W.L. Walker ISOTHERM 2000 Using Manual

Advertisement

Quick Links

Using the
ISOTHERM
2000
W. L. Walker Co., Inc.
Tulsa,Oklahoma
(888) 397- 9255

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ISOTHERM 2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for W.L. Walker ISOTHERM 2000

  • Page 1 Using the ISOTHERM 2000 W. L. Walker Co., Inc. Tulsa,Oklahoma (888) 397- 9255...
  • Page 2 Take the stainless steel probe and place it in the fluid to be measured. After the probe has been in the fluid for approximately 30 to 45 seconds, turn on the ISOTHERM 2000 and read the current temperature. For the greatest accuracy, leave the probe in the fluid for five minutes or more.
  • Page 3 EVEREADY NO. 1222, RAYOVAC NO. A1604. To replace the battery you will need to remove the bottom end of the ISOTHERM 2000's handle. It unscrews similar to a common flashlight. Inside the handle, there is a 9V battery attached to a long connector with leads back into the handle..
  • Page 4 CALIBRATING THE ISOTHERM 2000 You will need to purchase a Calibration Tool (W.L. Walker part number 69550) in order to perform a calibration. The Calibration Tool The Calibration Tool is a small box that allows you to calibrate the ISOTHERM. It has four buttons which allow you to adjust the displayed temperature up and down until the desired temperature is displayed.
  • Page 5 Plug the ISOTHERM 2000 into the Calibration Tool 2000 The plug is located in the ISOTHERM's handle or calibration compartment(Reel) NOTE: The Isotherm 2000 MUST remain plugged into the Calibration Tool during the calibration process. Unplugging the Isotherm when moving from the cold bath to the hot bath will cause wrong temperature readings after calibration.
  • Page 6 Returning Units for Repair When you need to return an ISOTHERM or Probe and Cable assembly to W.L. Walker for repair, you will need to fill out a copy of the ISOTHERM REPAIR TAG located at the end of this...
  • Page 7 Improper calibration Clear the calibration and then see if you get a stable reading.(see the document: "Calibrating the ISOTHERM 2000" for instructions on clearing the calibration) The probe & cable is Return the probe and cable for repair Malfunctioning and/or replace the probe &...
  • Page 8 ISOTHERM REPAIR TAG FOR FASTEST SERVICE, SHIP TO: W.L. Walker Co., Inc. 1201 South Main Street Tulsa, OK 74119 CUSTOMER INFORMATION: DATE:____/____/____ Name: _____________________________ Address: _____________________________ _____________________________ Contact: ____________ Phone #: (__ __ __)__ __ __ - __ __ __ __...