Biolin Scientific Q-Sense Quick Start
Hide thumbs Also See for Q-Sense:

Advertisement

Explorer Operation

Quick Start
The most important steps to a successful measurement
Explorer Quick Start Guide Rev B
09-09-2016
Switch the electronics unit on, and
1
start QSoft 401. Wait a few
seconds for the program to establish
connection to the instrument. When
"Tact" shows a value on the information
bar below the main screen, the connection
is established.
Click on the temperature info bar
2
to access the Temperature
Control settings window.
(Optional) To display the
temperature stabilization before
3
starting the measurement, the
"Preacquisition form" could be opened.
Connect tubing and pump to the
flow module.
4
Before inserting a sensor, make sure
that the o-ring is lying flat in its bed.
Insert the sensor crystal with the
backside electrode positioned as the
symbol indicates; close the flow module
(finger tight) and place on chamber
platform.
For liquid measurements, fill the system
with the start solution/ buffer and make
sure no air is introduced.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Q-Sense and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biolin Scientific Q-Sense

  • Page 1 Explorer Operation Quick Start The most important steps to a successful measurement Switch the electronics unit on, and start QSoft 401. Wait a few seconds for the program to establish connection to the instrument. When “Tact” shows a value on the information bar below the main screen, the connection is established.
  • Page 2 In the main menu, go to “Acquisition” and then “Setup Measurement” and pick the overtones to be measured, by ticking the boxes. ► For standard measurements, make sure “Automatically optimize all resonances” is checked. ► Start searching for the resonances by clicking “Find all”. D values in water should be within 20% from the typical values listed above.

Table of Contents