Page Erreur ! Argument de commutateur inconnu./15INSTRUCTION MANUAL SAFE 2000/MAXI SAFE 2000 Introduction You are now in possession of a high quality microprocessor controlled class II cabinet, H OLTEN 2000, designed to provide protection of the operator, the surroundings and the work process itself against particle/microbiological contamination.
SAFE 2000/MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur ! Argument de commutateur inconnu./15 The H 2000 class II cabinet will not provide any protection for operator or OLTEN surroundings against harmful gases or vapours. Remove hands from the work chamber prior to activating sliding mode for the front window ! Installation 1.
Page 6
Page Erreur ! Argument de commutateur inconnu./15INSTRUCTION MANUAL SAFE 2000/MAXI SAFE 2000 Figure Erreur ! Argument de commutateur inconnu.. 2. The installation site for the unit should be draught-free and should be selected so that frequent passing of persons in front of the work opening is avoided. 3.
SAFE 2000/MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur ! Argument de commutateur inconnu./15 Working Rules Rules to follow prior to use... - The cabinets should be used indoors in a normal clinical environment with a temperature between 18 °C and 28 °C, normal humidity and atmospheric pressure. - All the cabinets are designed and produced to be used in clean environments.
Page Erreur ! Argument de commutateur inconnu./15INSTRUCTION MANUAL SAFE 2000/MAXI SAFE 2000 Operation Fig. A: 1. Button ON / OFF for UV light (optional). - The UV light can only be switched on when the chamber light is off and front window brought in lowest position.
Page 9
SAFE 2000/MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur ! Argument de commutateur inconnu./15 ØHow to set time and day of the week... Turn the lock-switch to position 1 and..1. Press the [ENTER] button while connecting the cabinet to power 2. Press the [MENU] button once, and the display will flash for setting minutes. 3.
Page 10
Page Erreur ! Argument de commutateur inconnu./15INSTRUCTION MANUAL SAFE 2000/MAXI SAFE 2000 3. Press the [ENTER] button to start count down. When the stopwatch has counted down to 00:00 the acoustic alarm sounds. ØAlarm... When the alarm is activated, the acoustics alarm sounds and the display light flashes indicating that the cabinet does not provide full protection for the operator and the product.
Page 11
SAFE 2000/MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur ! Argument de commutateur inconnu./15 Erreur ! Argument de commutateur inconnu.
SAFE 2000/MAXI SAFE 2000 INSTRUCTION MANUALPage Erreur ! Argument de commutateur inconnu./15 Maintenance Recommended maintenance: • DAILY: The work area is disinfected. Also lift the table tops and carefully wipe the backside and the trough. • WEEKLY: Wipe the exterior of the unit with a mild household detergent. HOW TO CLEAN THE INSIDE FRONT WINDOW: 1.
Page 14
Page Erreur ! Argument de commutateur inconnu./15INSTRUCTION MANUAL SAFE 2000/MAXI SAFE 2000 Erreur ! Argument de commutateur inconnu.
Page 16
RECOMMENDED DISINFECTANTS FOR USE IN CLASS II CABINETS FROM H OLTEN AMIN Note: Most disinfectants are mixtures containing several components. Avoid the use of products containing iodine, chloramine, chlorine dioxide, hypochloride, phenol, and perchloric acid. Useable disinfectants contain alcohol or aldehydes. Recommended disinfectants are 70% (volume) pure ethanol 2% (volume) glutoric aldehyde...
Page 17
LOG-BOOK Type : Product name : Serial no. : Supplier : Service interval : Laboratory : Service / Fault Notified Repaired Remark / fault Caused by / action Service Carried out repair occurred on / date / date internal/external...
Page 18
Service / Fault Notified Repaired Remark / fault Caused by / action Service Carried out repair occurred on / date / date internal/external...
Page 19
Service / Fault Notified Repaired Remark / fault Caused by / action Service Carried out repair occurred on / date / date internal/external...
Page 21
Heto EC-Declaration of conformity Holten Form.nr.: 00243167 DS/EN 45014 Annexe A (Recommended form for declaration of conformity) Heto-Holten A/S Gydevang 17-19, DK-3450 Allerød declare under our sole responsibility that the product Model: S 2000 / M 2000 to which this declaration relates is in conformity with the following standard(s) or other normative document(s): EN 292-1:1993 Safety of machinery (Basic concepts - General principles) EN 292-2:1993 Safety of machinery...