drained away. equipment can’t be The battery installation is incorrect. Install the battery over again. turned on. The device works abnormally. Please contact the product distributor. The Spo2 Select the suitable size finger to The finger size is too big or small.
  • Page 13 Guidance and manufacturer´s declaration – electromagnetic emission The model YM201/YM301 is intended for use in the electromagnetic environment specified below. The customer or the user of the model YM201/YM301 should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
  • Page 14 Guidance and manufacturer´s declaration – electromagnetic immunity The Model YM201/YM301 are intended for use in the electromagnetic environment specified below. The customer or the user of the Model YM201/YM301 should assure that it is used in such an environment. IEC 60601...
  • Page 15 NOTE: UT is the a. c. mains voltage prior to application of the test level. Guidance and manufacturer’s declaration – electromagnetic immunity The YM201/YM301 is intended for use in the electromagnetic environment specified below. The customer or the user of the YM201/YM301 should assure that it is used in such an environment. Immunity...
  • Page 16 80 equipment should be used no closer to any part of the Models YM201/YM301, including 6 Vrms 150 cables, than the recommended separation 61000-4-6 kHz to 80 distance calculated from equation MHz outside applicable to the frequency of the transmitter.
  • Page 17 NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a. The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz;...
  • Page 18 RF compliance level above, the YM201should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the YM201/YM301. d. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
  • Page 19 EC REPRESENTATIVE Shanghai International Holding corp. Gmbh(Europe) Address: Eiffestrasse 80, 20537, Hamburg, Germany Email: shholding@hotmail.com M-PO-Y1E 0520 EN Certificates...
  • Page 20 Finger Pulse Oximeter Date:...