Troubleshooting

Cannot connect the iCOquit® Smokerlyzer® to the app
Make sure Bluetooth® is enabled on your smart device
Battery may be flat
NOTE: Once the iCOquit's battery has become flat, the iCOquit® will need to be disposed of, please
see disposal guidance in Disposal section on page 30.
Network Error Occurred
Please make sure you are connected to the internet.
Precautions to be taken in the event of changes in the performance of the iCOquit®
Smokerlyzer®
Leave in clean, dry area with fresh air (free from VOCs) for 24hrs.
Precautions to be taken regarding the exposure of the iCOquit® to reasonably foreseeable
environmental conditions
Leave in clean, dry area with fresh air (free from VOCs) for 24hrs.
What if I can't hold my breath for the recommended 15 second breath hold time?
Bedfont® recommends that if you are unable to hold your breath for the recommended 15 seconds,
to hold your breath for 10 seconds before exhaling into the device.
If any of these issues persist, please contact Bedfont® or its distributors.

Cleaning

1. The iCOquit® Smokerlyzer® product is moulded with ABS and PC plastic, which is easy to keep
clean.
2. After you have finished using the iCOquit® Smokerlyzer® to obtain your results, wipe the outside
of the device with a lightly damp cloth, ensuring not to allow any water to enter the breath or
exhaust ports in the device.
3. NEVER use alcohol or cleaning agents containing alcohol or other organic solvents as long-term
exposure to these vapours will damage the CO sensor inside.
4. Under no circumstances should the instrument be immersed in liquid or splashed with liquid.
25