Contents Section Overview Page Overview of the SPD111V Unit Installation of the SPD111V Unit How to Hook Up the SPD111V Unit Section Control Panel Page The Control Panel Description of the Control Panel Section Auto & Manual Run Page The Manual Run The Auto Run Section Rotor Selection...
1.0 Overview of the SPD111V Unit Figure 1 = Front View Figure 2 = Side View Figure 3 = Back View Fig. 1 Top Cover (Radiant Cover Optional) Control Panel Front Panel Left Side Panel Anti Skid/Vibration Fig. 2 Right Side Panel Chamber View Window Vacuum Port AC Socket...
1.1 INSTALLATION OF THE SPD111V Figure 4-Rear View of System Vacuum Tubing VLP Vacuum Pump SPD111V- SpeedVac ® Concentrator RVT4104- Refrigerated VaporTrap 1.2 HOW TO HOOK UP THE SPD111V SPEEDVAC SYSTEM ® Unpack unit from the box and verify that all the parts match packing list. Read instruction manual carefully! If assistance is required, contact Thermo If the unit is part of a larger system, use the vacuum tubing supplied to...
2.0 SPD111V CONTROL PANEL 2.1 DESCRIPTION OF SPD111V CONTROL PANEL PRE-HEAT - use to pre-heat chamber to 45 C prior to or between runs. ° Once run is initiated the pre-heat stops. RC ON/OFF - Use to add radiant heat to chamber. Manual activation ON/OFF at any point during run.
3.0 THE MANUAL RUN This type of run will allow continuous running of the unit until it is turned off by the user. The heat timer is activate and can be set to a user defined time. Connect the unit to its required voltage. Turn the power switch to ON, located on the back of the unit.
3.1 THE AUTO RUN An automated run will run the unit for a pre-defined time as selected by the user. Refer to the “MANUAL RUN” section for start up. Place sample tubes in rotor so load is balanced. Secure it with rotor supplied knob (hand tight).
5.0 GUIDELINES FOR SOLVENT CHOICE Part I: Solvents suited for the SPD111V unit • Ethanol • Methanol • Water • Acetonitrile Part II: Solvents NOT suited for the SPD111V unit • Methylene Chloride • Chloroform • Ethyl Acetate • Hydrochloric Acid •...
6.0 MAINTENANCE/SERVICE Maintenance: The SPD111V SpeedVac requires no scheduled ® maintenance. Cleaning: The SPD111V SpeedVac should be cleaned if solvents spill on ® or inside the unit. Always clean up any spills immediately using absorbent paper towels. Always wear gloves when cleaning and dispose of paper towel in appropriate designated refuse containers.
7.0 SPECIFICATIONS/WARRANTY Model: SPD111V Bleeder Valve: Integrated Automatic Bleeder Valve Temperature Range: C-80 ° ° Volume Range/Tube: 18 x 150 mm Maximum Tube Capacity: Maximum Carrier Capacity: 2 (96-deepwell plates) see chart on page 6 Dimensions: (W x D x H) in.: 14 x 18 x 13 36 x 45 x 33 Weight: lbs.