1. Introduction 1.1 The Infrared therapeutic lamp features a plate coated with a proprietary mineral formulation consisting of 33 elements. When activated by a built-in electric heating element, this mineral plate emits a special band of electro-magnetic waves ranging from 2 to 22 microns in wavelength and 25 mw/sq.
• Please read the instruction manual before use and use only according to this manual. As the installation and operation for this device are very simple,the intended operator only need to follow this Introduction Manual to operate,not necessary for professional training. •...
people responds to the treatment in different ways. We recommend you experiment with the device to discover the best way in which it can help you. The treatments described are for guidance only. For more specific problems consult your doctor. Timings are approximate and are intended to be flexible to suit your own comfort.
Page 6
● The Device can easily be damaged by improper handling. ● To safely turn off the device, reset the timer to “OFF” position (for models CQ-12, CQ27 & CQ29) or press “PAUSE” on the panel of the timer (for models CQ-33, CQ-36). Switch off the power switch to the "OFF"...
●Daily maintenance needs to pay attention to the service life of element board and heating plate. 3.6Warning for possible exposure to hazardous optical radiations ●To avoid possible exposure hazardous radiation,during treatment,both eyes should be avoided direct exposure to the lamp.If it’s necessary to irradiate tothe eyes,wear sunglasses or cover it with cotton fabric.
5. Package and assembling of device Contentsof the “Model CQ-12, CQ-27, CQ-29, CQ-33, CQ-36” IR Therapeutic Lamps Model:CQ-36 Model:CQ-12 Model:CQ-27 Model:CQ-29 Model:CQ-32...
Page 9
Instruction Manual 1 pc The device and its accessories are packed in a strong paper box. Model Long (cm) Wide (cm) High(cm) CQ-12(2 In 1) CQ-27 CQ-29 CQ-33 CQ-36 A) Eachshipped device is individual packed into a 3-ply Carton Box.
Page 10
700hPa to 1060hPa Model CQ-XXX Serial No. xxxxxxxx Manufacturer: Chongqing Xinfeng Medical Instruments Co., Ltd. 1st floor, NO.8 group ,Industrial factory Zone B, Nanxi Group ,Industrial Zone , Hechuan District, 401520,Chongqing, PEOPLE’S REPUBLIC OF CHINA EC-Representative: Novasan.S.A. C/Churruca,18 28004 MADRID SPAIN...
5.3 Assembling: (Floor Type) Open the packaging box with care and check the contents of the device. It should consist of: -Device headlamp (installed with mineral coated plate), with spring loaded arm and control box. -Extendable Stand Pole -Five Point Base with Castors Assemble the device according to Figure 1 (see right diagram).
Page 12
Item Description Illustration On-Off power switch, Power on indicator(white), timer switch Timer Top panel ON indicator(green) are installed in this panel On-Off power Press the On-Off switch to turn On switch thedevice White Power indicator pilot is lighting Power on Indicator while the power is switchedOn.
Page 13
● Do not use the device directly on wet skin. 6.2 Turning Unit On: For Mechanical Timer, model CQ-12 a) Insert the plug into a grounded 230±10 volt AC output (wall) socket and the “white” POWER indicator lit, indicates the power is connected. At this time, the “green”...
Page 14
indicator lit. ii)For preset time of treatment - Push “Reset”, the screen will display “00:00”. Then push “Timer” to set working time. (The increments of each pushing of Button is 5 mins) c) Now Push “Start/Pause” and “green” (WORK) indicator lit, indicates the unit is warming up and countdown for operation is started.
7. Position for Treatment As a general guideline, adjust the distance between the treatment surface and the headlamp at an advisable distance (minimum 20 cm or 8”) to produce a warm and comfortable feeling. Actual placement must be dictated by patient comfort. Note that an advisable distance will provide therapeutic heating of 40°...
the headlamp further away or discontinue treatment. If the headlamp touches onto the body during treatment, the temperature may become too high and the skin may overheat and cause a skin burn. To minimize the impact of dry skin after treatment, apply moisturizing lotion to the treatment area after each session.
12. Specification Model CQ-12, CQ-27, CQ-29, CQ-33, CQ-36 40° C-45° C at room temperature (an advisable Optimum skin temperature distance from the heater to surface of bare Skin) Voltage: 230V~ Power Source Frequency: 50 Hz Electrical Power 250 Watts Timer...
Page 18
Mechanical Speciation of different models of IR Therapeutic Lamp...
13. Explanation of Symbols The following symbols are used in this device. 14. Disposal There is generally no restricted “shelf-life” of the device. The device contains substances that can be harmful to the environment. When you discard the device, do not throw it away with the normal household waste but hand it in at an official collection point.
15. CE Compliance Model CQ-12/CQ-27/CQ-29/CQ-33/CQ-36 Infrared Therapeutic Lamp conforms to the requirements of Medical Device MDD 93/42/EEC. It is in compliance with the standards IEC 60601-1:2012 reprint and IEC 60601-1-2:2014. 16. Trouble shooting If your device does not seem to be operating properly, please check as: ●...
NOTE: MANUFACTURER will provide circuit diagrams, component parts list, descriptions and calibration instructions to assist to SERVICE PERSONNEL in parts repair. Attention: Detachable part to be replaced is limited to the power supply cord and headlamp. When replacing a new power supply cord, ensure this is done by a qualified Service Personnel, and pay attention to the following steps: Power OFF and unplug the device from the mains.
Page 22
Model CQ-12/CQ-27/CQ-29/CQ-33/CQ-36,including cables specified by the manufacturer.Otherwise,degradation of the performance of this equipment could result. Guidance and manufacturer’s declaration - electromagnetic immunity- for all EQUIPMENT and SYSTEMS...
Page 23
Guidance and manufacturer’s declaration – electromagnetic immunity- for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING...
Page 24
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM-...
Page 25
for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration – electromagnetic emissions- for all EQUIPMENT and SYSTEMS Guidance and manufacturer's declaration-electromagnetic immunity- for all EQUIPMENT and SYSTEMS...
Page 26
Guidance and manufacturer’s declaration – electromagnetic immunity – for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING...
Page 27
Recommended separation distances between portable and mobile...
RF communications equipment and the EQUIPMENT or SYSTEM for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING 18. Limited Warranty ● 90 days parts and labor. ● One (1) year parts. ● The customer is responsible for any transportation of shipping relative to the return of product during the warranty period.
Need help?
Do you have a question about the CQ-12 and is the answer not in the manual?
Questions and answers