Bone Injection Gun Reloading Device RBIG007
USER MANUAL
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | mail@waismed.com | www.waismed.com
Rev 3 28122010

Advertisement

loading

Summary of Contents for PerSys Medical Waismed BiG

  • Page 1 Bone Injection Gun Reloading Device RBIG007 USER MANUAL 11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | mail@waismed.com | www.waismed.com Rev 3 28122010...
  • Page 2: Notes And Precautions

    Dear customer, Waismed is doing its best to supply you with the top quality products at the highest quality levels. We hope you will find the Reloading System to your satisfaction. Should you encounter any problem or inquiry please contact us or your local dealer.
  • Page 3 In order to reload the BIG you will need the following parts: Rear Body II. Piston III. Washer (Ring) IV. Spring Needle VI. Safety Latch VII. Front Body 11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | mail@waismed.com | www.waismed.com Rev 3 28122010...
  • Page 4 Position the Reloading Device with Waismed sticker facing you. The handles should be in Position Remove the cap at the top of the Reloading Device. Twisting it counter clockwise and pull it up. 11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | mail@waismed.com | www.waismed.com Rev 3 28122010...
  • Page 5 Move the Washer (Ring) (III) on top of the Piston (II) to the small groove on the Needle- Gripping Prongs. Match the Piston (II) to the small Pin inside the Reloading Device. 11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | mail@waismed.com | www.waismed.com Rev 3 28122010...
  • Page 6 Slide the spring (IV) over the Piston (II). Slide the Rear Body of the B.I.G. (I) over the Piston (II) and Spring (IV); make sure that the Bolder Raised Arrow matches the groove on the back of the device and the Shallower Arrow is pointing at the red line on the front of the device.
  • Page 7 Move the handles to Position 1. With your fingers, press the top of the BIG and press it downwards. You should feel the spring’s resistance. While holding the handles in Position 1, close the cap by pushing it downwards and twisting it clockwise until the two red string holders match.
  • Page 8 Lower the handles to Position 2 and press it downwards until a click sound is heard. The handles base and the device base should touch one another. Do not lean on the far side of the handles as they might break. Raise the handles to Position 3.
  • Page 9 Lower the red clip on the cap’s back with your finger until you can insert the Safety Latch (IV). Insert the safety latch (VI). Make sure the safety latch (VI) is firmly in place and could be seen through the small hole on the other side of the cap.
  • Page 10 Remove the Cap by twisting it counter-clockwise and pulling it up. Remove the BIG from the cap and place the Needle (V) hub into the Needle Gripping Prongs, until the Needle (V) is firmly and securely held in place. (A click sound will be heard).
  • Page 11 Close the Front Body (VII) of the BIG by screwing it onto the Rear Body (I). Never take out the safety Latch (VI) when the device is loaded. Never point the BIG and/or the needle at a person. For further information, instructional movies, presentations and article please go to www.waismed.com LOOKING FOR A TRAINING? Just go to www.waismed.com and click the...
  • Page 12 11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | mail@waismed.com | www.waismed.com Rev 3 28122010...