Cleaning kit

Advertisement

Instructions for Use
Boule Cleaning Kit Medonic
Medonic CA620 and CA530
Intended use
The Boule Cleaning Kit is intended for use with the Medonic
Hematology Systems CA620 and CA530.
different cleaning solutions as indicated on their labels. Using this
cleaning kit at the stated regular intervals will ensure:
-
Longer life time of instrument tubing
-
Lower bacterial grow risk
-
Lower background counts
-
Increased mean time between failure (higher MTBF)
Contents
1.
Enzymatic Cleaner, 450 mL, blue-colored solution
2.
Hypochlorite Cleaner; 450 mL, yellow-colored solution
3.
Detergent Cleaner, 450 mL, red-colored solution
The contents can be used for three full and complete rinsing cycles.
Remaining cleaning solution should be disposed of.
Storage and stability
The Boule Cleaning kit can be transported at ambient temperature
(avoid freezing, allow adaption to room temperature before use).
Always store the bottles in an upright position to avoid any leakage
from the cap.
Cleaning interval
It is recommended the following time intervals:
Less than 50 samples/day →
More than 50 samples/day →
100-200 samples/day
Procedure
It is most important that the cleaning procedure is performed in the
correct order (or it may result in high PLT background count).
The following order must be followed:
1)
Enzymatic Cleaner (enzyme solution - blue cap)
2)
Hypochlorite (2% solution - white cap)
3)
Detergent Cleaner (detergent solution - red cap)
Proceed as follows:
1.
Firstly, remove the diluent and lyse probes from their
containers. (Note. Do NOT use any of the Cleaning Kit
solutions in the Detergent inlet. The Detergent probe shall
be left in its container during this whole procedure.)
Go to menu 8.3 'Empty System' and press <ENTER>.
2.
Wait for the system to empty.
3.
Wet a soft tissue with the proper cleaning solution (e.g. the
first time use Enzymatic Cleaner) and wipe both level
sensors clean on the outside. Then insert both level sensors
directly into the cleaning solution bottle.
Go to menu 8.2 'Fill System' followed by <ENTER>.
4.
Allow the system to fill up with the cleaning solution and
wait 15-30 minutes for the cleaner to work.
Remove both probes and go to menu 8.3 'Empty System'
5.
and press <ENTER>. Allow the system to empty
completely.
Boule Medical AB
Domnarvsgatan 4
SE-163 53 Spånga, Sweden
E-mail: info@boule.com
www.boule.com
11452-4
2019-06-18
The kit contains 3
every 6 months
every 3 months
every month
6.
Repeat step 3 to 5 using the Hypochlorite (white) and then
followed by using the Detergent cleaner (red).
7.
Finally take a soft tissue and wet it with diluent (not lyse)
and wipe off any remains of the cleaners from the probes.
When probes are clean insert both back into the original
reagent containers. (Be sure not to interchange the diluent
and lyse probes.)
Go to menu 8.2 'Fill System' followed by <ENTER> and
8.
wait until this cycle is finished.
Go to menu 6.2 'HGB photometer adjustment' and press
9.
digit '1' on the keyboard for the photometer to
automatically adjust and wait until this cycle is finished.
10. The background (blank) count should not exceed the
recommended level as stated in the user's manual. Perform
a minimum of 3 background (blank) counts before running
the next sample.
Notes
-
As the volume in the cleaning bottles will decrease during
the cleaning cycles, be sure to secure the bottles to avoid
tilting.
-
During the described cleaning process above, some
foaming might occur on the aspiration pipette; remove
such with a soft tissue.
-
In case of heavy contaminated systems, it is recommended
to warm the detergent to 45-55 ° C prior to cleaning the
system.
-
Only the Hypochlorite Cleaner has the ability to act as a
disinfectant, significantly reducing the number of viable
microorganisms.
-
It is recommended to check calibration by running a
control after cleaning.
Precautions
-
Boule Hypochlorite Cleaner (2%) contains sodium
hypochlorite (2%). The solution is skin irritating and
causes serious eye irritation. It is marked with a hazard
pictogram.
The other two solutions do not contain
chemicals at a level considered to be hazardous to health.
However, all solutions should be handled with proper care,
and skin and eyes should be protected. Please read the SDS
before use. Since hypochlorite solutions are oxidizing with
a basic pH, avoid contact with materials other than those
indicated in the instructions (spills give a remaining smell
on e.g. skin and stains (bleaching) on clothes etc).
-
The concentration of hypochlorite decreases with time
without invalidation of the functionality within the given
expiry. Close the container as soon as possible after use.
-
Cleaners should only be used before stated expiration date.
The cleaners should be disposed of according to local
regulations. When draining through the sewage system,
always use plenty of water.
ORDERING INFORMATION AND SERVICE
Contact your local distributor for orders and support. Please have the
product number ready for orders. For other assistance contact Boule
Medical AB at phone +46 8 7447700, fax +46 8 7447720 or
info@boule.com
For translation of this instruction and explanation to symbols see
www.medonic.se/support/.
Ordering no:
Description
1504111
Boule Cleaning Kit
Packaging
3 x 450 ml
1504093

Advertisement

loading

This manual is also suitable for:

Medonic m32Medonic ca620Medonic ca530