Quality Assurance / Contacts at Thermo Haake Quality Assurance Dear customer, Thermo Haake implements a Quality Management System certified according to EN 29001. This guarantees the presence of organizational structures which are necessary to ensure that our products are devel- oped, manufactured and managed according to our custom- ers expectations.
Page 4
Repairs, alterations or modifications must only be car- ried out by specialist personnel. Considerable damage can be caused by improper repairs. The Thermo Haake service department is at your disposal for repair work. Do not operate the unit with wet or oily hands.
1 Warranty card (please fill out and return) EK20 and EK30 Immersion coolers: 1 Set of mounting elements (2 platings, 2 screws) for fixing the heat exchanger to the Thermo Haake bath bridge H32, 1 Instruction manual, 1 Warranty card (please return)
Information concerning the CE sign 3.3 Information concerning the CE sign Thermo Haake measuring and control instruments carry the CE sign which confirms that they are compatible with the EU guideline 89/336/EEC (electromagnetic compatibility). The tests are carried out according to module H (official sheet L380 of the European Community) as our quality assurance system is certified according to DIN / ISO 9001.
Flow-Through Cooler DK15 The DK15 is designed to be integrated within the heat trans- fer liquid circuit of e.g. a heating circulator. The heat is re- moved from the heat transfer liquid as it is pumped through Heating / Open-bath circulator the cooler.
Thermo Haake baths with a volume of up to approx. 20 liters are cooled to below –5°C. The graphs in chapter 8 show tem- peratures and cooling times using Thermo Haake circula- tors.
Unit Description Immersion cooler EK20 and EK30 applications The external evaporator (cooling coil) is immersed into a bath and the coil must be completely immersed within the liq- uid. The lowest temperature which can be reached in a bath de- pends on the following factors: •...
Connecting the Hoses (only for the DK15) Connecting the Hoses (only for flow-through cooler DK15) Inlet Hose from external object Outlet Hose to return nozzle on circulator The hose connections pieces for 8 or 12 mm ∅ B and the unions E supplied with the unit are screwed onto the threaded nozzles A (7 + 8).
Thermo Haake Bath Bridge Slide the supplied metal plates into the slot above the cooling coil. Remove the covering plate on the Thermo Haake bath bridge by unscrewing two screws. Insert the evaporator through the opening and secure the fitted plates with 4 screws.
The flow-through cooler must only be used with heat transfer liquids with a freezing point below –30°C. If the heat transfer liquid were to freeze in the heat exchanger of the DK15, the resulting volume expansion could cause the heat exchanger to burst.
10. Accessories / Technical Specifications Accessories Order No. Clamp mounting for fixing the EK20 or EK30 to bath walls up to a thickness of 25 mm 333–0452 Technical specs. acc. to DIN 58966 DK15 EK20 EK30 °C Working temperature range –15..150 –20..150 –30..150 Cooling capacity at +20°C...
Accessories / Technical Specifications 10.1 Fuse values Unit type Mains voltage Fuse(s) at the rear panel 230V 2x5 A DK15 115V 1x10 A EK20 230V 2x5 A 115V 1x10 A 230V 2x5 A EK30 115V 1x10 A Subject to alterations Printed in Germany (FRG) 002–4348...