  • Page 1 Document No. 1800-45 IonBar™ (Ionizing Bar) © Copyright 2016 Terra Universal Inc. All rights reserved. • Revised November 2016 Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 2 For installation in a hood, the overall hood width should be at least 2" greater than the length of the desired IonBar™. NOTE Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 3 Figure 2: The airflow from the fan module is required to disperse the generated ions throughout the glovebox Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 4 Installation on Laminar Flow Hoods When ordered with a laminar flow hood, the IonBar™ is installed in the exact same way as with a fan/filter unit in a cleanroom. The mounting brackets will rest on the top edges of the hood and the FFU will be placed directly on top of the brackets, holding the IonBar™...
  • Page 5 28 seconds (from ± 1,000 volts to 100 volts), measured in a 4-foot laminar flow hood with air speed of 90 feet/minute. Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 6 See Section 3: Maintenance (Page 5) Poisonous/Hazardous Gas N/A; no hazardous gases Flammable Liquids N/A; no flammable liquids Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...
  • Page 7 See Terra’s “Policy and Procedure for Returned Goods.” Thank you for ordering from Terra Universal! Terra Universal, Inc. • TerraUniversal.com • 800 S. Raymond Ave. • Fullerton, CA 92831 • TEL: (714) 578-6000 • FAX: (714) 578-6020...