Technical Operating IFU
• Connection of head: Insert the attachment
Gentlefile-drive
head [4] onto
and push firmly to complete connection.
• Insertion of File: Insert the file into the
hole in the attachment head [4] by gently
pushing the file into the head. (grip by
friction power only)
•
Gentlefile drive operation:
•
1. Place the file in the canal and press the
on-off button [3] briefly.
•
2. Verify that the light indicator [2] is on,
and not blinking. The
will continue until the on-off button [3] is
pressed again briefly.
• 3.
Before extracting the file from the canal,
push the on-off button [3] again to stop
operation.
Charging & Maintenance
• Connect the charger's plug to the unit's
socket [1], located on the back lid of the
unit [2]. When the battery is full, charging
will
stop
automatically
indication on the charger will appear).
When charging is completed, unplug the
unit from the charger.
• Note: It is recommended to charge the
device prior to each procedure.
• Note: To prolong battery lifetime, it is
recommended to disconnect the device
from the charger after 4-5 hours.
• Alerts: When the battery charge is low, the
indicator light [2] on the back lid of the unit
will blink.
Control and Safety
• The Gentlefile drive includes inherent
safety mechanisms.
1. Motor Power control - when resistance
is encountered, the motor's rotation
speed declines, reaching a full stop in the
continue presence of high resistance.
2. Motor Load Control - When the resistance
in the canal rises, controller firmware
detects the increase in load and limits the
motor power to a safe predefined value.
3. After stoppage of the file (should that
occur), the file may be released by
manually pulling it clock-wise.
[5]
, joint
Gentlefile-drive
[5]
(Green
light
Drive
Single use
➍
Gentlefile - Drive
➎
➋
IFU
attachment head
GF-CAP-01
GF-DRV-02
Chahrging socket
➊
➌
Need help?
Do you have a question about the GF-CAP-01 and is the answer not in the manual?