1. Features, warranty & technical service This product is intended by the manufacturer Alerta Medical sells all its products through to be used on human beings for the purpose a network of distributors. In the event of of lifting, lowering and moving users. The user...
2. Safety precautions 2.2 Braking and moving Any person using or assembling these items must first read and understand the assembly, operating and cleaning instructions provided. While the hoist is stationary, the brakes Read all safety information contained in this should always be on, except when lifting manual (highlighted with the symbol or lowering a patient.
Page 5
2.4 Electrical safety precautions All the hoist’s electrical connections must comply with the International Electrotechnical Commission (IEC) standards. The power supply should be equipped with an earth- leakage circuit breaker with a maximum operating current of 30 mA, as prescribed by standard IEC 364-5-53.
Page 6
Model Alerta Stand 200 Powerlifter Picture 1 www.alertamedical.com PAGE 06...
3. Assembly 3.2 Joining the pre-assembled structures Remove packaging carefully. Please refer to page 8. No tools are necessary to assemble the Place the base structure on the ground with hoist, although some parts are considerably the back casters braked. Insert the platform heavy.
4. Operating towards the mast. To EXTEND the arm, release the button (L) and slide the mobile Instructions boom outwards and position it as much as you want. The security button (M) will remain 4.1 Recommendations fixed in the last position. To spread the legs, step on the left pedal.
4.3. Changing the battery • Using the hoist on terrain with a slope gradient greater than 5º (loaded with a To remove the battery from the charging base, patient). hold the upper handle, press and remove • Extreme, intensive use of the actuator carefully off its support.
the axis of vertebral column. • Dry the hoist thoroughly before using again. Adjust the belt for better security. • Use clean cloths, disposable or recyclable. • Use a diluted solution of detergent or 5.3. Raising up of user disinfectant, or a spray-on disinfectant. Before lifting the user, recline user’s back and check if the sling is correctly fastened.
Page 12
Contact your local components of the metal structure of the Alerta Medical distributor for any specific crane can be replaced. issue in connection with maintenance and/or 6.5. Transport and storage servicing.
7. Troubleshooting PROBLEM CAUSE REMEDY The actuator is The red emergency Turn the red button not working stop is activated clockwise to reactivate The battery is connected Unplug the black to the electricity mains power supply for recharging The battery is low Recharge the battery The battery is not Remove the battery...
8. Technical specifications Stand 200 Powerlifter www.alertamedical.com PAGE 14...
Page 15
Stand 200 Powerlifter Approx. 50-66 size (cm) 83,5 62,5 • Height: minimum 114cm (H) maximum 158cm (R) • Length: 113cm (X) • Internal width at maximum reach: 77cm • Reach from base with legs spread to 700mm: 57cm • Maximum reach from base: 6cm •...
Page 16
CS Lifesciences Europe Ltd The Black Church St. Mary’s Place Dublin 7 D07 P4AX Ireland eurep@cslifesciences.com Rev: 2021.10 Making cutting-edge medical equipment affordable Head Offi ce Alerta Medical 4 Symington Place Riverside Business Park Irvine KA11 5DE United Kingdom www.alertamedical.com...