Instructions for Use
File Prep
Always use new files or previously used instruments cleaned with ultrasonic cleaning prior to placing them into a new
piece of e-Foam insert for steam autoclave processing. Files can be placed directly into a new single use e-Foam
insert and then placed into a standard pack for steam autoclave processing. Avoid placing heavy objects onto the
autoclave pack to avoid deforming the foam.
A
B
Processing Instructions:
Single Use e-Foam Inserts
DO NOT REUSE
Jordco products are covered by one or more U.S. or foreign patents or patent applications owned by Jordco, Inc. of Beaverton, OR, USA. See www.jordco.com for details. JORDCO, ENDORING, E-FOAM, ENDOGEL, E-DX,
and FILECADDY are trademarks of Jordco, Inc., registered in the United States and/or other countries. The blue color of the e-Foam inserts and the endodontic organizer, the yellow color of the e-Foam inserts, and the
colors yellow and blue-yellow together on e-Foam insert packaging, are all trademarks of Jordco, Inc., registered in the United States. See the above website for further information. (1704)
A. Place the Jordco e-Foam insert into the EndoRing cup. Pull the
e-Foam insert out the bottom of the cup to seat the e-Foam
insert.
B. Arrange hand files, rotary files or other instruments such as
Gates Gliddens into the e-Foam insert. Remove all file bioburden
prior to autoclaving files in the e-Foam insert. When using rubber
stops, utilize the stop locks on either the built-in ruler or the
detachable metal ruler to increase or decrease the instrument's
working length. The rulers perform well with paper points, gutta
percha and many other endodontic instruments. To efficiently
clean instruments simply pass them in and out of the e-Foam
insert. You can also clean instruments by sliding them through
the corrugated area on top of the e-Foam insert. Rubber stops
can be safely added to instruments using the ports located on
the back of the EndoRing cup.
Steam autoclave a new e-Foam insert prior to use. Unused or ultrasonically cleaned
files may be placed in the e-Foam insert prior to processing. Autoclaving will not
remove bioburden left behind from prior usage. The e-Foam insert, along with files,
may be placed in an FDA approved autoclave pouch/bag prior to processing. Do not
exceed temperature of 275°F (135°C). To avoid deforming the e-Foam insert, do not
load or stack items on top of the e-Foam insert during autoclave processing.
Dispose of after each use.

Jordco e-Foam Inserts

REV1708

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ENDORING II and is the answer not in the manual?

Questions and answers

Summary of Contents for Jordco ENDORING II

  • Page 1 Jordco products are covered by one or more U.S. or foreign patents or patent applications owned by Jordco, Inc. of Beaverton, OR, USA. See www.jordco.com for details. JORDCO, ENDORING, E-FOAM, ENDOGEL, E-DX, and FILECADDY are trademarks of Jordco, Inc., registered in the United States and/or other countries. The blue color of the e-Foam inserts and the endodontic organizer, the yellow color of the e-Foam inserts, and the colors yellow and blue-yellow together on e-Foam insert packaging, are all trademarks of Jordco, Inc., registered in the United States.
  • Page 2 F O R U S E REF# ERK2-s ENDORING ® with Gelwells Please read before using the Endoring II. Jordco, Incorporated Beaverton, Oregon US 97006 503/531-3904 TEL, 503/531-3757 F X, 800/752-2812 toll free US www.jordco.com I N I T I...
  • Page 3 To measure gutta percha or paper points simply grasp them with the fine tip of a standard cotton pliers. Place them into the channel and grasp them from the stop lock to get a working length. ENDORING II INSTRUCTION GUIDE 2 OF 4 P GES JREV0903...
  • Page 4 This convenient attachment allows you to store your lubricating gels or liquids close to the operating field. Jordco’s EndoGel is an ideal lubricant that can be loaded onto files and taken directly to the root canal. The single use clear cups must be disposed of after each use.
  • Page 5 Inadequate water can cause excess chamber temperatures and result in damage to the ring and its components. W RNING: DO NOT IMMERSE THE ENDORING OR ITS COMPONENTS IN SURGIC L MILK. ENDORING II INSTRUCTION GUIDE 4 OF 4 P GES JREV0903...