Explorer Operation
Quick Start
The most important steps to a successful measurement
Explorer Quick Start Guide Rev B
09-09-2016
Switch the electronics unit on, and
1
start QSoft 401. Wait a few
seconds for the program to establish
connection to the instrument. When
"Tact" shows a value on the information
bar below the main screen, the connection
is established.
Click on the temperature info bar
2
to access the Temperature
Control settings window.
(Optional) To display the
temperature stabilization before
3
starting the measurement, the
"Preacquisition form" could be opened.
Connect tubing and pump to the
flow module.
4
►
Before inserting a sensor, make sure
that the o-ring is lying flat in its bed.
►
Insert the sensor crystal with the
backside electrode positioned as the
symbol indicates; close the flow module
(finger tight) and place on chamber
platform.
►
For liquid measurements, fill the system
with the start solution/ buffer and make
sure no air is introduced.